Vista de página 11
12
General characteristics
Power supply • Universal 100 V
ac
to 240 V
ac
±10%
• AC line frequency 45 Hz to 66 Hz (360 Hz to 440 Hz for 100/120 V operation)
Power consumption 150 VA maximum
Current Input Fuse 3.15 A, 500 V FF fuse (on front panel)
Display Highly visible vacuum-fluorescent display (VFD)
Operating environment • Operating temperature from 0 °C to +55 °C
• Relative humidity up to 80% at 30 °C RH (non-condensing)
• Altitude up to 2000 meters
• Pollution degree 2
• For indoor use only
Storage compliance –40 °C to 70 °C
Safety compliance Certified with:
• IEC 61010-1:2001/EN61010-1:2001 (2nd Edition)
• Canada: CAN/CSA-C22.2 No. 61010-1-04
• USA: ANSI/UL 61010-1:2004
EMC compliance Certified with:
• IEC61326-1:2005 / EN61326-1:2006
• CISPR 11:2003 / EN55011:2007 (Group 1 Class A)
• Canada: ICES/NMB-001:2004
• Australia/New Zealand: AS/NZS CISPR11:2004
Shock and vibration Tested to IEC/EN 60068-2
Remote interface • GPIB IEEE-488
• Full Speed USB 2.0 (Type B)
• USBTMC 488.2 Class device
• USB-CDC
Programming language Standard Commands for Programmable Instruments (SCPI)
Measurement category • CAT II, 300 V
• CAT I, 1000 V
dc
, 750 V
ac
rms
• 2500 V
pk
transient over voltages
Dimensions (H × W × D) 105 × 255 × 329 mm (with rubber bumpers)
87 x 215 x 312 mm (without rubber bumpers)
Weight 3.775 kg (with rubber bumpers)
3.535 kg (without rubber bumpers)
Warranty Three years
Calibration cycle One year
Warm-up time 60 minutes