USER’S GUIDE
Multi-Cell Charger/Discharger
Agilent Model E4370A
Powerbus Load
Agilent Model E4371A
64-Channel Charger/Discharger
Agilent Models E4374A and E4375A
Agilent Part No. 5964-8138
Microfiche No. 5964-8139
September 2001
Vista de página 0
1 2 3 4 5 6 ... 140 141

Resumo do Conteúdo

Página 1 - Multi-Cell Charger/Discharger

USER’S GUIDEMulti-Cell Charger/DischargerAgilent Model E4370APowerbus LoadAgilent Model E4371A64-Channel Charger/DischargerAgilent Models E4374A and E

Página 2 - Warranty Information

1 - General Information10Basic Functions♦ Charger – The Agilent MCCD can deliver accurately controlled current and voltage into a cell forproper formi

Página 3 - Safety Summary

6 - Language Dictionary100cfSetShutdownDelaySyntaxint cfSetShutdownDelay(CF_HANDLE server, float delay);DescriptionSets the delay between the assertio

Página 4 - Document Scope

Language Dictionary - 6101cfSetVoltageCAUTION Direct output control should not be used for charging cells. There is no protectionagainst overcharging

Página 5 - Table of Contents

6 - Language Dictionary102cfStateListSyntaxint cfStateList(CF_HANDLE server, char *buffer);DescriptionReturns a comma-separated and null terminated li

Página 6 - 6 - LANGUAGE DICTIONARY 67

Language Dictionary - 6103cfWriteSerialSyntaxint cfWriteSerial(CF_HANDLE server, CF_SERIAL_PORT port, char*port_data, int count);DescriptionWrites cou

Página 8

1057C Program ExamplesExample 1This following C program shows you how to implement the example discussed in the beginning ofchapter 5 using the API ce

Página 9 - General Information

7 - C Program Examples106} while(presentState != CF_INITIATED);/* Start the sequence */cfTrigger(server);/* Wait for the sequence to end */do {cfGetRu

Página 10 - Hardware Description

C Program Examples - 7107Example 2This following C program shows you how to implement the example discussed at the end of chapter 1using the API cell

Página 11 - General Information - 1

7 - C Program Examples108/***********************************************************************Main function****************************************

Página 12 - Agilent E4371A Powerbus Load

C Program Examples - 7109/* Configure resistance limit for output probe test. */cfSetOutputProbeTest(hServer, 0.1f);/* Mark outputs 65 - 256 as unused

Página 13 - POWERBUS LOAD

General Information - 1111FaultReadySYSTEMPowerReadyActiveInternalExternalFAULTOffOnLINEE4370AMULTICELL CHARGER/DISCHARGER2FaultReady3FaultReady4Fault

Página 14 - External Power Source

7 - C Program Examples110{cfGetDigitalPort(hServer, &nDigitalPort);if (nDigitalPort & DIG_START_BUTTON)break;/* Sleep for 1 second. */Sleep(10

Página 15

C Program Examples - 7111/* Measure the internal resistance of all cells. */cfMeasACResistance(hServer, CF_ALL_CELLS, fCellResistance);/** Turn off th

Página 16 - Measurement Capability

7 - C Program Examples112Example 3You can control up to 16 Agilent MCCDs from one PC and still achieve good system responsiveness,depending on the app

Página 17 - Capacity Measurements

C Program Examples - 7113// Signal threads to start reading logs.SetEvent(ThreadInfo[0].hStart);SetEvent(ThreadInfo[1].hStart);// Wait for threads to

Página 18 - Data Logging

7 - C Program Examples114/*****************************************************************************Return a timestamp string.*********************

Página 19 - Protection Features

115ASpecificationsHardware SpecificationsSpecifications in Table A-1 are warranted. Specifications apply over an ambient temperature range of 0°C to 4

Página 20 - If AC Power Fails

A - Specifications116Tables A-2 through A-4 list the supplemental characteristics of the Agilent MCCD System.Requirements for the external power bus s

Página 21 - Remote Programming Interface

Specifications - A117Table A-2. Agilent E4370A/E4374A/E4375A MCCD Characteristics (continued)Parameter Condition ValueAuxiliary bias output voltage a

Página 23 - Installation

119BCalibrationCalibration TypesThere are three types of calibration available for the Agilent MCCD System.♦ Full calibration, which calibrates the Ag

Página 24 - Company Address Contact

1 - General Information12RS-232PORT ARS-232PORT BABCDEF+ and - Power bus connectors(- bus bar is connected to chassis ground)Calibration status LEDsCo

Página 25 - Channel Connections

A - Specifications120Transfer CalibrationNOTE: Transfer calibration does not require an external voltmeter. It can be performedindependently of the fu

Página 26

Specifications - A121Figure B-1. Calibration ConnectionsRS-232PORT A3458AMultimeterH ILOPowerOnOffMENUNUMERIC/USERFUNCTION/RANGEIInputSense4 3 2 1G

Página 27 - Remote Sense Connections

A - Specifications122Accessing CalibrationCalibration control is accessible by one of three methods:♦ the Agilent MCCD Configuration Screens,♦ API cal

Página 28 - Power Bus Connections

Specifications - A123Rear panel transfer calibration switchThis push button switch is accessible through a recessed hole on the rear panel. When press

Página 30

125CDimension DrawingsFigure C-1 shows a simplified outline diagram of the Agilent E4370A MCCD mainframe. Figure C-2shows a simplified outline diagram

Página 31

C - Dimension Drawings126HOT EXHAUST AIRSIDE VIEW425.5 mm540.5mmLINEOnOffE4371APOWERBUS LOADFRONT VIEWAIR FLOWTOP VIEW28.0mm68.0mm221.5mmFigure C-2. A

Página 32 - Digital Connections

127DSense and Power Connector PinoutsThe figures and tables in this appendix document the sense and power pinout assignments on the front ofthe Agilen

Página 33 - Wiring Guidelines

D - Sense and Power Connector Pinouts128Note: Unlabeled pins are the minus connections of each pair.Figure D-1. Card 1 Sense and Power Connector Cell

Página 34 - RS-232 Connections

Sense and Power Connector Pinouts - D129Note: Unlabeled pins are the minus connections of each pair.Figure D-3. Card 3 Sense and Power Connector Cell

Página 35 - Auxiliary Output Connection

General Information - 113The Agilent E4371A Powerbus Load has a + and a − power bus connector on its rear panel. There is alsoa ground connection. To

Página 36 - Visual C++ Configuration

D - Sense and Power Connector Pinouts130Table D-1. Card 1 Sense and Power Pinout AssignmentsSense Pins Cell Number Power Pins Sense Pins Cell Number P

Página 37 - Configuration

Sense and Power Connector Pinouts - D131Table D-2. Card 2 Sense and Power Pinout AssignmentsSense Pins Cell Number Power Pins Sense Pins Cell Number P

Página 38 - Configure

D - Sense and Power Connector Pinouts132Table D-3. Card 3 Sense and Power Pinout AssignmentsSense Pins Cell Number Power Pins Sense Pins Cell Number P

Página 39 - Network Configuration

Sense and Power Connector Pinouts - D133Table D-4. Card 4 Sense and Power Pinout AssignmentsSense Pins Cell Number Power Pins Sense Pins Cell Number P

Página 41 - Configuring the Digital I/O

135EIn Case of TroubleIntroductionThe Agilent E4370A MCCD System has a built-in self test capability which is performed at power-on.Additionally a mor

Página 42

E - In Case of Trouble136Agilent E4374A/E4375A Fault1, 2, 3, 4Indicates an internal hardware fault such asSelftest failure.Calibration error.Hardware

Página 43

137Index—A—abort, 70airflowmainframe, 25powerbus load, 25amp-hour capacity measurement, 17API, 21API functionsguidelines, 67summary, 68API libraryinst

Página 44

Index138cfSetOutputProbeTest, 94cfSetOutputState, 94cfSetSense, 95cfSetSenseProbeTest, 95cfSetSeqStep, 95cfSetSeqTest, 97cfSetSeqTestMult, 99cfSetSeri

Página 45 - Agilent MCCD User Interface

Index139sequence step time, 78sequence test parameters, 77, 78serial port configuration, 78serial port status, 78step number, 79trigger source, 79user

Página 46 - Using the Interface

1 - General Information14External Power SourceFor the charging cycle, each Agilent MCCD mainframe requires an external dc power source to powerthe cel

Página 47

Index140serial port data, 86test log data, 87rear panelmainframe, 12remote sensing, 27repacking, 23reset, 58, 87reset sequence, 87resistance measureme

Página 48

141Agilent Sales and Support OfficeFor more information about Agilent Technologies test and measurement products, applications, services,and for a cu

Página 49 - Programming Overview

General Information - 115Agilent E4370A(256 channels)(256 channels)(256 channels)(256 channels)(256 channels) +4 E4374A cards(256 channels)(256 chann

Página 50 - Cell Forming Example

1 - General Information16Agilent E4370A(256 channels)(256 channels)(256 channels)(256 channels)(256 channels) +4 E4375A cards(256 channels)(256 chann

Página 51

General Information - 117sensing over local sensing is that when the remote sense leads are connected to the cell, the actual voltageof the cell will

Página 52 - 5 - Programming Overview

1 - General Information18Cell ResistanceIn addition to continuous voltage, current, and capacity measurements, the Agilent MCCD can alsomeasure ac and

Página 53 - Function Call Overview

General Information - 119The following events can be used to trigger critical measurements:Change in voltage(∆∆∆∆V)If the trigger is ∆∆∆∆V, a data log

Página 54 - Step/Test Functions

2Warranty InformationCERTIFICATIONAgilent Technologies certifies that this product met its published specifications at time of shipment from the facto

Página 55 - Sequence Control

1 - General Information20Internal Protection FunctionsThere are internal relays between the power bus and the Agilent E4374A/E4375A Charger/Discharger

Página 56 - Output Configuration

General Information - 121When power fails, the power bus is also disconnected from the Agilent MCCD because of the biaspowered relays inside the Agile

Página 57 - Instrument Protection

1 - General Information22Figure 1-7. Typical Cell Forming Station♦The control PC sends a signal via the LAN to the digital I/O to turn on the Ready li

Página 58 - Instrument State Storage

232InstallationInspectionWhen you receive your equipment, inspect it for any obvious damage that may have occurred duringshipment. If there is damage,

Página 59

2 - Installation24Table 2-2. Accessories (continued)Item Manufacturer’s PartNumberDescriptionDocumentation Package Agilent E4373A Contains user docume

Página 60 - Measurement Log

Installation - 225LocationAgilent E4370A MCCD MainframeThe outline diagrams in Appendix C give the dimensions of your Agilent MCCD mainframe. Themainf

Página 61 - Output Measurements

2 - Installation26Table 2-4. Channel ConfigurationCard Connector Number Number 1234567 811 - 8 9 - 16 17 - 24 25 - 32 33 - 40 41 - 48 49 - 56 57 - 64

Página 62 - General Server functions

Installation - 227♦ It is good engineering practice to either twist or shield the sense and power wires.♦ Twist the power wires together and keep them

Página 63 - Selftest

2 - Installation28Power Bus ConnectionsCAUTION: Observe polarity when making the power bus connections to both the Agilent MCCDmainframe and the Agile

Página 64 - Digital port

Installation - 229Table 2-6. Ampacity and Resistance of Stranded Copper ConductorsAWG No. Areain mm2Ampacity Resistancein ΩΩΩΩ/meterResistancein ΩΩΩΩ/

Página 65 - Probe check

3Safety SummaryThe following general safety precautions must be observed during all phases of operation of this instrument.Failure to comply with thes

Página 66

2 - Installation30(256 channels)(256 channels)Power Source(24 V @ 160 A)Powerbus LoadPower Source(24 V @ 160 A)+_+_+_+_+_maximumdischargingcurrent = 1

Página 67 - Language Dictionary

Installation - 2311. Multiply the power used by one cell times the number of cells in the Agilent MCCD. Divide theresult by the efficiency of the unit

Página 68 - API Function Summary

2 - Installation32Digital ConnectionsEach Agilent E4370A MCCD mainframe has a 16-bit digital I/O port. Digital I/O configuration can bedone with the A

Página 69 - Language Dictionary - 6

Installation - 233ExternalTriggerThis external trigger input is used to start the cell forming sequence.Power FailDepending on how the system is confi

Página 70 - API Function Definitions

2 - Installation34The following figure illustrates some typical DIO hardware connections.A) Relay Driver Example Circuit. . . .+ 16.5 V maximum. .

Página 71

Installation - 2351 2 3 4 56 7 8 9DB-9 male connectorPin123456789Input/OutputInputOutputCommonOutputInputDescriptionno connectionRec

Página 72

2 - Installation36Installing the API Library and Measurement Log UtilitySoftware for the Agilent MCCD consists of the API library and a measurement ut

Página 73

373ConfigurationConfiguring the LANThe connection to the LAN is through a standard 8-pin 10Base-T connector on the rear panel, whichmust first be conf

Página 74

3 - Configuration38In the COMProperties box:select the following port settings:Bits per second 9600Data bits 8Parity NoneStop Bits 1Flow control NoneT

Página 75 - *config);

Configuration - 339For now, you will only be accessing the Network Configuration and the Identification Configurationscreens.Network ConfigurationNO

Página 76

4Safety SymbolsSAFETY SYMBOLSDirect current Caution, risk of electric shockEarth (ground) terminal Caution, hot surfaceProtective earth (ground) termi

Página 77

3 - Configuration40The Agilent MCCD is shipped from the factory without a password being set. A network Password canbe assigned to the Agilent E4370A

Página 78

Configuration - 341Miscellaneous ConfigurationIn the Initial Screen, select 5 to configure the language used in the Agilent MCCD User Interface. You

Página 79

3 - Configuration42To continue configuring the Digital I/O, press 3. The pin numbers of the Digital I/O connector appear onthe screen. Refer to Figure

Página 80

Configuration - 343All pins do not have to be configured in the same way. Some can be used as isolated outputs while othersare single-ended I/O. Fun

Página 81

3 - Configuration44Mixed Configuration ExampleThe following example illustrates a mixed digital I/O configuration. In this example,♦ Pins 0, 2, 4, an

Página 82

454Agilent MCCD User InterfaceDescriptionThe Agilent MCCD User Interface lets you interactively monitor and control the Agilent MCCD System.This inter

Página 83

4 - User Interface46LocalizationThe user interface pages are provided in English and Japanese. You can specify the default languageduring installation

Página 84

User Interface - 447Using the Agilent MCCD Measurement Log UtilityIf you are using the Agilent MCCD User Interface to create and run a cell forming se

Página 85

4 - User Interface48Raw logTransfers all of the logged data in the order that it was logged.Sorted by cellTransfers all of the logged data sorted by c

Página 86

495Programming OverviewA Cell Forming OverviewThe cell forming process of the Agilent E4370A MCCD consists of a series of steps or actions that areper

Página 87

5Table of ContentsWarranty Information 2Safety Summary 3Document Scope 4Table of Contents 51 - GENERAL INFORMATION 9Agilent MCCD System Capabilities 9

Página 88

5 - Programming Overview50NEXTFAILGo to NEXT stepRemove cell fromsequenceTest action?Test outcome?Start STEPContinue in stepFALSETRUEFigure 5-1. Test

Página 89

Programming Overview - 551Step 1In Step 1, all cells are set to charge at a constant current of 0.295 amperes until the voltage reaches 4.2volts. It c

Página 90

5 - Programming Overview52VOLTAGETIME(35)(30)(25)(20)(15)(5)(10)(40)(1)(2)(3)(4)(volts)(minutes)STEP 1ENDRestSTEP 2ENDDischargeRestChargeSTEP 3ENDSTEP

Página 91

Programming Overview - 553Function Call OverviewThe driver function calls that control the cell forming process of the Agilent E4370A MCCD areclassifi

Página 92

5 - Programming Overview54Grouping FunctionsThe group handle returned by cfOpenGroup can be used with any of the functions in the list below.These fun

Página 93 - CF_MAX_GROUP_NAME_LEN

Programming Overview - 555To program one test to cause a cell to fail if the voltage does not exceed 4 volts within 30 minutes, andanother test to cau

Página 94

5 - Programming Overview56After selftest is completed and there is dc voltage on the power bus, the instrument moves to theCF_IDLE state. In this stat

Página 95

Programming Overview - 557Instrument ProtectionThe following diagram shows the various protection states of the instrument.CF_HW_FAILEDFailed selftest

Página 96

5 - Programming Overview58Power Fail OperationThe Agilent E4370A MCCD can operate in one of two power-fail shutdown modes. The mode is set bythe cfSet

Página 97

Programming Overview - 559The power-on and cfReset instrument settings are:Output State = OFFOutput Voltage = 0 voltsOutput Current = 0 amperesGroups

Página 98

63 - CONFIGURATION 37Configuring the LAN 371. Configure the HyperTerminal program 372. Connect the Agilent E4370A MCCD to the COM port on the PC 383.

Página 99

5 - Programming Overview60Measurement LogThe Agilent E4370A MCCD logs measurement data at the beginning, end, and can be programmed to logmeasurement

Página 100 - Description

Programming Overview - 561NOTE: The measurement log contents are cleared when a sequence is initiated.Set and query the Measurement logging voltage an

Página 101

5 - Programming Overview62The output voltage is regulated and measured at the power output terminals unless the remote sensefunction is used. The Agil

Página 102

Programming Overview - 563SelftestThe Agilent E4370A MCCD has a built in selftest capability, which is performed at power-on. Thislimited selftest ver

Página 103 - See Also

5 - Programming Overview64Since calibration can take up to 15 minutes for an Agilent MCCD with 256 channels, calibrationfunctions do not wait for cali

Página 104

Programming Overview - 565Probe checkProbe check includes three separate functions: a continuity check, a power probe resistance check, and asense pro

Página 106 - 7 - C Program Examples

676Language DictionaryAPI Usage GuidelinesThis Application Programming Interface lets you create an application program on a PC to control theoperatio

Página 107 - Example 2

6 - Language Dictionary68Password ProtectionAn application program must provide a password to open a connection to a server. As shipped from thefactor

Página 108

Language Dictionary - 669cfMeasDCResistancemeasures the dc resistance of a cell or all cellscfMeasOutputProbeResistancemeasures the output probe resis

Página 109 - C Program Examples - 7

7cfCalTransfer 71cfClose 71cfDeleteGroup 71cfGetCellStatus 72cfGetCellStatusString 72cfGetCurrent 72cfGetDigitalConfig 73cfGetDigitalPort 73cfGetGroup

Página 110

6 - Language Dictionary70API Function DefinitionscfAbortSyntaxint cfAbort(CF_HANDLE server);DescriptionAborts a forming sequence, which sets the run s

Página 111

Language Dictionary - 671cfCalTransferCAUTION: Make sure that no cells are connected when executing cfCalTransfer.Syntaxint cfCalTransfer(CF_HANDLE se

Página 112 - Example 3

6 - Language Dictionary72cfGetCellStatusSyntaxint cfGetCellStatus(CF_HANDLE server, int cell, CF_CELL_STATUS *status);DescriptionReturns a value in th

Página 113

Language Dictionary - 673cfGetDigitalConfigSyntaxint cfGetDigitalConfig(CF_HANDLE server, int bitnum, CF_EXT_SIGNAL*signal, CF_POLARITY *polarity, CF_

Página 114

6 - Language Dictionary74Examplevoid query_groups(CF_SERVER server){char names[CF_MAX_GROUPS][CF_MAX_GROUP_NAME_LEN];int starts[CF_MAX_GROUPS];int siz

Página 115 - Specifications

Language Dictionary - 675cfGetMeasLogIntervalSyntaxint cfGetMeasLogInterval(CF_HANDLE server, int step_number, float*volt_interval, float *curr_interv

Página 116

6 - Language Dictionary76cfGetOutputStateSyntaxint cfGetOutputState(CF_HANDLE server, CF_OUTPUT_STATE *state);DescriptionReturns the output state of t

Página 117 - bandwidth = 20Hz - 20MHz)

Language Dictionary - 677cfGetSenseProbeTestSyntaxint cfGetSenseProbeTest(CF_HANDLE server, CF_BOOLEAN *on_off);DescriptionReturns the setting of the

Página 118

6 - Language Dictionary78cfGetSeqTestAndSyntaxint cfGetSeqTestAnd(CF_HANDLE server, CF_READP *read_pos, int*step_number, CF_SEQ_TEST *meas_test_type,

Página 119 - Calibration

Language Dictionary - 679cfGetShutdownDelaySyntaxint cfGetShutdownDelay(CF_HANDLE server, float *delay);DescriptionReturns the delay value that is set

Página 120 - Calibration Connections

8cfSetDigitalPort 92cfSetErrorFunction 92cfSetGroup 93cfSetMeasLogInterval 93cfSetOutputConfig 93cfSetOutputProbeTest 94cfSetOutputState 94cfSetSense

Página 121 - Multimeter

6 - Language Dictionary80cfGetVoltageSyntaxint cfGetVoltage(CF_HANDLE server, float *voltage);DescriptionReturns the idle state voltage setting set by

Página 122 - Accessing Calibration

Language Dictionary - 681cfMeasCapacityWSSyntaxint cfMeasCapacityWS(CF_HANDLE server, int cell, float *reading);DescriptionReturns the accumulated cap

Página 123 - Calibration Error Messages

6 - Language Dictionary82To make an effective probe resistance measurement, there should be some significant currentthrough the probe contacts to the

Página 124

Language Dictionary - 683cfMeasVoltageSyntaxint cfMeasVoltage(CF_HANDLE server, int cell, float *reading);DescriptionReturns the measured cell voltage

Página 125 - Dimension Drawings

6 - Language Dictionary84Example#define MY_GROUP "1.5Ahour"/** Define group named "1.5Ahour" containing 64 cells* starting at cell

Página 126 - C - Dimension Drawings

Language Dictionary - 685The measurement log contains measurements acquired during the forming sequence. For thesequence step types CF_CHARGE, CF_DISC

Página 127

6 - Language Dictionary86For sequence steps of type CF_CHARGE, CF_DISCHARGE, or CF_REST, the format is: cell-number step-number time status entry-typ

Página 128 - 104103102101100999897

Language Dictionary - 687cfReadTestLogSyntaxint cfReadTestLog(CF_HANDLE server, CF_READP *read_pos, int bufsize,char *buffer, int *retcount);Descripti

Página 129

6 - Language Dictionary88cfRestartSyntaxint cfRestart(CF_HANDLE server);DescriptionThis command causes the Agilent MCCD to recall a previously saved r

Página 130

Language Dictionary - 689cfSetAutoConnectSyntaxint cfSetAutoConnect(CF_HANDLE server, CF_BOOLEAN on_off);DescriptionThis command turns the automatic r

Página 131

91General InformationAgilent MCCD System CapabilitiesThe Agilent Multi-Cell Charger/Discharger (MCCD) System has been designed to address the uniquere

Página 132

6 - Language Dictionary90cfSetDigitalConfigSyntaxint cfSetDigitalConfig(CF_HANDLE server, int bitnum, CF_EXT_SIGNALsignal, CF_POLARITY polarity, CF_RE

Página 133

Language Dictionary - 691When an output signal is programmed, the pin is driven by an open collector transistor. Writing aword to the port using cfSet

Página 134

6 - Language Dictionary92cfSetDigitalPortSyntaxint cfSetDigitalPort(CF_HANDLE server, int data);DescriptionWrite data to the digital I/O port. Data mu

Página 135 - In Case of Trouble

Language Dictionary - 693cfSetGroupSyntaxint cfSetGroup(CF_HANDLE server, char *name, int start, int size);DescriptionDefines a group of cells by spec

Página 136 - Selftest Error Messages

6 - Language Dictionary94cfSetOutputProbeTestNOTE: The Agilent MCCD must be configured for remote voltage sensing to perform outputprobe testing. No o

Página 137

Language Dictionary - 695cfSetSenseSyntaxint cfSetSense(CF_HANDLE server, CF_SENSE sense);DescriptionSets voltage sense to remote or local sense. The

Página 138

6 - Language Dictionary96CF_CHARGEConstant voltage/constant current chargeCF_DISCHARGEConstant voltage/constant current dischargeCF_RESTRest (output i

Página 139

Language Dictionary - 697cfSetSeqTestSyntaxint cfSetSeqTest(CF_HANDLE server, int step_number, CF_SEQ_TESTmeas_test_type, float limit, CF_TIME_TEST ti

Página 140

6 - Language Dictionary98CF_NEG_DIDT_LEThe change in the magnitude of current during the standard measurementinterval that is negative, and the magnit

Página 141

Language Dictionary - 699cfSetSeqTestAndSyntaxint cfSetSeqTestAnd(CF_HANDLE server, int step_number, CF_SEQ_TEST*meas_test_type, float *limit, CF_TIME