User’s and Service Guide Supplement 15
E8362B, E8363B and E8384B Option H85 General Overview
Making High Power Measurements With Option H85
Final Setup and Response Calibration
23. Verify that all of the power and attenuator settings are correct, and set the following
measurement. Press [Menu/Dialog] then tab to Trace. In the pull down menu select
Measure then S21.
24. Connect the test port cables of the analyzer to form a thru configuration.
25. Press [Menu/Dialog] then tab to Calibration. In the pull down menu select Calibration
Wizard then Unguided Calibration Use Mechanical Standards. Select THRU
Response. Follow analyzers window prompts to finish calibration.
26. Make the connection as shown in Figure 9. Turn on the DUT and measure the S21 gain
of the amplifier under test to confirm the proper operation of the measurement test
setup.
Figure 9 Forward High Power Measurement Configuration
27. Continue with any other high power measurements.