13
DC power supply specifications (continued)
Performance specifications (continued)
Table 8. DC power supply performance specifications
Parameter
Specifications
S1S2 S1 S1m S2 S2m
Output ratings AUTO 30 V/1 A 100 mA/30 V 8 V/3 A 1000 mV/3 A
Front terminal load regulation
1
±(% of output + offset)
Voltage < 3 mV +
(6 mV/A)
< 3 mV +
(6 mV/A)
< 3 mV +
(6 mV/A)
< 3 mV +
(6 mV/A)
< 0.3 mV +
(6 mV/A)
Current < 0.03% +
0.3 mA
< 0.03% +
0.3 mA
< 0.03% +
0.03 mA
< 0.03% +
0.3 mA
< 0.03% +
0.3 mA
Rear terminal load regulation
±(% of output + offset)
Voltage < 0.01% +
3 mV
< 0.01% +
3 mV
< 0.01% +
3 mV
< 0.01% +
3 mV
< 0.01% +
0.3 mV
Current < 0.03% +
0.3 mA
< 0.03% +
0.3 mA
< 0.03% +
0.03 mA
< 0.03% +
0.3 mA
< 0.03% +
0.3 mA
Line regulation
Voltage 3 mV typical 3 mV typical 3 mV typical 3 mV typical 0.3 mV typical
Current 1.5 mA
typical
1.5 mA
typical
0.15 mA
typical
1.5 mA
typical
1.5 mA
typical
Programming resolution
Voltage 1 mV 1 mV 1 mV 1 mV 0.1 mV
Current 0.1 mA 0.1 mA 0.01 mA 0.1 mA 0.1 mA
Readback resolution
Voltage 1 mV 1 mV 1 mV 1 mV 0.1 mV
Current 0.1 mA 0.1 mA 0.01 mA 0.1 mA 0.1 mA
Front panel resolution
Voltage 1 mV 1 mV 1 mV 1 mV 0.1 mV
Current 0.1 mA 0.1 mA 0.01 mA 0.1 mA 0.1 mA
Transient response time
Less than 100 µs for output to recover to within 15 mV following a change in
output current from full load to half load or vice versa
Command processing time
Average time for output voltage to begin to change after receipt of digital data
when instrument is connected directly to the USB or GPIB is less than 100 ms
Over-voltage protection
(for CC mode)
Accuracy: 0.5% + 0.5 V
Activation time
2
: < 2 ms
Over-current protection
(for CV mode)
Accuracy: 0.5% + 0.05 A
Activation time
2
: < 2 ms
1. The terminal sense is related to the resistance of the contacts or leads, and proportional to the load condition.
2. Average time for the detection of OVP or OCP condition. The output will be dropped down and set to standby within 20 ms.