This document describes the performance and features of the following products:• Agilent 8753ES S-parameter vector network analyzer• Agilent 8753ES Op
10Test-Port Output Characteristics7Frequency characteristicsRange 30 kHz to 3 GHz Option 006 30 kHz to 6 GHzResolution 1 HzStability9±7.5 ppm (0° to 5
11External source mode1(CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements (typical)Power level 0 to -25 dBmSpectral
12Test Port Output CharacteristicsFrequency characteristicsRange 300 kHz to 3 GHz Option 006 30 kHz to 6 GHz Resolution 1 Hz Stability typically ±7.
13Frequency offset mode3Frequency range 300 kHz to 3 GHzOption 006 300 kHz to 6 GHz R-channel input requirements Power level300 kHz to 3 GHz 0 to -35
14Test-Port Output CharacteristicsFrequency characteristicsRange 300 kHz to 3 GHzOption 006 300 kHz to 6 GHzResolution 1 HzStability4±7.5 ppm (0° to
15External source mode1(CW time sweep only)Frequency range 300 kHz to 6 GHz R-channel input requirements2Power level 0 to -25 dBmSpectral purity Ma
MeasurementNumber of display channelsFour display channels available.Number of measurement channels Two primary and two auxiliary measurementchannels
17Sweep triggerSet to either continuous, single, group sweep, orexternal trigger. Set external trigger to take a com-plete sweep or to measure individ
18Trace FunctionsDisplay data Display current measurement data, memory data,or current measurement with measurement andmemory data simultaneously. Tr
19Frequency offset on/offSets the RF source to be swept at a fixed offset frequency above the receiver as required in a sweptRF/IF, fixed LO, mixer te
341021212224Table of ContentsDefinitions and Test ConditionsSystem Performance SummariesSpecifications and CharacteristicsMeasurement Throughput Summa
20Front Panel Connectors 8753ES test ports (without Option 011) Connector type 7 mm, precision Impedance 50 ohms (nominal) 8753ES Option 011 test po
21General ConditionsRFI and EMI susceptibility: defined by VDE 0730,CISPR Publication 11, and FCC Class B Standards.ESD (electrostatic discharge): mus
22Agilent 85046A/B S-Parameter Test SetsThe 85046A/B S-parameter test sets provide thecapability to measure reflection and transmissioncharacteristics
23Agilent 85047A S-Parameter Test SetThe 85047A S-parameter test set provides thecapability to simultaneously measure the reflec-tion and transmission
24Calibration KitsVector accuracy enhancement procedures requirethat the systematic errors of the measurement sys-tem be characterized by measuring kn
25Agilent 85038 Family of Calibration KitsThe 85038A calibration kit contains open circuits,short circuits, and terminations with both maleand female
26Verification KitsMeasuring known devices other than the standardsused in calibration is an easy way to verify the proper operation of an 8753ES meas
27Connectors 50-ohm type-N (f) to75-ohm type-N (m)standard Option 004, 50-ohm type-N (m) to 75-ohm type-N (f)Dimensions 14-mm D x 70-mm L(0.56 in x 2
Agilent 8347A RF AmplifierThe 8347A RF amplifier delivers increased power across a 300 kHz to 3 GHz frequency range. The 8347A provides leveled output
3Specifications describe the instrument’s warrantedperformance after a half-hour warm-up and overthe temperature range of 25° ± 5° C, unless other-wis
4Agilent 8753ES (50-ohm systems) 7-mm test portsConfigurationNetwork analyzer 8753ES Standard and Option 006Calibration kit 85031BTest-port cables 118
5Agilent 8753ES (50-ohm systems)Type-N test portsConfigurationNetwork analyzer 8753ESStandard and Option 006Calibration kit 85032BTest-port cables
6Agilent 8753ES (50-ohm systems)3.5-mm test portsConfigurationNetwork analyzer 8753ESStandard and Option 006Calibration kit 85033DTest-port cables 118
Agilent 8753ES (75-Ohm Systems)Type-N test portsConfigurationNetwork analyzer 8753ES Option 075Calibration kit 85036BTest-port cables 11857BSystem Dyn
8Agilent 8753ES (75-ohm systems)Type-F test portsConfigurationNetwork analyzer 8753E Option 075Calibration kit 85039BTest-port cables 11857BSystem dyn
91. One-port or enhanced-response calibration.2. Enhanced-response or response-only calibration.3. 2 dB less with Option 004.4. Option 004 may degrade