4- 40 Agilent B1500A User’s Guide, Edition 13
Specifications
Hardware
Table 4-22 Minimum timing parameters for current measurement
1
Voltage applied to DUT 10 V
Current applied to DUT 100 nA 1 A 10 A 100 A 1 mA 10 mA
Applied
voltage
condition
Recommended
minimum pulse
width
a
a. Recommended minimum pulse width = settling time + recommended minimum measurement
window.
47 s 38.7 s 6.8 s 950 ns 240 ns 145 ns
Current
measurement
condition
Measurement Range 1 A 1 A 10 A 100 A 1 mA 10 mA
Recommended
minimum
measurement
window
10 s 1.64 s 1 s 130 ns 40 ns 20 ns
Settling time
b
b. The time until the measured value settles to within ± 0.6 % of the final result value after the out-
put voltage is changed from the initial value (0 V). Minimum rise/fall time of 70 ns is recom-
mended for minimizing overshoot.
37 s 37 s 5.8 s 820 ns 200 ns 125 ns
Noise (rms)
c
c. RMS noise measured over the recommended minimum measurement window.
160 pA 425 pA 2.5 nA 47 nA 280 nA 1.9 A
1. Measurement conditions: The DUT is a resistive load chosen to adjust the flowing current to the
specified current in the table above. The capacitance of the cable between the RSU and the DUT
is 20 pF. Voltage is applied to the DUT by a channel of WGFMU/RSU in Fast IV mode and in
the 10 mA range, and current measurement is performed by another channel at 0 V in Fast IV
mode.