Página 1 - EMI Receiver

Agilent TechnologiesGetting Started and Troubleshooting GuideAgilent X-Series EMI ReceiverThis manual provides documentation for the following:MXE EMI

Página 2 - Safety Notices

10 Quick-StartInitial InspectionInitial InspectionInspect the shipping container and the cushioning material for signs of stress. Retain the shipping

Página 3 - Compliance

100 Using Microsoft Windows XPWindows Shortcuts and Miscellaneous TasksWindows startup folderAll Windows XP systems include a special folder, called t

Página 4

101Agilent X-Series Signal AnalyzersGetting Started Guide6 Troubleshooting“Check the Basics” on page 102“Problems with Microsoft Windows XP” on page 1

Página 5 - 4 Instrument Operating System

102 TroubleshootingCheck the BasicsCheck the Basicso Is there power at the receptacle? o Is the instrument turned on? Check to see if the green LED be

Página 6

103TroubleshootingCheck the Basicso If the previously performed alignments did not resolve the problem, press System, Alignments, Restore Align Defaul

Página 7

104 TroubleshootingProblems with Microsoft Windows XPProblems with Microsoft Windows XPThe Microsoft Windows XP operating system settings have been op

Página 8

105TroubleshootingReturning an EMI Receiver for ServiceReturning an EMI Receiver for ServiceCalling Agilent Technologies Agilent Technologies has offi

Página 9

106 TroubleshootingReturning an EMI Receiver for ServiceRead the Warranty The warranty for your instrument is in the front of your Specifications Guid

Página 10 - Initial Inspection

107TroubleshootingReturning an EMI Receiver for Service3. Surround the equipment with three to four inches of packing material and prevent the equipme

Página 11 - Cooling and rack mounting

108 TroubleshootingReturning an EMI Receiver for Service

Página 12

IndexGetting Started Guide 109Symbols.NET, 49Aactive function, 29adda network connection, 47a printer, 48Administrator logon, 43Adobe Acrobat Reader,

Página 13

11Quick-StartInstrument Location and Rack Mounting RequirementsInstrument Location and Rack Mounting RequirementsLocating the instrumentMake sure that

Página 14

IndexGetting Started Guide 110Tab settings, 48user accounts, 47virus protection, 47Microsoft Windows XP system settings, 47mouseless help system use,

Página 15

IndexGetting Started Guide 111set time and date, 48Shortcut keys, 97Startup Folder, 100System Properties, 48Tab settings, 48taskbarauto-hide, 99user a

Página 16

112 Getting Started GuideIndex

Página 17

12 Quick-StartTurning on the EMI Receiver the First TimeTurning on the EMI Receiver the First TimeInitial power-on of the instrument can be accomplish

Página 18

13Quick-StartTurning on the EMI Receiver the First Time3. Launch Microsoft Setup • Highlight Microsoft Windows XP Professional and press Enter to star

Página 19

14 Quick-StartTurning on the EMI Receiver the First Time4. Reboot and logon • Click OK When the instrument restarts, the following message window appe

Página 20 - Instrument Information

15Quick-StartTurning on the EMI Receiver the First Time7. View a signal a. Press Mode, Spectrum Analyzer.b. Press Input/Output, RF Calibrator, 50 MHz.

Página 21

16 Quick-StartTurning on the EMI Receiver the First TimeInitializing the instrument using front-panel navigation The following keys are used in the in

Página 22 - Instrument Maintenance

17Quick-StartTurning on the EMI Receiver the First Time2. Launch Microsoft Setup • Highlight Microsoft Windows XP Professional and press Enter to star

Página 23

18 Quick-StartTurning on the EMI Receiver the First Time3. Reboot and logon • Press Enter When the instrument restarts, the following message window a

Página 24

19Quick-StartAnti-Virus Software and FirewallsAnti-Virus Software and FirewallsNo anti-virus software is shipped with the EMI Receiver. It is recommen

Página 25 - Features

Notices© Agilent Technologies, Inc. 2012No part of this manual may be reproduced in any form or by any means (including electronic storage and retriev

Página 26 - Front-Panel Features

20 Quick-StartShortening the Instrument Power-On TimeShortening the Instrument Power-On TimeThe instrument desktop includes a shortcut to the Configur

Página 27 - Front and Rear Panel Features

21Quick-StartInstrument InformationAC power cordThe EMI Receiver is equipped with a three-wire power cord, in accordance with international safety sta

Página 28

22 Quick-StartInstrument InformationInstrument MaintenanceCleaning the instrumentWARNING To prevent electrical shock, disconnect the instrument from m

Página 29 - Overview of key types

23Quick-StartInstrument InformationBattery InformationThe EMI Receiver uses a lithium battery located on the CPU board. This is not an operator replac

Página 30

24 Quick-StartInstrument Information• Be sure that all instruments are properly earth-grounded to prevent build-up of static charge.• Perform work on

Página 31

25Agilent X-Series Signal AnalyzersGetting Started Guide2 Front and Rear Panel FeaturesThis section describes the following features:“Front-Panel Feat

Página 32

26 Front and Rear Panel FeaturesFront-Panel FeaturesFront-Panel FeaturesItemDescription# Name1 Menu Keys Key labels appear to the left of the menu key

Página 33 - Display Annotations

27Front and Rear Panel FeaturesFront-Panel Features8 Back Space Key Press this key to delete the previous character when entering alphanumeric informa

Página 34

28 Front and Rear Panel FeaturesFront-Panel Features17 Ctrl Key Ctrl works the same as a PC keyboard. Use it to navigate in Windows applications or to

Página 35 - Rear-Panel Features

29Front and Rear Panel FeaturesFront-Panel FeaturesOverview of key typesThe keys labeled FREQ Channel, System, and Marker Functions are all examples o

Página 36

3WarrantyThis Agilent technologies instrument product is warranted against defects in material and workmanship for a period of one year from the date

Página 37 - Front and Rear Panel Symbols

30 Front and Rear Panel FeaturesFront-Panel FeaturesWhen a menu first appears, one key label is highlighted to show which key is the default selection

Página 38

31Front and Rear Panel FeaturesFront-Panel FeaturesIn some key menus, a key label is highlighted to show which key has been selected from multiple ava

Página 39 - 3 Using the EMC32 Software

32 Front and Rear Panel FeaturesFront-Panel Features

Página 40 - Configuring the MXE

33Front and Rear Panel FeaturesDisplay AnnotationsDisplay AnnotationsThis section describes the display annotation as it is on the EMI Receiver displa

Página 41 - 4 Instrument Operating

34 Front and Rear Panel FeaturesDisplay Annotations5 Settings panel - displays system information that is not specific to any one application.• Input/

Página 42 - Agilent Software Installed

35Front and Rear Panel FeaturesRear-Panel FeaturesRear-Panel FeaturesEMI ReceiverItemDescription#Name1 EXT REF IN Input for an external frequency refe

Página 43 - User Accounts

36 Front and Rear Panel FeaturesRear-Panel Features10 TRIGGER 2 OUTA trigger output used to synchronize other test equipment with the instrument. Conf

Página 44 - Trial licenses

37Front and Rear Panel FeaturesFront and Rear Panel SymbolsFront and Rear Panel Symbols This symbol is used to indicate power ON (green LED). This

Página 45 - After Initial Purchase

38 Front and Rear Panel FeaturesFront and Rear Panel Symbols

Página 46

39Agilent X-Series Signal AnalyzersGetting Started Guide3 Using the EMC32 SoftwareThe MXE is designed to be compatible with EMC32 software. The follow

Página 48

40 Using the EMC32 SoftwareHow to Set Up and Run the EMC32 SoftwareHow to Set Up and Run the EMC32 SoftwareTo be compatible with the EMC32 software, t

Página 49

41Agilent X-Series Signal AnalyzersGetting Started Guide4 Instrument Operating SystemThis chapter describes the Microsoft Windows XP configuration and

Página 50 - Configuring LAN

42 Instrument Operating SystemAgilent Software InstalledAgilent Software InstalledAgilent signal analyzer softwareThe N9060B Agilent Spectrum Analyzer

Página 51 - Windows Security

43Instrument Operating SystemUser AccountsUser AccountsUser loginThe default user account that ships from the factory is “Instrument” with the passwor

Página 52 - Automatic updates

44 Instrument Operating SystemAgilent X-Series EMI Receiver LicensingCustomer creation of accountsYou can create additional user accounts and decide o

Página 53 - Virus protection

45Instrument Operating SystemLicensing New Measurement Application Software - After Initial PurchaseExample: N9071A-TRLIf a product has multiple tiers

Página 54 - Spyware protection

46 Instrument Operating SystemLicensing New Measurement Application Software - After Initial Purchase3. Load the license file • Connect the USB storag

Página 55 - System Maintenance

47Instrument Operating SystemWindows ConfigurationWindows ConfigurationThe Windows settings have been optimized for the best measurement performance.

Página 56 - USB Connections

48 Instrument Operating SystemWindows ConfigurationSettings that must not be changedAvoid changing any settings in this section. Changes to the follow

Página 57

49Instrument Operating SystemWindows ConfigurationIn addition, DO NOT:• Add, delete, or modify disk drive partitions.• Delete or modify Agilent regist

Página 58

Contents 51 Quick-StartInitial Inspection 10Shipping Problems? 10Instrument Location and Rack Mounting Requirements 11Locating the instrument 11Coolin

Página 59 - Disk Drive Recovery Process

50 Instrument Operating SystemConfiguring PrintersConfiguring PrintersPrinters are configured using the Microsoft Windows Control Panel. It is easily

Página 60

51Instrument Operating SystemWindows SecurityWindows SecurityMicrosoft recommends the following to ensure the instrument's Windows XP operating s

Página 61 - Updating the software

52 Instrument Operating SystemWindows SecurityWindows FirewallThe instrument is shipped with the Windows Firewall enabled.NOTE The window may look sli

Página 62

53Instrument Operating SystemWindows SecurityNOTE Be aware that downloading and installing Windows Updates can be network and CPU usage intensive (imp

Página 63

54 Instrument Operating SystemWindows SecuritySpyware protectionThere is no anti-spyware software installed on the instrument. This should not be a pr

Página 64

55Instrument Operating SystemSystem MaintenanceSystem MaintenanceBackupIt is recommended that you have a regular backup strategy. Your IT department m

Página 65 - 5 Using Microsoft Windows XP

56 Instrument Operating SystemUSB ConnectionsUSB ConnectionsAll of the USB ports are compatible with the USB 2.0 and 1.1 specification. The two USB po

Página 66

57Instrument Operating SystemDisk Drive Partitioning and UseAgilent Technologies does not support or warrant correct instrument operation if additiona

Página 67 - Help for that key)

58 Instrument Operating SystemDisk Drive Partitioning and UseBacking-up the factory calibration dataTo back-up the factory calibration data you will n

Página 68

59Instrument Operating SystemDisk Drive Recovery ProcessDisk Drive Recovery ProcessThe Agilent Recovery System can be used to repair errors on the ins

Página 69

6 Contents3rd party software verified by Agilent 42Installation of other 3rd party software 42User Accounts 43User login 43Administrator login 43Agile

Página 70

60 Instrument Operating SystemDisk Drive Recovery ProcessAfter exiting the Agilent Recovery System, the instrument reboots. If the original factory in

Página 71

61Instrument Operating SystemDisk Drive Recovery ProcessNOTE After the final reboot, you might see an error dialog box appear on the XSA applications’

Página 72

62 Instrument Operating SystemDisk Drive Recovery Process4. When prompted to completely remove the selected application and all of its featuresa. Sele

Página 73

63Instrument Operating SystemDisk Drive Recovery ProcessA window appears showing the extraction process.Following the extraction screen, there may be

Página 74

64 Instrument Operating SystemDisk Drive Recovery ProcessIt is possible a newer version of the instrument software is available on the web at http://w

Página 75

65Agilent X-Series Signal AnalyzersGetting Started Guide5 Using Microsoft Windows XPNOTE The capabilities described in this section are Microsoft Wind

Página 76

66 Using Microsoft Windows XPNavigating Windows Without a MouseNavigating Windows Without a MouseKey Presses ActionsEsc Exits/closes a Windows dialog

Página 77

67Using Microsoft Windows XPNavigating Windows Without a MouseUsing the interactive Help system without a mouseTo open the Help window for any menu Ch

Página 78

68 Using Microsoft Windows XPNavigating Windows Without a MouseNavigating Acrobat (PDF) without a mouseWhen you open an Acrobat (PDF) document, it is

Página 79

69Using Microsoft Windows XPNavigating Windows Without a MouseTo select Thumbnails, and Display pages (from Pages Tab): Press the Left/Right/Up/Down k

Página 80

Contents 7Backing-up the factory calibration data 58Disk Drive Recovery Process 59Updating the software 615 Using Microsoft Windows XPNavigating Windo

Página 81

70 Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments RemotelyRemote Desktop: Using the X-Series Instruments RemotelyWindows Rem

Página 82

71Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments Remotely Basic setup for remote desktop operationSetting up a remote deskto

Página 83

72 Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments RemotelySetting up the remote computerThe procedure depends on whether the

Página 84 - The Virtual Front Panel

73Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments RemotelyHow to locate the computer name of the instrumentTo connect a remot

Página 85

74 Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments RemotelyRunning a Remote Desktop sessionInitializing a Remote Desktop sess

Página 86

75Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments RemotelyNOTE Only the current User or an Administrator can remotely log in

Página 87

76 Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments RemotelySetting Remote Desktop optionsStep Notes1. On the Remote Desktop C

Página 88

77Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments Remotely3. Click the Display tab.a. Under Remote desktop size, you may sele

Página 89

78 Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments Remotely4. Click the Local Resources tab.a. Selecting the Disk Drives chec

Página 90 - Step Notes

79Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments RemotelyAccessing menus for Remote Desktop operationUsing the mouse, right-

Página 91 - Selecting the Get Data tab

8 ContentsRead the Warranty 106Service Options 106Packaging the Instrument 106

Página 92 - Selecting the Get Image tab

80 Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments RemotelyKeycode commands for Remote Desktop operationWhen using the instru

Página 93 - Selecting the SCPI Telnet tab

81Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments RemotelyDelDeleteDown Arrow Down ArrowEnterEnter (Return)FileCtrl+Shift+LFR

Página 94

82 Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments RemotelyRight Arrow Right ArrowSaveCtrl+SSelect Space BarSingleCtrl+Alt+SS

Página 95 - Selecting the Help tab

83Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments RemotelyNOTE The effect of pressing Ctrl+Alt+Delete on the keyboard of the

Página 96

84 Using Microsoft Windows XPRemote Desktop: Using the X-Series Instruments RemotelyThe Virtual Front PanelThe Virtual Front Panel is a software equiv

Página 97

85Using Microsoft Windows XPEmbedded Web Server: Using the X-Series Instruments RemotelyEmbedded Web Server: Using the X-Series Instruments RemotelyTh

Página 98

86 Using Microsoft Windows XPEmbedded Web Server: Using the X-Series Instruments Remotely2. Click Advanced information about this Web-Enabled MXEThis

Página 99

87Using Microsoft Windows XPEmbedded Web Server: Using the X-Series Instruments RemotelySelecting the View and Modify Configuration tabStep Notes1. Cl

Página 100 - Windows startup folder

88 Using Microsoft Windows XPEmbedded Web Server: Using the X-Series Instruments Remotely3. Enter new settings as required, then click Apply to cause

Página 101 - 6 Troubleshooting

89Using Microsoft Windows XPEmbedded Web Server: Using the X-Series Instruments RemotelyAccessing the Web Control SA tabSelecting this tab lets you vi

Página 102 - Check the Basics

9Agilent X-Series Signal AnalyzersGetting Started Guide1 Quick-StartThis section explains how to initialize the EMI Receiver.The following topics can

Página 103

90 Using Microsoft Windows XPEmbedded Web Server: Using the X-Series Instruments RemotelyStep Notes

Página 104

91Using Microsoft Windows XPEmbedded Web Server: Using the X-Series Instruments RemotelySelecting the Get Data tabSelecting the Get Data tab allows yo

Página 105 - Calling Agilent Technologies

92 Using Microsoft Windows XPEmbedded Web Server: Using the X-Series Instruments RemotelySelecting the Get Image tabSelecting the Get Image tab captur

Página 106 - Service Options

93Using Microsoft Windows XPEmbedded Web Server: Using the X-Series Instruments RemotelySelecting the SCPI Telnet tabSelecting the SCPI Telnet tab ope

Página 107

94 Using Microsoft Windows XPEmbedded Web Server: Using the X-Series Instruments RemotelySCPI commands and queries may be entered using the command li

Página 108 - Troubleshooting

95Using Microsoft Windows XPEmbedded Web Server: Using the X-Series Instruments RemotelySelecting the Help tabSelecting the Help tab displays basic he

Página 109 - Getting Started Guide 109

96 Using Microsoft Windows XPCapturing/Printing Displays and WindowsCapturing/Printing Displays and WindowsYou need an external keyboard and mouse to

Página 110 - Setting up, 72

97Using Microsoft Windows XPWindows Shortcuts and Miscellaneous TasksWindows Shortcuts and Miscellaneous TasksThis section provides a list of windows

Página 111 - 32-bit, 72

98 Using Microsoft Windows XPWindows Shortcuts and Miscellaneous TasksNavigating an application menu without a mouse or keyboardThis example uses Micr

Página 112 - 112 Getting Started Guide

99Using Microsoft Windows XPWindows Shortcuts and Miscellaneous TasksWindows taskbar auto-hide featureThe Windows taskbar should always be in the auto

Comentários a estes Manuais

Sem comentários