Operation and Features 2
U2741A User’s Guide 37
Testing Continuity
The continuity testing function has the following features:
Uses a 1 mA ± 0.2% constant current source.
Open circuit voltage is limited to less than 4.5 V for all
ranges.
Continuity threshold is fixed at 10 Ω.
Response time is 60 samples/second.
Make the connection as shown below.
Agilent Measurement Manager operation
Select the Cont-
)
)
) function. The reading is displayed and
updated continuously.
SCPI commands
The following example shows how to make a Continuity Test
measurement using SCPI commands.
MEASure:CONTinuity?
Open or closed circuit
Current
Te s t
User's Guide.book Page 37 Friday, August 8, 2008 9:11 AM