17
Jitter Tolerance Test Specifi cations
Figure 23. Interference channel connectors
Figure 24. Typical S21 parameter for ISI channel of 9 inch length
Table 17. Typical ISI (measured in UI) for traces depending on data rate, pattern and trace length.
Data rate 1.25 Gb/s 2.5 Gb/s 3.125 Gb/s 5 Gb/s 6.25 Gb/s 11 Gb/s
ISI trace
length
PRBS
2
7
-1
PRBS
2
15
-1
CJPAT PRBS
2
7
-1
PRBS
2
15
-1
CJPAT PRBS
2
7
-1
PRBS
2
15
-1
CJPAT PRBS
2
7
-1
PRBS
2
15
-1
CJPAT PRBS
2
7
-1
PRBS
2
15
-1
CJPAT PRBS
2
7
-1
PRBS
2
15
-1
CJPAT
3.5 inches 0.007 0.017 0.016 0.016 0.022 0.014 0.020 0.031 0.029 0.037 0.062 0.038 0.054 0.085 0.056 0.099 0.146 0.131
9 inches 0.026 0.034 0.037 0.039 0.066 0.039 0.057 0.080 0.079 0.104 0.157 0.092 0.147 0.216 0.138 0.329 0.504 0.405
16 inches 0.045 0.068 0.051 0.103 0.138 0.106 0.137 0.191 0.117 0.279 0.365 0.260 0.346 0.543 0.376
20 inches 0.058 0.088 0.062 0.152 0.184 0.123 0.181 0.276 0.167 0.366 0.529 0.349 0.580
24 inches 0.081 0.109 0.103 0.182 0.260 0.171 0.254 0.356 0.274 0.529
28 inches 0.098 0.141 0.147 0.241 0.334 0.238 0.319 0.483 0.373
32 inches 0.128 0.153 0.120 0.289 0.395 0.295 0.389 0.570 0.376
36 inches 0.156 0.188 0.168 0.375 0.493 0.390 0.507
40 inches 0.172 0.228 0.199 0.458 0.626 0.423 0.617
44 inches 0.199 0.262 0.247 0.571
Sinusoidal interference (Option J20)
This adds common mode, differential or
single-ended sinewave signal on top of the
data outputs, to test common mode rejection
of a receiver and to emulate vertical eye
closure. Sinusoidal interference is injected
before the signal passes through the ISI
board traces (“near end”) when using P1 as input. For “far end”
injection P2 has to be used as input.
Table 18. Specifi cations for sinusoidal interference (SI)
Amplitude
1
0 to 400 mV common mode, single
ended and differential (differential
amplitude 0 to 800 mV)
Frequency 10 MHz to 3.2 GHz in 100 kHz steps
Level accuracy ± 10% ±10 mV typical
1. The output signal amplitude is reduced by 3 dB when sinusoidal interference
is enabled.
S21
Interference channel (Option J20)
The Option J20 is only available in addition to Option J10. It
includes the semi-rigid cable set to connect data outputs to P1
and P1 (N4915A-008).
Interference channel input and output (P1, P2)
User selectable board traces are switched into
the signal path to emulate a backplane.
Table 16. Specifi cations for inter symbol interference (ISI)
Trace length 3.5” (minimum), 9” (minimum with S.I.
enabled) 16”, 20”, 24”, 28”, 32”, 36”, 40”,
44” inches of board trace type Nelco 4000-6.
When using in combination with sinusoidal
interference, minimum trace length is 9 inches
S
21
parameter See Figure 25
Range See Table 17
Max input levels -5.5 V to +5.5 V
Connectors 2.4 mm, female
Electro Rent Corporation
an Agilent Authorized Technology Partner
www.ElectroRent.com/BuyNew