42 N2780/81/82/83 User’s and Service Guide
Test Coil Construction for the N2782A/B and N2783A/B
Construct a 5- turn coil with a diameter of about 4 inches
(10 cm) using transformer wire.
Amplitude Accuracy Test
The following procedure measures the accuracy of the
current probe.
1 Connect the equipment as shown below. Allow all
instruments 30 minutes for warm up before starting the
5 turns
medium_standard.book Page 42 Friday, September 10, 2010 2:45 PM