5-116 N9360A Multi UE Tester W-CDMA User Manual
5 Screen Reference
Softkey Menu Field
Softkey Menu 1
Trigger Sing/Cont: Selects the trigger mode from "Sing"(single)
and "Cont"(continuous). This softkey is activated only at
Connection step.
When the trigger mode is changed from "Sing" to "Cont",
continuous measurement starts.
When the trigger mode is changed from "Cont" to "Sing",
softkey disappears and while the measurement is aborted.
Message Field
The following message is displayed in the message field in the
[Measuring] screen:
"Press the desired softkey."
Table 5-34 Measurement items of HSDPA Manual Test
Measurement Item Description
Throughput R Measures the throughput R.
It is calculated by the following equation.
(at the interim reporting)
Throughput (bps) = [TB size] (bit) * [Number of Ack from the beginning to the interim
reporting]/[the measurement period from start to interim reporting] (s)
(at the end of measurement)
Throughput (bps) = [TB size] (bit) * [Number of Ack during measurement]/[the entire
measurement period] (s)
Median CQI (at the interim reporting)
Measures the Median CQI from the beginning to the interim reporting.
(at the end of measurement)
Measures the Median CQI for the entire measurement period.