Agilent Technologies AGILENT U2761A Manual de Serviço Página 31

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 30
Calibration 3
U2761A Service Guide 17
Test Consideration
For optimum performance, all procedures should comply
with the following recommendations.
Ensure that the calibration ambient temperature is stable
and between 23 °C ± 3 °C.
The ambient relative humidity must be less than 80%.
Allow a one-hour warm-up period before verification or
adjustment.
Keep the measurement cables as short as possible and
consistent with the impedance requirement.
Use RG-58 or equivalent 50
Ω cable.
Vista de página 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 66 67

Comentários a estes Manuais

Sem comentários