4
High output power
Frequency range 20 GHz models 40 GHz models
Standard (dBm) Option 1EA (dBm) Standard (dBm) Option 1EA (dBm)
250 kHz to 3.2 GHz +13 to –20 +16 to –20 +9 to –20 +15 to –20
>3.2 to 20 GHz +13 to –20 +20 to –20 +9 to –20 +18 to –20
>20 to 40 GHz +9 to –20 +14 to –20
Wide output power range with optional step attenuator
Frequency range 20 GHz models 40 GHz models
Standard (dBm) Option 1EA (dBm) Standard (dBm) Option 1EA (dBm)
250 kHz to 3.2 GHz +11 to –135 +15 to –135 +7 to –135 +14 to –135
>3.2 to 20 GHz +11 to –135 +18 to –135 +7 to –135 +16 to –135
>20 to 40 GHz +7 to –135 +12 to –135
The performance you demand
Typical maximum available power
Frequency (MHz)
Power (dBm)
26
24
22
20
18
16
0 10000 20000 30000 40000
Typical maximum available power
Frequency (MHz)
Power (dBm)
26
24
22
20
18
16
0 5000 10000 15000 20000
Reduce cost of test and simplify test configurations
20 GHz models with option 1EA power
versus frequency
40 GHz models with option 1EA power
versus frequency
CW level accuracy with option 1E1 (dB)
Specifications apply over the 15 to 35 deg. C temperature
range and are degraded typically 0.3 dB outside of that range.
The use of Type-N RF connectors avove 18 GHz degrades
specification typically by 0.2 dB. Level accuracy is not specified
below -110 dBm.
Frequency >+10 dBm +10 to –10 dBm –10 to –70 dBm –70 to –90 dBm –90 to –110 dBm
250 kHz to 2 GHz ±0.6 ±0.6 ±0.7 ±0.8 ±1.4
>2 to 20 GHz ±0.8 ±0.8 ±0.9 ±1.0 ±1.7
>20 to 40 GHz ±1.0 ±0.9 ±1.2 ±2.0
Typical level accuracy +5 dBm option 1E1
Frequency (MHz)
Level error (dB)
1.0
0.5
0.0
-0.5
-1.0
0 5000 10000 15000 20000
Typical level accuracy +5 dBm option 1E1
Frequency (MHz)
Level error (dB)
1.0
0.5
0.0
-0.5
-1.0
0 10000 20000 30000 40000