Chapter 3: Testing Performance
Offset Accuracy Test
22
b When the Acquisition Setup window is displayed, enable averaging and set the # of
averages to 256 as shown below.
3
Configure the scope to measure Average voltage on channel 1 as follows:
a Change the vertical sensitivity of channel 1 to 5 mV/div.
b Click the V avg measurement on the left side of the screen.
V avg
measurement