Agilent Technologies E1345A Manual do Utilizador Página 54

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 93
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 53
Relay Multiplexer Command Reference 55
INITiate
The INITiate subsystem selects continuous scanning cycles and starts the scanning
cycle.
Subsystem Syntax INITiate
:CONTinuous <
mode>
:CONTinuous?
[:IMMediate]
:CONTinuous
INITiate:CONTinuous <mode> enables or disables continuous scanning cycles for
the switchbox.
Parameters
Comments Continuous Scanning Operation: Continuous scanning is enabled with the
INITiate:CONTinuous ON or INITiate:CONTinuous 1 command. Sending the
INITiate[:IMMediate] command closes the first channel in the channel list. Each
trigger from a trigger source selected by the
TRIGger:SOURce command
advances the scan through the channel list. A trigger at the end of the channel
list closes the first channel in the list and the scan cycle repeats.
Non-Continuous Scanning Operation: Non-Continuous scanning is enabled
with the
INITiate:CONTinuous OFF or INITiate:CONTinuous 0 command.
Sending the
INITiate[:IMMediate] command closes the first channel in the
channel list. Each trigger from a trigger source selected by the
TRIGger:SOURce command advances the scan through the channel list. A
trigger at the end of the channel list opens the last channel in the list and the
scanning cycle stops.
Stopping Continuous Scans: See the ABORt command.
Related Commands: ABORt, ARM:COUNt, TRIGger, TRIGger:SOURce
*RST Condition: INITiate:CONTinuous OFF
Example Enabling Continuous Scans
INIT:CONT ON
Enables continuous scanning.
SCAN (@100:115)
Sets channel list.
INIT
Starts scanning cycle.
Parameter Name Parameter Type Range of Values
<
mode>
boolean 0|1|ON|OFF
Vista de página 53
1 2 ... 49 50 51 52 53 54 55 56 57 58 59 ... 92 93

Comentários a estes Manuais

Sem comentários