2-10
Operating the Instrument
Exercise 5. Modifying the Setup
Figure 2-2. Signal vertical display
Changing the time base
5
To display the Time Base menu and change the horizontal scale, press:
Time base
,
Scale
Use the knob, arrow keys or keypad to change the scaling so the horizontal
width of the positive portion of the signal is about eight divisions.
Note — Agilent 83480A only
You can automatically adjust the vertical scale, instead of following the preceding pro-
cedure by pressing:
Channel,
Channel autoscale
The amplitude scale is optimized for the input signal while the timebase settings are left
alone.