Agilent-technologies 85225F Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Hardware Agilent-technologies 85225F. Agilent Technologies 85225F User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 148
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes

Resumo do Conteúdo

Página 1 - Agilent 85225F

Agilent TechnologiesAgilent 85225F Performance Modeling SystemInstallation and User’s Guide

Página 2

10 Installation and User’s GuideIn This Guide...This guide provides instruction on installing, verifying, and servicing the system, as well as an int

Página 3 - Warnings, Cautions, and Notes

100 Installation and User’s Guide3 Verifying System Functionality This completes the functional verification procedure.15 For systems with the Agilent

Página 4

Verifying System Functionality 3Installation and User’s Guide 101 If you encounter a problem✔ Check the system connections and settings:1 system conne

Página 5 - Operator Safety Requirements

102 Installation and User’s Guide3 Verifying System Functionality

Página 6

103 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent Technologies4Servicing the SystemTo troubleshoot the system 104To r

Página 7

104 Installation and User’s Guide4 Servicing the System To troubleshoot the systemMost system problems are caused by faulty cabling or switch configur

Página 8

Servicing the System 4Installation and User’s Guide 105 To remove or replace a system componentFollow these steps to remove a system component for per

Página 9 - For Technical Assistance

106 Installation and User’s Guide4 Servicing the System Following the replacement of a system component, complete the procedure in “Performing the Sy

Página 10 - In This Guide

Servicing the System 4Installation and User’s Guide 107 To receive additional assistanceIf you would like assistance, visit the online assistance web

Página 11

108 Installation and User’s Guide4 Servicing the System To package the system for transportFollow these instruction to package the system in a shippin

Página 12 - Software

Servicing the System 4Installation and User’s Guide 109 17 Add tip indicators and appropriate labeling18 Secure crate to packaging base using band str

Página 13 - Contents

Installation and User’s Guide 11D CV Subsystem Functional Verification TestsThis appendix includes an Agilent 4284A precision LCR meter functional ve

Página 14 - 2 Installing the System

110 Installation and User’s Guide4 Servicing the System

Página 15

111 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent TechnologiesAEnhancing Measurement AccuracyTo enhance measurement a

Página 16

112 Installation and User’s GuideA Enhancing Measurement Accuracy To enhance measurement accuracyFollow these instructions to enhance the accuracy of

Página 17 - Performance Modeling System

Enhancing Measurement Accuracy AInstallation and User’s Guide 113 5 Always perform a measurement calibration on the network analyzer prior to making d

Página 18

114 Installation and User’s GuideA Enhancing Measurement Accuracy Understanding System Measurement CalibrationMeasurement accuracy is degraded by the

Página 19

Enhancing Measurement Accuracy AInstallation and User’s Guide 115 Performing a Coaxial System Measurement CalibrationComplete the following steps to p

Página 20 - The DC Subsystem

116 Installation and User’s GuideA Enhancing Measurement Accuracy If you encounter a problem✔ Inspect the connectors on the load, open, short, and thr

Página 21 - The Bias Networks

Enhancing Measurement Accuracy AInstallation and User’s Guide 117 Periodic System Component CalibrationA complete calibration verifies that the system

Página 22 - Component Integration

118 Installation and User’s GuideA Enhancing Measurement Accuracy

Página 23 - Figure 2 System Components

119 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent TechnologiesBDC Subsystem Functional Verification TestUnderstanding

Página 24

12 Installation and User’s GuideTyp efac eConventionsThis guide uses the following typeface conventions to describe various aspects of a particular h

Página 25

120 Installation and User’s GuideB DC Subsystem Functional Verification Test Understanding the DC Subsystem Functional Verification TestUse this proce

Página 26

DC Subsystem Functional Verification Test BInstallation and User’s Guide 121 Performing the DC Subsystem Functional Verification TestComplete these st

Página 27

122 Installation and User’s GuideB DC Subsystem Functional Verification Test For Agilent E5260A/70B DC subsystems:If you encounter a problem✔ Refer to

Página 28

123 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent TechnologiesCRF Subsystem Functional Verification TestUnderstanding

Página 29

124 Installation and User’s GuideC RF Subsystem Functional Verification Test Understanding the RF Subsystem Functional Verification TestUse the proced

Página 30

RF Subsystem Functional Verification Test CInstallation and User’s Guide 125 Performing the RF Subsystem Functional Verification TestIf the performanc

Página 31

126 Installation and User’s GuideC RF Subsystem Functional Verification Test 5 Test the forward transmission mode for channel 1.a From the Trace drop

Página 32 - Systems with Agilent 4156C

RF Subsystem Functional Verification Test CInstallation and User’s Guide 127 If you encounter a problem✔ Check the GPIB cable and connection.✔ Check t

Página 33

128 Installation and User’s GuideC RF Subsystem Functional Verification Test

Página 34 - The CV Subsystem

129 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent TechnologiesDCV Subsystem Functional Verification TestUnderstanding

Página 35 - Figure 9 System Components

Installation and User’s Guide 13 Contents1 Introducing the Agilent 85225F Performance Modeling SystemPerformance Modeling System Configuration Overvie

Página 36

130 Installation and User’s GuideD CV Subsystem Functional Verification Test Understanding the CV Subsystem Functional Verification TestUse the proced

Página 37

CV Subsystem Functional Verification Test DInstallation and User’s Guide 131 Performing the CV Subsystem Functional Verification TestIf the performanc

Página 38

132 Installation and User’s GuideD CV Subsystem Functional Verification Test If you encounter a problem✔ Check the GPIB cable and connection.✔ Check t

Página 39

133 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent TechnologiesE1/f Noise Subsystem Functional Verification TestUnders

Página 40

134 Installation and User’s GuideE 1/f Noise Subsystem Functional Verification Test Understanding the 1/f Noise Subsystem Functional Verification Test

Página 41

1/f Noise Subsystem Functional Verification Test EInstallation and User’s Guide 135 Performing the 1/f Noise Subsystem Functional Verification TestIf

Página 42

136 Installation and User’s GuideE 1/f Noise Subsystem Functional Verification Test If you encounter a problem✔ Check the GPIB cable and connection.✔

Página 43

137 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent TechnologiesFUnderstanding the Bias NetworksFeatures 138Characteris

Página 44

138 Installation and User’s GuideF Understanding the Bias Networks FeaturesUsing the Agilent 11612V K11 and K21 bias networks, you can simultaneously

Página 45

Understanding the Bias Networks FInstallation and User’s Guide 139 CharacteristicsTable 20 11612V Option K11/K21 Bias Network CharacteristicsParameter

Página 46 - The 1/f Noise Subsystem

14 Installation and User’s Guide 1/f Noise, CV, RF, and DC Measurement System Configuration 45Figure 16. System Block Diagram 46Figure 17. 1/f Noise M

Página 47

140 Installation and User’s GuideF Understanding the Bias Networks OperationEach bias network includes two bias tees, one for force and one for sense.

Página 48

141 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent TechnologiesGNetwork Analyzer Performance Specification SummaryNetw

Página 49 - Figure 18 System Components

142 Installation and User’s GuideG Network Analyzer Performance Specification Summary Network Analyzer System PerformanceThe following specifications

Página 50

Network Analyzer Performance Specification Summary GInstallation and User’s Guide 143 Measurement Port CharacteristicsMeasurement UncertaintyMeasureme

Página 51

144 Installation and User’s GuideG Network Analyzer Performance Specification Summary Transmission Measurements

Página 52

Installation and User’s Guide 145 IndexNumerics11612V. See bias networks2.4 mm (f-to-f) adapter, 69, 942.4 mm (f-to-m) adapter, 69, 81, 83, 85, 942.4

Página 53

146 Installation and User’s GuideIndex Hhandle kit, rack mount and, 70, 71, 105hard disk requirements, controller, 59hardware setup, IC-CAP, 89, 96IIC

Página 54

IndexInstallation and User’s Guide 147 schematic diagramsDC/RF signal flow, 32, 40, 41, 54, 55semi-rigid cables, 71shipment reception procedure, 65sit

Página 55

148 Installation and User’s GuideIndex

Página 56 - Instrument Control Interface

Installation and User’s Guide 15 Precautions for Avoiding Electrostatic Discharge 79To connect the bias networks 80Agilent 4156C Systems 80Agilent 415

Página 57 - The LAN/GPIB Gateway

16 Installation and User’s Guide If you encounter a problem 122C RF Subsystem Functional Verification TestUnderstanding the RF Subsystem Functional Ve

Página 58

17 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent Technologies1Introducing the Agilent 85225F Performance Modeling Sys

Página 59 - The System Controller

18 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Performance Modeling System Configuration OverviewThe sta

Página 60 - The Rack Cabinet

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 19 RF and DC Measurement System ConfigurationIn conjunction

Página 61 - Interference Standards

Notices© Agilent Technologies, Inc. 2005No part of this manual may be reproduced in any form or by any means (including electronic storage and retriev

Página 62 - Bias Network Characteristics

20 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System The RF SubsystemS- parameter device characterization is p

Página 63 - Installing the System

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 21 Agilent E5270B 8-Slot Precision Parametric Measurement Ma

Página 64

22 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Component IntegrationSystem component integration is perf

Página 65 - To receive the system

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 23 Figure 2 System Components

Página 66

24 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Table 3 Front Panel System Connections, with Agilent 4156

Página 67

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 25 Figure 3 Front Panel Connections with Agilent 4156C

Página 68 - To verify the shipment

26 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Table 4 Front Panel System Connections, with Agilent E526

Página 69

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 27 Figure 4 Front Panel Wiring Diagram with Agilent E5260A o

Página 70

28 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Table 5 Rear Panel System Connections with Agilent 4156CC

Página 71

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 29 Figure 5 Rear Panel Wiring Diagram with Agilent 4156C

Página 72

Installation and User’s Guide 3Safety and Regulatory InformationWarnings, Cautions, and NotesThis installation and user’s guide utilizes the followin

Página 73

30 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System CV, RF, and DC M easure ment Sys tem C onfigu rationTabl

Página 74 - To install the work surface

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 31 Figure 6 Rear Panel Wiring with Agilent E5260A or E5270B

Página 75

32 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Figure 7 DC/RF Cabling Diagram - DC and RF ConfigurationS

Página 76

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 33 CV, RF, and DC Measurement System ConfigurationWith the a

Página 77

34 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System The CV SubsystemThe Agilent 4284A precision LCR meter pro

Página 78

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 35 Figure 9 System Components

Página 79 - Installing the System 2

36 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Table 7 Front Panel System ConnectionsComponent Informati

Página 80 - To connect the bias networks

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 37 Figure 10 Front Panel Wiring Diagram

Página 81

38 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Table 8 Rear Panel System ConnectionsComponent Informatio

Página 82

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 39 Figure 11 Rear Panel Wiring Diagram

Página 83

4 Installation and User’s GuideSafety Symbols and Instrument MarkingsSymbols and markings in documentation and on instruments alert you to potential

Página 84 - Agilent E5260A/70B Systems

40 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Figure 12 DC/RF Cabling Diagram - DC and RF Configuration

Página 85

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 41 Figure 13 DC/RF Cabling Diagram - Parametric Configuratio

Página 86

42 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System The Low Leakage Switch MainframeThe Agilent E5250A is use

Página 87

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 43 Figure 14 Rear Panel Wiring Diagram including Low Leakage

Página 88

44 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Figure 15 DC/RF Cabling Diagram - Parametric Configuratio

Página 89

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 45 1/f Noise, CV, RF, and DC Measurement System Configuratio

Página 90

46 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System The 1/f Noise SubsystemThe Agilent 35670A dynamic signal

Página 91

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 47 Figure 17 1/f Noise Measurement Block Diagram

Página 92

48 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Component IntegrationSystem component integration is perf

Página 93

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 49 Figure 18 System Components

Página 94

Installation and User’s Guide 5Operator Safety RequirementsThe following general safety precautions must be observed during all phases of operation o

Página 95

50 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Table 10 Front Panel System ConnectionsComponent Informat

Página 96

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 51 Figure 19 Front Panel Wiring Diagram

Página 97

52 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Table 11 Rear Panel System ConnectionsComponent Informati

Página 98

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 53 Figure 20 Rear Panel Wiring Diagram

Página 99

54 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Figure 21 DC/RF Cabling Diagram - DC and RF Configuration

Página 100

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 55 Figure 22 DC/RF Cabling Diagram - Parametric Configuratio

Página 101 - If you encounter a problem

56 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Instrument Control InterfaceInstrument control interface

Página 102

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 57 The LAN/GPIB GatewayThe Agilent E5810A LAN/GPIB gateway p

Página 103 - Servicing the System

58 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System Figure 23 Rear Panel Wiring Diagram for LAN/GPIB Gateway

Página 104 - To troubleshoot the system

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 59 The System ControllerA customer- furnished UNIX workstati

Página 105 - To remove a system component

6 Installation and User’s Guidesafety standards of design, manufacture, and intended use of the product. Agilent Technologies, Inc. assumes no liabil

Página 106 - To order replacement parts

60 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System The Rack Cabinet The system is housed in a 1.6 meter rack

Página 107

Introducing the Agilent 85225F Performance Modeling System 1Installation and User’s Guide 61 Performance Characteristics and SpecificationsSupplementa

Página 108 - 4 Servicing the System

62 Installation and User’s Guide1 Introducing the Agilent 85225F Performance Modeling System RF Subsystem Performance SpecificationsThe overall perfor

Página 109

63 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent Technologies2Installing the SystemTo prepare the installation site 6

Página 110

64 Installation and User’s Guide2 Installing the System To prepare the installation siteFollow these steps to prepare the site for system installation

Página 111 - Agilent Technologies

Installing the System 2Installation and User’s Guide 65 To receive the systemFollow these steps to store, inspect, and confirm the system shipment.To

Página 112

66 Installation and User’s Guide2 Installing the System To unpack the shipment crate containing the rack cabinetThe racked system is shipped upright s

Página 113

Installing the System 2Installation and User’s Guide 67 3 Unload the system a Remove the two brace bolts attaching the side brace assembly to the bott

Página 114 - Required Tools

68 Installation and User’s Guide2 Installing the System To verify the shipmentUse Table 18 “System Receiving Checklist and Replaceable Parts” to:• co

Página 115

Installing the System 2Installation and User’s Guide 69 12 Complete the receiving checklist.a Compare the Bill of Materials to the system components r

Página 116

Installation and User’s Guide 7Mains powerGround the systemBefore applying powerCAUTIONThe mains cable shall be permanently connected to the premise

Página 117

70 Installation and User’s Guide2 Installing the System 16494A Option 001 Triaxial cable, 1.5 meter16494A Option 002 Triaxial cable, 3 meter16494B Opt

Página 118

Installing the System 2Installation and User’s Guide 71 5063-9221 Rack mount kit with handles, 3-EIA5063-9222 Rack mount kit with handles, 4-EIA5063-9

Página 119

72 Installation and User’s Guide2 Installing the System E5252A 10 x 12 switch matrixE5260A 8-slot high speed parametric measurement mainframeE5260A Op

Página 120

Installing the System 2Installation and User’s Guide 73 E8364B Option 022 Extended memoryE8364B Option 080 Frequency offsetE8364B Option 081 Reference

Página 121

74 Installation and User’s Guide2 Installing the System To install the work surfaceIf the system will not be used with a probe station, install the wo

Página 122

Installing the System 2Installation and User’s Guide 75 3 Attach the work surface to the support rails.a Slide the work surface onto the support rails

Página 123

76 Installation and User’s Guide2 Installing the System To ensure your safety while using the system This product has been designed and tested in acco

Página 124

Installing the System 2Installation and User’s Guide 77 4 Be aware of potential shock hazards during floating-ground measurements.a Do not touch any o

Página 125

78 Installation and User’s Guide2 Installing the System Precautions for Performing Floating-Ground MeasurementsIC- CAP measurements can be performed w

Página 126

Installing the System 2Installation and User’s Guide 79 Precautions for Avoiding Electrostatic DischargeNever operate the system without taking precau

Página 127

8 Installation and User’s GuideFuses and breakersBefore cleaning the systemOvercurrent protectionStatement of Compliance and Declaration of Conformit

Página 128

80 Installation and User’s Guide2 Installing the System To connect the bias networksFollow these steps to connect the bias networks to the system, and

Página 129

Installing the System 2Installation and User’s Guide 81 3 Connect and route the triaxial cable from 4156C HRSMU3 FORCE connector to the port 2 bias ne

Página 130

82 Installation and User’s Guide2 Installing the System Agilent 4156C Systems with Agilent 41501B Expander BoxTo connect the bias networksStep Action

Página 131

Installing the System 2Installation and User’s Guide 83 4 Connect the Agilent E8364B test ports to the bias networks.a Connect one end of the port 1

Página 132

84 Installation and User’s Guide2 Installing the System Agilent E5260A/70B SystemsTo connect the bias networksStep Action Notes1 For Agilent E5260A/70

Página 133 - Verification Test

Installing the System 2Installation and User’s Guide 85 3 Connect the triaxial cables from the E5260A/70B SMUs to the bias networks.a Connect the tria

Página 134

86 Installation and User’s Guide2 Installing the System 6 Connect the bias networks to the device under test.a Connect one semi-rigid cable to the RF/

Página 135

Installing the System 2Installation and User’s Guide 87 To switch on power to the systemNOTEProper system operation is dependent on the sequence in wh

Página 136

88 Installation and User’s Guide2 Installing the System 13 On the 4156C, press System > [MISCELLANEOUS], highlight the 4156C value in the GPIB ADDR

Página 137

Installing the System 2Installation and User’s Guide 89 To configure the LAN/GPIB gateway for functional verificationIf your system includes an Agilen

Página 138 - Features

Installation and User’s Guide 9Compliance with German Noise RequirementsThis is to declare that this instrument is in conformance with the German Re

Página 139 - Characteristics

90 Installation and User’s Guide2 Installing the System

Página 140 - Operation

91 Agilent 85225F Performance Modeling SystemInstallation and User’s GuideAgilent Technologies3Verifying System FunctionalityTo choose a verification

Página 141 - Specification Summary

92 Installation and User’s Guide3 Verifying System Functionality To choose a verification processSystem functionality can be verified using several di

Página 142 - Dynamic Range

Verifying System Functionality 3Installation and User’s Guide 93 2 If you would like to manually check the functionality of the Agilent E8364B PNA Ser

Página 143 - Measurement Uncertainty

94 Installation and User’s Guide3 Verifying System Functionality Understanding the System Functional Verification TestThe system functional verificati

Página 144 - Transmission Measurements

Verifying System Functionality 3Installation and User’s Guide 95 Performing the System Functional Verification TestComplete the following steps to ver

Página 145

96 Installation and User’s Guide3 Verifying System Functionality 3 Start the IC-CAP software.*PC version:a Select Start > Programs > IC-CAP 2004

Página 146

Verifying System Functionality 3Installation and User’s Guide 97 7 Open the attenuator test model in IC-CAP.a From the IC-CAP/Main window menu bar, ch

Página 147

98 Installation and User’s Guide3 Verifying System Functionality 10 Configure the instrument state for calibration.a Disconnect the attenuator from th

Página 148

Verifying System Functionality 3Installation and User’s Guide 99 13 Make the DC and RF measurements in IC-CAP.a Reconnect the attenuator.b In the Atte

Comentários a estes Manuais

Sem comentários