55
PERFORMANCE TEST CARD - 1 YEAR LIMITS
Test Performed By
Keysight Model 3458A Digital Multimeter
Serial Number
Date
DC CURRENT TESTS
CAL? 60
TEMP?
Difference - (must be less than 5 degrees C)
Perform an ACAL OHMS
Test # 3458A
Input
3458A
Range
Transfer Standard
Reading
Unit Under Test
Reading
Difference Limit (Std) Limit (Opt 002) Pass Fail
OFFSET TESTS
(NOTE: Math Null is Disabled)
1Open
100 µA
N/A N/A
000.00095
µA 000.00095 µA
2 Open 1 mA N/A N/A 0.0000065 mA 0.0000065 mA
3Open
10 mA
N/A N/A 00.000065 mA 00.000065 mA
4 Open 100 mA N/A N/A 000.00065 mA 000.00065 mA
5
Open 1 A
N/A N/A 0.0000115 A 0.0000115 A
GAIN TEST
1
100
µA 100 µA 000.00356 µA 000.00356 µA
2 1 mA 1 mA 0.0000323 mA 0.0000323 mA
3 10 mA 10 mA 00.000323 mA 00.000323 mA
4 100 mA 100 mA 000.00489 mA 000.00489 mA
5 1 A 1 A 0.0001349 A 0.0001349 A