Getting Started Tutorial 1
34450A User’s Guide 17
Measuring AC (RMS) or DC current up to 10 A
Measurement Range (AC): 1.00000 A, 10.0000 A
Measurement Range (DC): 1.00000 A, 10.0000 A
Speed (AC): Slow-2 Hz, Medium-20 Hz, Fast-200 Hz
Speed (DC): Slow, Medium, Fast
•Default Setting: Autoranging, Slow measurement speed
Shunt Resistance: 0.01 Ω for 1 A and 10 A range
Input Protection: Internal 11 A, 1000 V fuse for the 10 A terminal
Figure 1-10 ACI rms or DCI (A) terminal connection and display
34450A UG.book Page 17 Tuesday, July 23, 2013 2:43 PM