2-90
Assembly Replacement
A32 10 MHz Crystal Oscillator (Option UNR)
A32 10 MHz Crystal Oscillator (Option UNR)
Tools Required
T-10 driver
needle-nose pliers
Removal Procedure
Refer to Figure 2-43 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to "Outer Instrument Cover" on page 2-5
and "Inner Instrument Cover" on page 2-7.
3. Position the signal generator with the Rear Panel facing you.
4. Using the needle-nose pliers, disconnect the W47 flexible cable (A31 Motherboard J1041) from
A32J2.
5. Using the needle-nose pliers, disconnect the W48 flexible cable (10 MHz EFC) from A32J1.
6. Disconnect the A32W1 bias cable from the A21 Rear Panel Interface P2.
7. Using the T-10 driver, remove the three screws (1) that attach the A32 10 MHz Crystal Oscillator
to the rear panel.
8. Remove the A32 10 MHz Crystal Oscillator.
Replacement Procedure
Reverse the order of the removal procedure.
Torque all T-10 screws to 9 in-lbs.
Perform the post-repair adjustments and performance tests that pertain to this removal
procedure.