Vista de página 22
10%-90% variations
As designs move to smaller process technology the
edge rates are increasing
USB 2.0 specification limits RT/FT to 500ps as
measured at 10%-90%
Errors in consistency of these measurements due to:
Shape of signal edge
Noise
Amplitude variations
Probe peaking due to insufficient BW
Slow corners give 10/90
rise time that is slower than
actual edge rate
RT/FT faster than 400ps have observed
Oscilloscope AND Probe BW must be at least 2.5Ghz
to accurately observe and measure these fast edges
Rise and Fall time