Decibel (dB) for U1272A
[1] - [3]
dB base Reference Default reference
1 mW (dBm) 1 to 9999 Ω 50 Ω
1 V(dBV) 1 V 1 V
1. The reading of dBm is indicated in decibels of power above or below 1 mW, or decibels of voltage above or below 1 V. The
formula is calculated according to the voltage measurement and specified reference impedance. Its accuracy is
depended on the accuracy of the voltage measurement. See Decibel (dBV) accuracy table below.
2. Auto-ranging mode is used.
3. The bandwidth is according to voltage measurement.
Decibel (dBV) accuracy
Range
dBV range Accuracy
Minimum Maximum 45 to 65 Hz
20 Hz to
1 kHz
1 Hz to
5 kHz
5 kHz to
20 kHz
20 Hz to
100 kHz
30 mV -56.48 -30.46 0.06 0.07 0.09 0.1 0.32
300 mV -36.48 -10.46 0.06 0.07 0.09 0.1 0.32
3 V -16.48 +9.54 0.06 0.09 0.14 0.19 0.32
30 V +3.52 +29.54 0.06 0.09 0.14 0.19 0.32
300 V +23.52 +49.54 0.06 0.09 0.14 0.19 -
1000 V +33.98 +60 0.06 0.09 0.14 - -
Measurement rate (approximate)
Function
Times/second
U1271A U1272A
ACV 7 7
DCV 7 7
Ω1414
Ω with offset compensation - 3
Diode 14 14
Auto Diode - 3
Capacitance 4 (<100 µF) 4 (<100 µF)
DCI 7 7
ACI 7 7
Temperature 7 7
Frequency 2 (>10 Hz) 2 (>10 Hz)
Duty cycle 1 (>10 Hz) 1 (>10 Hz)
Pulse width 1 (>10 Hz) 1 (>10 Hz)