130 Installation and User’s Guide
D CV Subsystem Functional Verification Test
Understanding the CV Subsystem Functional Verification Test
Use the procedure provided in “Performing the CV Subsystem Functional
Verification Testto manually confirm the functionality of the
Agilent 4284A precision LCR meter. This procedure does not verify
performance to specification. This procedure confirms that the precision
LCR meter is ready for performance verification and/or operation by
confirming the following hardware functionality:
memory card read/write test
LED display test
LCD display test
handler I/F test
scanner I/F EEPROM read/write test
scanner I/F I/O test
bias current I/F I/O test
Required Tools
See Chapter 10 of the Agilent 4284A precision LCR meter operation
manual.