Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Adjusting Intermodule Skew
11. From the menu bar in the Waveform display, select Options..., then select
Reference trigger..., then select the timing analyzer as the trigger
reference.
12. Select the Group Run icon to update both data sets.
13. Place Markers (see page 106) on a common point and measure the
interval.
14. From the Intermodule window, select Intermodule Skew..., then Adjust
the skew (see page 106) to align displayed waveforms visually.
Configure the Timing Analyzer
The Sampling Tab
Configure the Sampling tab as shown below.
The Format Tab
1. Configure the Format tab with one label called Analyzer.
2. Assign the bit that maps to the input probe signal.