SCPI Command Reference 6
N9340A User’s Guide 131
[:SENSe]:POWer Subsection
Input Attenuation
[:SENSe]:POWer[:RF]:ATTenuation <para>
[:SENSe]:POWer[:RF]:ATTenuation?
This command sets the value of the attenuator.
Input Port Attenuator Auto
[:SENSe]:POWer[:RF]:ATTenuation:AUTO
OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
This command selects the input port attenuator
range to be set either automatically or manually.
On – Input attenuation is automatically set as
determined by the Reference Level Setting.
Off – Input attenuation is manually set.
Input Port Power Gain
[:SENSe]:POWer[:RF]:GAIN[:STATe]
OFF|ON|0|1
[:SENSe]:POWer[:RF]:GAIN[:STATe]?
This command turns the internal preamplifier on
or off.
*RST Value: 20 dB
Range: 0 dB to 51 dB
Available unit: dB
Key Access: [AMPTD] > {Attenuation}
*RST State: On
Key Access: [AMPTD] > {Attenuation}