5-68 N9360A Multi UE Tester GSM User Manual
5 Screen Reference
Sensitivity/ RX Quality/ RX Level
The measured data for BER, FER, RX Quality, and RX Level are
shown as follows.
Any data indicating a failure is highlighted.
Table 5-21 [Manual Test] GSM Mode Phase Error/Frequency Error Screen Field
Screen Field Description
RMS Measured Phase Error (RMS) and the HI Limit are shown. The HI Limit is set on the [Configuration:
Test Condition] screen.
Phase Peak Measured Phase Error (Peak), the HI and the LO Limit are shown. A set of the HI and the LO Limit is
set on the [Configuration: Test Condition] screen.
Marker Measured Phase Error at the marker position, the HI and the LO Limit are shown. Measured data is
shown when the marker is placed on the measurement trace. The HI Limit is set on the
[Configuration: Test Condition] screen.
Frequency Measured Frequency Error, the HI and the LO Limit are shown. The HI and the LO Limit are set on the
[Configuration: Test Condition] screen.
Figure 5-40 [Measuring] Sensitivity/ RX Quality/ RX Level Screen