Agilent Technologies E6474A Manual de Serviço Página 47

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 46
47
A
A
Appendix A- Calibration of USB Audio
Module
Measurement may normally be performed at the
recommended settings for the ASIO USB control application,
refer to “Setting up the USB Audio Module” on page 26.
However, attenuation or boosting of voice amplitude may
occur along the communication network (which includes the
mobile handset), such that undesirable low SNR
(signal- to- noise- ratio) or clipping occurs. Hence, it may be
necessary to adjust the gain setting for downlink (ear gain)
or uplink (month gain). Use the ASIO USB Control
application to set the levels.
1 To verify if calibration is necessary, start the
measurement and let it run for about 3 minutes before
terminating measurement (may also setup the system to
run 2 iterations).
2 Start the VQT application.
Vista de página 46
1 2 ... 42 43 44 45 46 47 48 49 50 51 52

Comentários a estes Manuais

Sem comentários