Device under test (DUT)
Your DUT contributes to the overall noise fi gure uncertainty based on its
individual noise fi gure, gain, port match, and noise parameters. In general,
there are two scenarios to consider when choosing the Y-Factor method. When
the output noise of the DUT is well above the input noise of the analyzer, the
analyzer with the best instrument uncertainty gives the most accurate results,
and the MXA signal analyzer is the best option. If the output noise of the DUT is
smaller, select the NFA noise fi gure analyzer, which gives the lowest uncertainty.
Refer to Table 2 for nominal specifi cation comparison of these solutions at 1 GHz.
The following graph shows how DUT gain affects noise fi gure uncertainty when
using a spectrum analyzer or noise fi gure analyzer. This example is at 1 GHz with
a 346A noise source and assumes the DUT has a 2 dB noise fi gure and 1.5:1 VSWR.
Figure 3. As the gain of a DUT decreases, Y-factor noise fi gure measurement uncertainty increases;
below 10 dB of gain, there are signifi cant differences between the various instrument choices
The values in the graph above were created via the noise fi gure uncertainty
calculator and nominal specifi cations at 1 GHz shown in Table 2. The uncertainty
calculator can be found at www.agilent.com/fi nd/nfu. The uncertainty calculator
can be used for either of the following cases:
Modeling the performance of your system: For this purpose, defaults are available
for Agilent's noise fi gure instruments and noise sources. These defaults have typical
values associated with them and can be useful for estimating the effect individual
parameters have in overall uncertainty levels.
Making actual calculations of the uncertainty of your system: You will need to
obtain accurate values of all the associated parameters in question, such as match
and gain. Please consult the calibration certifi cates of your instruments to obtain the
measured uncertainty parameters of the equipment being used.
Find out more www.agilent.com/fi nd/nf