8712ET/ES and 8714ET/ES Adjustments
RF Network Analyzers Source Power Correction, Adjustment #104
Service Guide 2-17
a. Insert a formatted disk into the internal disk drive.
b. Press .
This creates a file (or writes over an existing file) where the correction constants are
stored.
In Case of Difficulty
If the system locks up or the power meter does not respond to the analyzer, make sure the
GPIB cable is correctly connected and the GPIB address on the power meter is set to 13.
If a “log error” occurs on the power meter during the low power adjustment and the test
stops, you will need to do the test again. Save CCs to disk and then switch the analyzer off
and preset the power meter.
NOTE If you must cycle the analyzer's line power, and you did not store the
correction constants to disk before starting this adjustment, you need to
repeat all of the adjustments done up to this point, then repeat the source
power correction. If you did save the correction constants to disk, recall the
correction constants data from disk as described in Chapter 8, “Correction
Constants and Firmware,then go back to step 1 to repeat this adjustment.
SYSTEM OPTIONS
Service
Update Corr Const
Store CC to Disk