Error Messages 7
N9340A User’s Guide 147
130 Meas uncal Indicates the measurement is
uncalibrated due to fast sweeping
through a narrow RBW filter.
Check the sweep time, span, and
bandwidth settings, or use auto
coupling.
119 RF Board Changed Indicates the RF board was
changed, the analyzer needs
re-load the alignment data.
116 Cannot
communicate with
RF
Indicates the MCU cannot find
the RF board.
Error Code Error Message Description