Specifications – Agilent 1260 Infinity Standard Autosampler
Specifications for 1260 Infinity Standard Autosampler
Valid with standard 100 µL metering head installed
Injection range
0.1- 100 µL in 0.1 µL increments
Up to 1500 µL with multiple draw (hardware modification required)
Precision <0.25% RSD from 5-100 µL, <1% from 1-5 µL, variable volume
Sample viscosity range 0.2 - 5 cp
Sample capacity
100 x 2-mL vials in 1 tray
40 x 2-mL vials in 1/2 tray
15 x 6-mL vials in 1/2 tray (Agilent vials only)
Injection cycle time Typically 50 s depending on draw speed and injection volume
Carryover Typically < 0.1 %, < 0.05 % with external needle cleaning
Pressure operating range Up to 600 bar (60 MPa or 8702 psi)
Specifications for 1260 Infinity Standard Autosampler
Valid with standard 900 µL metering head installed
Injection range
0.1 - 900 µL in 0.1 µL increments (recommended 1 µL increments)
Up to 1800 µL with multiple draw (hardware modification required)
Precision
Typically < 0.5 % RSD of peak areas from 5 - 2000 µL
Typically < 1 % RSD of peak areas from 2000 - 5000 µL
Typically < 3 % RSD of peak areas from 1 - 5 µL
Variable volume
Sample viscosity range 0.2 - 5 cp
Sample capacity
100 x 2-mL vials in 1 tray
40 x 2-mL vials in 1/2 tray
15 x 6-mL vials in 1/2 tray (Agilent vials only).
Injection cycle time Typically 50 s depending on draw speed and injection volume
Carryover Typically < 0.1 %, < 0.05 % with external needle cleaning
Pressure operating range Up to 200 bar (20 MPa or 2901 psi)