E - IEC Mode Command Summary
172
FORMat:BORDer
This command sets the byte order of IEEE floating point values returned within Arbitrary Block Response
Data. When NORMal is selected, the first byte sent is the sign bit and seven most significant bits of the
exponent, and the last byte sent is the least significant byte of the mantissa. This ordering is most useful
for big-endian controllers such as those that use Motorola processors.
When SWAPped is selected, the least significant byte of the mantissa is sent first and the sign bit and
seven most significant bits of the exponent are sent last. This ordering is most useful for little-endian
controllers such as those that use Intel processors.
Command Syntax
FORMat:BORDer <CRD>
Parameters
NORMal | SWAPped
*RST Value
NORMal
Examples
FORM:BORD SWAP
Query Syntax
FORMat:BORDer?
Returned Parameters
<CRD>
Related Commands
FORM[:DATA] MEAS:ARR:CURR:DC?
MEAS:ARR:VOLT:DC? MEAS:ARR:CURR:HARM[:AMPL]?
MEAS:ARR:VOLT:FLUC:ALL?