Installation Note 83650-90007 13
Figure 13. RF Deck in the Service Position
46. Connect the new cable W51 to the AT3 attenuator on the RF deck. Position the cable so when the
RF deck is installed it will connect to FL2 without any stress on the cable. Make sure the connection
to AT3 is tight before installing the RF deck.
47. Replace the deck in the synthesizer. Connect W51 to FL2. Be careful that the cable connection to
AT3 remains tight and the cable is not stressed or bent to make the connection. If necessary, remove
the RF deck, loosen the W51 connection to AT3 and reposition the cable as required.
NOTE If the RF deck does not fit easily back into the casting, loosen the RF deck bracket screws
on the side of the synthesizer, slide the RF deck into place, and tighten the screws again.