ECG Reproduction Quality
B-2
8Battery
Rated 7.2V DC 2450mAh
Rechargeable NI-MH Battery Pack
8ECG Reproduction Quality
Meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiograph
Devices. Meets frequency response standard by methods A, D, and E when
configured with 0.15-150 Hz filters.
Accuracy of input signal reproduction meets or exceeds AAMI EC11-1991
Standard for diagnostic electrocardiograph devices
8Patient Safety
8Patient Isolation
Less than 50 µΑ leakage current with 100-240 VAC, at 50/60 Hz, with patient
cable.
8Defibrillation Protection
Protected against damage from 400 joule defibrillator discharges.