Vista de página 0
Agilent X-Series Signal
Analyzer
This help file provides documentation for the following
X-Series Analyzers:
PXASignal Analyzer N9030A
MXASignal Analyzer N9020A
EXASignal Analyzer N9010A
N9061A Remote Language
Compatibility Measurement
Application User's
&Programmer's Reference
Vista de página 0
1 2 3 4 5 6 ... 145 146

Resumo do Conteúdo

Página 1 - Analyzer

Agilent X-Series SignalAnalyzerThis help file provides documentation for the followingX-Series Analyzers:PXASignal Analyzer N9030AMXASignal Analyzer

Página 2

Table Of ContentsLegacy Products 252Description 252C2 [two] (A Minus B Into A) 253Syntax 253Legacy Products 253Description 253CA (Couple Attenuation)

Página 3

3 Programming the AnalyzerList of Supported SCPI CommandsU:UNIT:POWer "Y Axis Unit" on page 533100 Remote Language Compatibility Measuremen

Página 4 - 3 Programming the Analyzer 73

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRestartMore InformationThe Restart function first aborts the current sweep/measurement as

Página 5

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveSaveThe Save menu lets you choose what you want to save and where you want to save it.

Página 6

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveExample MMEM:STOR:STATe "MyStateFile.state"This stores the current instrumen

Página 7

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveThe Listed beloware the functions of the various fields in the dialog, and the corresp

Página 8

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSavedefault path for this type of file and remembers the last path you used to save this t

Página 9

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveMode AllNotes Brings up Save As dialog for saving a State Save TypeInitial S/W Revisio

Página 10 - Table Of Contents

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveWhile the save is being performed, the floppy icon appears briefly in the Meas bar.Fil

Página 11

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveThis key corresponds to the icon of a folder with the "*" that is in the too

Página 12

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveThe Listed beloware the functions of the various fields in the dialog, and the corresp

Página 13

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSavedefault path for this type of file and remembers the last path you used to save this t

Página 14

3 Programming the AnalyzerIEEE 488.2 Common CommandsIEEE 488.2 Common CommandsThe instrument supports the following subset of IEEE 488.2 Common Comma

Página 15

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveMode AllNotes Brings up Save As dialog for saving a State Save TypeInitial S/W Revisio

Página 16

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveWhile the save is being performed, the floppy icon appears briefly in the Meas bar.Fil

Página 17

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveThis key corresponds to the icon of a folder with the "*" that is in the too

Página 18

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveThe Listed beloware the functions of the various fields in the dialog, and the corresp

Página 19

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSavedefault path for this type of file and remembers the last path you used to save this t

Página 20

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveMode AllNotes Brings up Save As dialog for saving a State Save TypeInitial S/W Revisio

Página 21

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveWhile the save is being performed, the floppy icon appears briefly in the Meas bar.Fil

Página 22

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveThis key corresponds to the icon of a folder with the "*" that is in the too

Página 23

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveThe Listed beloware the functions of the various fields in the dialog, and the corresp

Página 24

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSavedefault path for this type of file and remembers the last path you used to save this t

Página 25

3 Programming the AnalyzerIEEE 488.2 Common CommandsCondition message “Align Now, RF required” is generated, and bits 11 and 12 are set in the Status

Página 26

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveMode AllNotes Brings up Save As dialog for saving a State Save TypeInitial S/W Revisio

Página 27

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveWhile the save is being performed, the floppy icon appears briefly in the Meas bar.Fil

Página 28

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveThis key corresponds to the icon of a folder with the "*" that is in the too

Página 29

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveOR“(empty)" if no prior save operation has been performed to this register.Initia

Página 30

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveRegister 1 thru Register 16Selecting any one of these register menu keys causes the St

Página 31

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveAlthough these 16 registers are the only registers available from the front panel, the

Página 32

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveAfter the save completes, the corresponding register menu key annotation is updated wi

Página 33

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveExample *SAV 1Range 1–16 from front panel, 1–128 from SCPIReadback Date and time with

Página 34

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveOR“(empty)" if no prior save operation has been performed to this register.Initia

Página 35

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveRegister 1 thru Register 16Selecting any one of these register menu keys causes the St

Página 36

3 Programming the AnalyzerIEEE 488.2 Common CommandsWhile Align Now, All is performing the alignment, bit 0 in the Status Operation register is set.C

Página 37

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveAlthough these 16 registers are the only registers available from the front panel, the

Página 38

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveAfter the save completes, the corresponding register menu key annotation is updated wi

Página 39

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveExample *SAV 1Range 1–16 from front panel, 1–128 from SCPIReadback Date and time with

Página 40

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveOR“(empty)" if no prior save operation has been performed to this register.Initia

Página 41 - 1 About the Analyzer

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveRegister 1 thru Register 16Selecting any one of these register menu keys causes the St

Página 42 - Viewing a License Key

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveAlthough these 16 registers are the only registers available from the front panel, the

Página 43 - 1 About the Analyzer

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave:MMEMory:REGister:STATe:LABel? <reg number>Example :MMEM:REG:STAT:LAB 1,”my labe

Página 44

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSavesettings, even though they are outside of the Mode’s state, because they are needed to

Página 45 - Getting Started Guide

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveBoth single and double quotes are supported for any filename parameter over remote.Aft

Página 46 - Display Annotations

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveAmplitude CorrectionPressing this key selects Amplitude Corrections as the data type t

Página 47 - Window Control Keys

3 Programming the AnalyzerIEEE 488.2 Common CommandsKey PathNo equivalent key. Related key System, Show Errors, Clear Error QueueRemote Command*CLSEx

Página 48 - Next Window

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveLine# Type of field Example Notes“Amplitude Correction”2 File Description (in quotes)

Página 49 - :DISPlay:WINDow[:SELect]?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveLines 2 through 5 can be empty but must appear in the file. Lines 6 through 11 are op

Página 50 - Mouse and Keyboard Control

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveA.02.06,N9020AP13 EA3 UK6,01Frequency Unit,MHzAntenna Unit,dBuV/mFrequency Interpolati

Página 51

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSavePreset Not part of a Preset, but is reset to Correction 1 by Restore Input/Output Defa

Página 52 - PC Keyboard

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveKey PathSave, Data, Amplitude CorrectionPreset Not part of a Preset, but is reset to C

Página 53

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveCouplings When you select which trace to save, it makes that trace the current trace,

Página 54

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave• Sweep Time• Start Frequency• Stop Frequency• Average Count (actual; not the limit fo

Página 55

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave• RBW Filter BW: 3dB, 6dB, Noise, Impulse• Sweep Type: Swept, FFT• PreAmp State: On, O

Página 56

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveAverage Count 0Average Type Power(RMS)RBW 51000RBW Filter GaussianRBW Filter BW 3dBVBW

Página 57 - Application

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveY Axis Unit dBmDATA1.6009301E+07 4.82047E+011.6018694E+07 4.69737E+011.6028087E+07 4.8

Página 58

3 Programming the AnalyzerIEEE 488.2 Common CommandsStandard Event Status Register QueryQueries and clears the standard event status event register.

Página 59 - Instrument Updates

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveNotes auto returnCouplings When you select which trace to save, it makes that trace th

Página 60 - General Rules and Limitations

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSavecurrently selected trace, selected in this menu or in the Trace/Det, Import Data, Reca

Página 61 - Predefined Functions

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveKey PathSave, Data, TraceNotes auto returnCouplings When you select which trace to sav

Página 62 - Supported Commands

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveSelect TraceThese softkeys let you pick which Trace to save. Once selected, the key re

Página 63 - OA Parameter

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveThis menu is the same as the Select Trace menu under Trace. The trace selected on that

Página 64

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveLimits File ContentsLimits may be exported into a data file with a .csv extension. The

Página 65

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveIf the sign of the margin is inappropriate for the limit type (for example a positive

Página 66 - Verify the Installation

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveLimit SelectionThese keys let you pick which Limit Line to save. Once selected, the ke

Página 67

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveLimit SelectionThese keys let you pick which Limit Line to save. Once selected, the ke

Página 68 - Setting up N9061A

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSavemessage is generatedIf a save of Spectrogram results is requested and the Spectrogram

Página 69 - Hints and Tips

3 Programming the AnalyzerIEEE 488.2 Common CommandsOperation CompleteThe *OPC command sets bit 0 in the standard event status register (SER) to “1”

Página 70 - AC and DC Coupling

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveMeas-urementResultSwept SAA.01.40_R0017 N9020A526 B25 PFR P26EA31Result Type MarkerTab

Página 71 - Service and Calibration

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaverangeRF Coupling ACFFT Width 411900Ext Ref 10000000Input RFRF Calibrator OffAttenuatio

Página 72

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave39 Fixed 3 InverseTime4.0000E-+01–30.0-5Off 0.0000E-+000 None10 Normal 3 Frequen-cy1.1

Página 73 - 3 Programming the Analyzer

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveThen the Meas Results file, when opened, would show the header data (the same as for t

Página 74 - Resource Description

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveMeasurementResultSwept SAA.01.40_R0017 N9020A526 B25 PFR P26 EA3 1Result Type Peak Tab

Página 75

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveRef Level Offset 0External Gain 0X Axis Units HzY Axis Units dBmPeak Threshold –85Peak

Página 76

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveSpectrogramThis section discusses the Spectrogram Results file format. The Spectrogr

Página 77

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveFor the purpose of this example, we have set the Average/Hold Number to 10, thus we ha

Página 78

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveResult Type SpectrogramDRD EA3 EDP EMC EP1 ERC ESCESP EXM FSA HBA K03 LFE MPBP03 P08 P

Página 79

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveResult Type SpectrogramRef Level Offset 0External Gain 0Trace Type ClearwriteDetector

Página 80

3 Programming the AnalyzerIEEE 488.2 Common CommandsQuery Instrument OptionsReturns a string of all the installed instrument options. It is a comma s

Página 81

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave5999984490 –74.027120795999984515 –73.57213005ooo6000009315 –75.91831036000009340 –79.

Página 82

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveMeasurement ResultsPressing this key selects Meas Results as the data type to be expor

Página 83

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveImagine that, at the point where a Marker Table Meas Result is requested, the followin

Página 84

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveStart Frequency 10000000Stop Frequency 26500000-000Average Count 0Average Type LogPowe

Página 85

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveMKR MODE TR-CSCL X Y FUNC-TIONFUNC-TIONWIDTHFUNC-TIONVALUEFUNC-TIONUNIT1 Normal 1 Freq

Página 86

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveThe metadata header is very similar to the metadata used in the trace data .csv files.

Página 87

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave• Peak Readout (All|AboveDL|BelowDL)• Peak Sort (Freq|Amptd)These fields are then foll

Página 88

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveRF Coupling ACFFT Width 411900Ext Ref 10000000Input RFRF Calibrator OffAttenuation 10R

Página 89

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave15 9.5140E+05 –60.8916 9.5920E+05 –60.9017181920SpectrogramThis section discusses the

Página 90

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveTraces that have not yet been filled in the Spectrogram display are empty; there is no

Página 91

3 Programming the AnalyzerIEEE 488.2 Common Commands*RST (Remote Command Only)*RST is equivalent to :SYST:PRES;:INIT:CONT OFF, which is a Mode Preset

Página 92

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveResult Type SpectrogramSwept SAA.11.00.01 N9020A503 508 513 526 ALL ALV B1C B1XB25 B2X

Página 93

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveResult Type SpectrogramInput RFRF Calibrator OffAttenuation 14Ref Level Offset 0Extern

Página 94

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave5999984415 –80.981978825999984440 –80.981978795999984465 –75.831421325999984490 –74.02

Página 95

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveMeasurement ResultsPressing this key selects Meas Results as the data type to be expor

Página 96

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveImagine that, at the point where a Marker Table Meas Result is requested, the followin

Página 97

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveStart Frequency 10000000Stop Frequency 26500000-000Average Count 0Average Type LogPowe

Página 98

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveMKR MODE TR-CSCL X Y FUNC-TIONFUNC-TIONWIDTHFUNC-TIONVALUEFUNC-TIONUNIT1 Normal 1 Freq

Página 99

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveThe metadata header is very similar to the metadata used in the trace data .csv files.

Página 100 - 3 Programming the Analyzer

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave• Peak Readout (All|AboveDL|BelowDL)• Peak Sort (Freq|Amptd)These fields are then foll

Página 101 - IEEE 488.2 Common Commands

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveRF Coupling ACFFT Width 411900Ext Ref 10000000Input RFRF Calibrator OffAttenuation 10R

Página 102 - :CALibration[:ALL]?

3 Programming the AnalyzerIEEE 488.2 Common CommandsService Request EnableThis command enables the desired bits of the service request enable registe

Página 103 - Clear Status

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave15 9.5140E+05 –60.8916 9.5920E+05 –60.9017181920SpectrogramThis section discusses the

Página 104 - Standard Event Status Enable

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveTraces that have not yet been filled in the Spectrogram display are empty; there is no

Página 105 - Identification Query

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveResult Type SpectrogramSwept SAA.11.00.01 N9020A503 508 513 526 ALL ALV B1C B1XB25 B2X

Página 106 - Operation Complete

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveResult Type SpectrogramInput RFRF Calibrator OffAttenuation 14Ref Level Offset 0Extern

Página 107 - Recall Instrument State

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave5999984415 –80.981978825999984440 –80.981978795999984465 –75.831421325999984490 –74.02

Página 108 - Save Instrument State

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveMeasurement ResultsPressing this key selects Meas Results as the data type to be expor

Página 109 - Trigger

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveImagine that, at the point where a Marker Table Meas Result is requested, the followin

Página 110 - Wait-to-Continue

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveStart Frequency 10000000Stop Frequency 26500000-000Average Count 0Average Type LogPowe

Página 111 - Remote Measurement Functions

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveMKR MODE TR-CSCL X Y FUNC-TIONFUNC-TIONWIDTHFUNC-TIONVALUEFUNC-TIONUNIT1 Normal 1 Freq

Página 112 - MEASure Command

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveThe metadata header is very similar to the metadata used in the trace data .csv files.

Página 113 - FETCh Command

Table Of ContentsDescription 262CLRW (Clear Write) 263Syntax 263Legacy Products 263Description 263CONTS (Continuous Sweep) 265Syntax 265Legacy Product

Página 114 - READ Command

3 Programming the AnalyzerIEEE 488.2 Common CommandsRemote Command*TRGExample *TRG Triggers the instrument to take a sweep or start a measurement, de

Página 115 - :CONFigure?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave• Peak Readout (All|AboveDL|BelowDL)• Peak Sort (Freq|Amptd)These fields are then foll

Página 116 - :CALCulate:DATA[n]?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveRF Coupling ACFFT Width 411900Ext Ref 10000000Input RFRF Calibrator OffAttenuation 10R

Página 117 - Option Description

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave15 9.5140E+05 –60.8916 9.5920E+05 –60.9017181920SpectrogramThis section discusses the

Página 118

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveTraces that have not yet been filled in the Spectrogram display are empty; there is no

Página 119

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveResult Type SpectrogramSwept SAA.11.00.01 N9020A503 508 513 526 ALL ALV B1C B1XB25 B2X

Página 120

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveResult Type SpectrogramInput RFRF Calibrator OffAttenuation 14Ref Level Offset 0Extern

Página 121 - Variable Description

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave5999984415 –80.981978825999984440 –80.981978795999984465 –75.831421325999984490 –74.02

Página 122

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveSave As . . .When you press “Save As”, the analyzer brings up a Windows dialog and a m

Página 123 - FREQuency

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveFor all of the Limit Data Files:My Documents\<mode name>\data\limitsFor all of t

Página 124 - Data Format Commands

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveKey PathSave, DataMode AllNotes The key location is mode-dependent and will vary.Bring

Página 125

3 Programming the AnalyzerRemote Measurement FunctionsRemote Measurement FunctionsThis section contains the following topics:l "Measurement Grou

Página 126 - 4

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveSave As . . .When you press “Save As”, the analyzer brings up a Windows dialog and a m

Página 127 - STATus Subsystem

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveFor all of the Limit Data Files:My Documents\<mode name>\data\limitsFor all of t

Página 128 - What Are Status Registers

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveKey PathSave, DataMode AllNotes The key location is mode-dependent and will vary.Bring

Página 129

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveScreen ImagePressing Screen Image accesses a menu of functions that enable you to spec

Página 130

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveFor versions previous to A.01.55, if you initiate a screen image save by navigating th

Página 131 - Using a Status Register

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSave3D ColorSelects a standard color theme with each object filled, shaded and colored as

Página 132

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveSave As…When you press “Save As”, the analyzer brings up a Windows dialog and a menu e

Página 133 - Status Register System

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveThe default path for Screen Images isMy Documents\<mode name>\screen.where <m

Página 134 - The Status Byte Register

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveKey PathSave, Screen ImageNotes Brings up Save As dialog for saving a Screen Image Sav

Página 135

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveSave As…When you press “Save As”, the analyzer brings up a Windows dialog and a menu e

Página 136

3 Programming the AnalyzerRemote Measurement FunctionsMEASure Command:MEASure:<measurement>[n]?This is a fast single-command way to make a meas

Página 137

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveNotes The string must be a valid logical path.Changes the default directory for a mass

Página 138

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveRemote Command:MMEMory:DATA <file_name>, <data>:MMEMory:DATA? <file_nam

Página 139

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSaveMass Storage Remove Directory (Remote Command Only)Remote Command:MMEMory:RDIRectory &

Página 140 - Operation Register

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSingle (Single Measurement/Sweep)Single (Single Measurement/Sweep)Sets the analyzer for Si

Página 141 - :STATus:OPERation:ENABle?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourceSourceThere is no Source control functionality for this measurement. When this key i

Página 142 - :STATus:OPERation[:EVENt]?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourceAmplitudeAllows you to access the Amplitude sub-menu to control various amplitude pa

Página 143 - Questionable Register

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourceThis alias is for the ESA tracking generator. It specifies the source output power

Página 144 - :STATus:QUEStionable:ENABle?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourceKey PathSource, AmplitudeMode SAScope Mode GlobalRemote Command:SOURce:POWer:SWEep &

Página 145 - :STATus:QUEStionable[:EVENt]?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourcel SOURce:POWer:MODE SWEep is equivalent to :SOURce:POWer:SWEep:STATeONPreset This is

Página 146

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourceAmptd Step Auto/ManAllows you to set the step size associated with the Source > A

Página 147

3 Programming the AnalyzerRemote Measurement FunctionsMeasurement settings persist if you initiate a different measurement and then return to aprevio

Página 148

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourceSource Frequency = (Analyzer Frequency *Multiplier Numerator / Multiplier Denominato

Página 149

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourceMultiplier DenominatorThe multiplier denominator parameter offsets the source freque

Página 150

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourcesweeping in the opposite direction from the analyzer, and it will not be possible tr

Página 151

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourceMode SAScope Mode GlobalRemote Command:SOURce:FREQuency:OFFSet <freq>:SOURce:F

Página 152

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourceAs stated above, when the Source Mode is set to Tracking, the analyzer acquires cont

Página 153

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourceWhen you set the Source Mode to OFF, it releases the Source (and puts it into Local)

Página 154

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourceAgilent EXG N5172B X X X X XAgilent MXG N5182B X X X X XAgilent PSG E8257D X X XAgil

Página 155

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourceFor X-Series software versions earlier than A.10.01, hardware triggering was unavail

Página 156

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourceFor example, if no Source has yet been selected, the statement at the bottom of the

Página 157

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSource[None] on a Restore Input/Output Defaults.If an internal Tracking Generator is insta

Página 158

3 Programming the AnalyzerRemote Measurement FunctionsSee each command for details of what types of scalar results or trace data results are availabl

Página 159

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourcel Amplitude Step = Autol Power Sweep = 0 dBl Amplitude Offset = 0 dBl Source Sweep R

Página 160

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSourceSource Setting Query (Remote Command Only)This query can be used to get certain sett

Página 161

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSPAN X ScaleSPAN X ScaleActivates the Span function and displays a menu of span functions.

Página 162

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSPAN X ScaleIf Source Mode is set to Tracking, and the Span is therefore limited by the li

Página 163

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSPAN X ScaleSpan PresetsThe following table provides the Span Presets for the Spectrum Ana

Página 164

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSPAN X ScaleNote that if you are in External Mixing, the maximum Span will be equal to the

Página 165

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSPAN X ScaleThe center frequency for the lower window is limited by the start and stop fre

Página 166

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSPAN X ScaleThe span is dependent on the currently selected Input (see the Section “Input/

Página 167

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSPAN X ScaleZero Span Swept SpansX axis is time X axis is frequencyThere is no auto-RBW se

Página 168 - Questionable Power Register

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSPAN X ScaleIf this key is pressed while in a nonzero span, and the previous value of span

Página 169

3 Programming the AnalyzerRemote Measurement Functionschanges from Channel Power back to ACP and, using the previous ACP settings, themeasurement is

Página 170

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSPAN X ScaleSignal Track can only function properly if the trace Marker 1 is on is updatin

Página 171

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSPAN X ScaleIf the signal is lost, an attempt will be made to find it again and continue t

Página 172

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlSweep/ControlAccesses a menu that enables you to configure the Sweep and Cont

Página 173

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlTime takes into account both the data acquisition time and the processing tim

Página 174

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlZero Span, the Auto/Man state of Sweep Time that existed before entering Zero

Página 175

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlSwept Sweep Type (either manually or automatically chosen) and not when in FF

Página 176 - Series Description

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlSetting Sweep Time Rules to SA-Accuracy will result in slower sweep times tha

Página 177

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlReadback SA - NormalInitial S/W Revision Prior to A.02.00SA - AccuracyChooses

Página 178

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Controlnarrowband filters. The number of measurement points can also be reduced to s

Página 179

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlPreset AUTOBackwardsCompatibility SCPI[:SENSe]:SWEep:TYPE AUTOsets sweep type

Página 180

3 Programming the AnalyzerRemote Measurement Functionsl Returns a 0 if the measured results pass when compared with the current limits.l Returns a 1

Página 181

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlState Saved Saved in instrument stateReadback SweptInitial S/W Revision Prior

Página 182

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlPreset DRANgeState Saved Saved in instrument stateBackwards CompatibilityNote

Página 183

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlExample SWE:TYPE:AUTO:RUL DRAN sets the auto rules to dynamic range.Couplings

Página 184

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlRemote Command[:SENSe]:SWEep:FFT:WIDTh <real>[:SENSe]:SWEep:FFT:WIDTh?E

Página 185

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlKey PathSweep/Control, Sweep SetupRemote Command[:SENSe]:SWEep:FFT:WIDTh:AUTO

Página 186

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlFurther improvement in dynamic range is possible by changing the FFT IF Gain

Página 187

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlNote that the FFT Width setting will have no effect unless in an FFT sweep.Se

Página 188

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Controlstill less than the intended segment width. The “Span” used in this computati

Página 189

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Control–8 dBm at the input mixer). But FFT processing experiences overloads at the m

Página 190

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Controlincluding the Span you have set. Usually the instrument picks the optimal FF

Página 191

3 Programming the AnalyzerRemote Measurement Functionsthe first burst in each frame. The command can also be used to identify the best curve fit for

Página 192

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlOption B40, 40 MHz.BackwardsCompatibility SCPI[:SENSe]:SWEep:FFT:SPAN:RATio &

Página 193

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlAlthough narrowing the segment width can allow higher dynamic ranges some cas

Página 194

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlFFT WidthThis menudisplays and controls the width of the FFT’s performed whil

Página 195

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlState Saved Saved in instrument stateMin 4.01 kHzMax The maximum available FF

Página 196

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlHowever, in order to increase dynamic range, most X-series models provide a s

Página 197

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlThe 10 MHz path is always used for Swept sweeps. It is always used for FFT sw

Página 198

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlKey PathSweep/Control, Sweep SetupRemote Command[:SENSe]:SWEep:FFT:WIDTh <

Página 199

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlInitial S/W Revision Prior to A.02.00Modified at S/WRevisionA.04.00Key PathSw

Página 200

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Controldrive level. If you reduce the width of an FFT, an analog filter is placed be

Página 201

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Controlthe current setup; but on occasion you may wish to limit the FFT Width to be

Página 202

3 Programming the AnalyzerRemote Measurement FunctionsOption DescriptionMEAN("Note 1" onpage 120)Returns a single value that is the arithme

Página 203

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlBackwardsCompatibility SCPI[:SENSe]:SWEep:FFT:SPAN:RATio <integer>[:SEN

Página 204

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlAlthough narrowing the segment width can allow higher dynamic ranges some cas

Página 205 - 5 Legacy Command Descriptions

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlFFT WidthThis menudisplays and controls the width of the FFT’s performed whil

Página 206 - Command Syntax

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlState Saved Saved in instrument stateMin 4.01 kHzMax The maximum available FF

Página 207 - Syntax Elements

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlHowever, in order to increase dynamic range, most X-series models provide a s

Página 208 - Command Descriptions

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlThe 10 MHz path is always used for Swept sweeps. It is always used for FFT sw

Página 209 - Description

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlKey PathSweep/Control, Sweep SetupRemote Command[:SENSe]:SWEep:FFT:WIDTh <

Página 210

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlInitial S/W Revision Prior to A.02.00Modified at S/WRevisionA.04.00Key PathSw

Página 211

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Controldrive level. If you reduce the width of an FFT, an analog filter is placed be

Página 212 - A4 [four] (Blank Trace A)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Controlthe current setup; but on occasion you may wish to limit the FFT Width to be

Página 213

3 Programming the AnalyzerRemote Measurement FunctionsOption DescriptionEquation 4: RMS Value of Data Points for Specified Region(s)where Xi is a dat

Página 214

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlBackwardsCompatibility SCPI[:SENSe]:SWEep:FFT:SPAN:RATio <integer>[:SEN

Página 215

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlAlthough narrowing the segment width can allow higher dynamic ranges some cas

Página 216

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlFFT WidthThis menudisplays and controls the width of the FFT’s performed whil

Página 217

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlState Saved Saved in instrument stateMin 4.01 kHzMax The maximum available FF

Página 218

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlHowever, in order to increase dynamic range, most X-series models provide a s

Página 219

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlThe 10 MHz path is always used for Swept sweeps. It is always used for FFT sw

Página 220

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlWhen the Gate Function is on, the selected Gate Method is used along with the

Página 221

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Controll When Gate is on, Offset Res BW and Offset Video BW are ignored (if you set

Página 222

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Control• When Gate View is turned on, the instrument is set to Zero Span.• Gate View

Página 223

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlA sample of the Gate View screen in other measurements is shown in the follow

Página 224

Table Of ContentsDET (Detection Mode) 273Syntax 273Legacy Products 273Description 273DL (Display Line) 274Syntax 274Legacy Products 274Description 275

Página 225

3 Programming the AnalyzerRemote Measurement FunctionsOption Descriptionwhere |Xi| is the magnitude of an I/Q pair, X is the mean of the magnitudes f

Página 226

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Controlchanges, even if the window is not being updated. In Gated LO and Gated Video

Página 227 - Query Data Type Details

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Control[:SENSe]:SWEep:EGATe:TIME?Example SWE:EGAT:TIME 500 msDependencies Gate View

Página 228

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlGate DelayControls the length of time from the time the gate condition goes T

Página 229

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlThe key is also grayed out if Gate Control = Level.Preset 461.6 usWiMAX OFDMA

Página 230

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlWhen in zero span, there is no actual sweep performed. But data is only taken

Página 231

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlGated FFT cannot be done in zero span since the instrument is not sweeping. S

Página 232 - AMB (A minus B into A)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlLinePressing this key, when it is not selected, selects the line signal as th

Página 233

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlExternal 1Pressing this key, when it is not selected, selects an external inp

Página 234 - ANNOT (Annotation)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlBackwardsCompatibility SCPI:TRIGger[:SEQuence]:EXTernal:LEVelFor backward com

Página 235

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlTRIG:<meas>:SOUR EXT2 Measurements other than Swept SADependencies I

Página 236 - AT (Input Attenuation)

3 Programming the AnalyzerRemote Measurement FunctionsNon-Constant EnvelopeSee table below for explanation of variables.Variable Description<soffs

Página 237

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlTrig SlopeControls the trigger polarity. It is set positive to trigger on a r

Página 238

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Controlmessage ("Waiting for trigger") is generated if no trigger signal a

Página 239

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlMin –200 dBmMax 100 dBmDefault Unit depends on the current selected Y-Axis un

Página 240 - AUTOCPL (Auto Coupled)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Control3. If the new absolute RF Burst level differs from the previous by more than

Página 241

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlRemote Command:TRIGger[:SEQuence]:RFBurst:SLOPe POSitive|NEGative:TRIGger[:SE

Página 242

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlThis feature selects the internal periodic timer signal as the trigger. Trigg

Página 243

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlPeriodSets the period of the internal periodic timer clock. For digital commu

Página 244

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlOffsetAdjusts the accumulated offset between the periodic timer events and th

Página 245 - B4 [four] (Blank Trace B)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlPreset 0 sState Saved Saved in instrument stateMin –10.000 sMax 10.000 sDefau

Página 246

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlPreset OffGSM/EDGE: RFBurstState Saved Saved in instrument stateReadback The

Página 247 - BLANK (Blank Trace)

3 Programming the AnalyzerRemote Measurement FunctionsVariable DescriptionIts unit is seconds for time-domain traces, or a dimensionless index (0 to

Página 248

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlThe Status Operation Register bit 5 "Waiting for Trigger" is set at

Página 249

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Control:TRIGger[:SEQuence]:EXTernal1:SLOPe?Example TRIG:EXT1:SLOP NEGCouplings This

Página 250

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlInitial S/W Revision Prior to A.02.00Modified at S/WRevisionA.03.00Trigger Le

Página 251

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlPreset POSitiveState Saved Saved in instrument stateBackwardsCompatibility SC

Página 252

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlAbsolute Trigger LevelSets the absolute trigger level for the RF burst envelo

Página 253 - C2 [two] (A Minus B Into A)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Control:TRIGger[:SEQuence]:RFBurst:LEVel:TYPE?Example TRIG:RFB:LEV:TYPE RELsets the

Página 254 - CA (Couple Attenuation)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/Controlneed to send the :TRIGger[:SEQuence]:RFBurst:LEVel:TYPE command, above.The re

Página 255 - CARROFF (Carrier Off Power)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlTVPressing this key, when it is not selected, selects the TV input signal as

Página 256 - CARRON (Carrier On Power)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlSCPI command is sent while the key is blanked, an error is returned.Readback

Página 257 - CF (Center Frequency)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlKey PathTrigger, TVRemote Command:TRIGger[:SEQuence]:TV:FMODe ENTire|ODD|EVEN

Página 258

3 Programming the AnalyzerRemote Measurement Functions:CALCulate:DATA[1]|2|...6:PEAKs? <threshold>,<excursion>[,AMPLitude |FREQuency | TI

Página 259 - CHANNEL (Channel Selection)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlMaximum line is 313 for formats PAL-B, D, G, H, I, PAL-N, PAL-N Combin, and S

Página 260 - CHANPWR (Channel Power)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlPreset MNTSState Saved Saved in instrument stateReadback Displays Readback va

Página 261

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlKey PathTrigger, TV, StandardExample TRIG:TV:STAN MPALReadback PAL-MInitial S

Página 262 - CLRAVG (Clear Average)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlKey PathTrigger, TV, StandardExample TRIG:TV:STAN PAL60Readback PAL-NInitial

Página 263 - CLRW (Clear Write)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlGate HoldoffLets you increase or decrease the wait time after a gate event en

Página 264

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlExample SWE:EGAT:HOLD 0.0002SWE:EGAT:HOLD?SWE:EGAT:HOLD:AUTO ONSWE:EGAT:HOLD:

Página 265 - CONTS (Continuous Sweep)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlThis command sets gate parameter values to the ESA preset values, as follows:

Página 266 - COUPLE (Input Coupling)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlSWE:EGAT:POL?Preset POSitiveState Saved Saved in instrument stateBackwardsCom

Página 267

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlKey PathSweep/ControlRemote Command[:SENSe]:SWEep:POINts <integer>[:SEN

Página 268

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlBackwards CompatibilityNotes1. In ESA and PSA, Sweep Points was adjustable wi

Página 269 - CT (Couple Sweep Time)

3 Programming the AnalyzerRemote Measurement Functionsl TIME - lists the peaks in order of occurrence, left to right across the x-axis.lPeaks vs. Dis

Página 270 - CV (Couple Video Bandwidth)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlWhen Paused, the label on the key changes to Resume. Pressing Resume un-pause

Página 271 - DA (Display Address)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSweep/ControlStatus Bits/OPCdependenciesThe STATus:OPERation register bits 0 through 8 are

Página 272

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemSystemOpens a menu of keys that access various configuration menus and dialogs.Key P

Página 273 - DET (Detection Mode)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemrelation to the size of the SCPI queue. If the queue extends onto a second page, a s

Página 274 - DL (Display Line)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemnumbers in the X-Series and those of earlier products. Error, event, and status proc

Página 275

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemHistoryThe History and Status keys select the Errors view. The Status key has a seco

Página 276 - DLE (Display Line Enable)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemState Saved NoRange On | OffInitial S/W Revision Prior to A.02.00RefreshWhen pressed

Página 277 - DLYSWP (Delay Sweep)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystem:SYSTem:ERRor:OVERload OFF command. In either case, Input Overloads always set the

Página 278 - DONE (Done)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathSystem, ShowMode AllExample SYST:SHOW SYSTBackwards CompatibilityNotesThe ha

Página 279 - DR (Display Read)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemNotes dedicated Show Hardware Statistics screen in the Service Menu.Initial S/W Revi

Página 280 - E1[one] (Peak Marker)

3 Programming the AnalyzerRemote Measurement Functionsan error. Sending ASCII data when a definite block is expected generates message –161 "Inv

Página 281

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemShow System contents (Remote Command Only)A remote command is available to obtain th

Página 282

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemThe Previous Page is grayed-out if the first page of information is presently displa

Página 283

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemMode and Input/Output DefaultsWhen the analyzer is powered on in Mode and Input/Outp

Página 284 - EDITDONE (Edit Done)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemhappens behind the scenes when a controlled shutdown is requested by using the front

Página 285 - EDITLIML (Edit Limit Line)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemTDSCDMA | VSA | VSA89601 | WCDMA | WIMAXOFDMA:SYSTem:PON:MODE?Example SYST:PON:MODE

Página 286 - ERR (Error)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemPreloading enables you to “preload” at startup, to eliminate the runtime delay. Pr

Página 287

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemYELLOW: the applications you have selected can all fit into the analyzer’s memory,

Página 288

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathSystem, Power On, Configure ApplicationsInitial S/W Revision A.02.00Select/D

Página 289 - ET (Elapsed Time)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemInitial S/W Revision A.02.00Modified at S/WRevisionA.04.00Exit Without SavingPressin

Página 290

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemConfigure Applications - Remote CommandsThe following topics provide details on usin

Página 291 - FA (Start Frequency)

3 Programming the AnalyzerRemote Measurement Functionsl NORMal order begins with the most significant byte (MSB) first, and ends with the leastsignif

Página 292

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemConfiguration Memory Total (Remote Command Only)Queries the limit of Virtual Memory

Página 293 - FB (Stop Frequency)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathSystemInitial S/W Revision Prior to A.02.00Auto AlignConfigures the method f

Página 294

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemNormalAuto Align, Normal turns on the automatic alignment of all measurement systems

Página 295 - FDSP (Frequency Display Off)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemThe Auto Align, Alert mechanism will notify you when alignments have expired. One so

Página 296 - FOFFSET (Frequency Offset)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemWhen Auto Align, Off is selected the Settings Panel indicates ALIGN OFF with a warni

Página 297

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemAlertThe instrument will signal an Alert when conditions exist such that you will ne

Página 298 - FPKA (Fast Preselector Peak)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystem24 hoursWith Auto Align Alert set to 24 Hours the instrument will signal an alert af

Página 299 - FREF (Frequency Reference)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemNo confirmation is required when Alert is configured through a remote command.Key Pa

Página 300 - FS (Full Span)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemNo confirmation is required when Alert is configured through a remote command.Key Pa

Página 301

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathSystem, AlignmentsInitial S/W Revision Prior to A.02.00AllImmediately execut

Página 302

3 Programming the AnalyzerSTATus SubsystemSTATus SubsystemThe following diagram shows the entire Status Register Subsystem implementation of the XSe

Página 303 - GATE (Gate)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemattenuator and Electronic attenuator alignments on the RF Preselector path. The pu

Página 304 - GATECTL (Gate Control)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemSee additional remarks described with :CALibration[:ALL]?Everything about :CALibrati

Página 305 - GD (Gate Delay)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystem:CALibration:NRF? returns 1 if failedWhile Align Now, All but RF is performing the a

Página 306 - GL (Gate Length)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathSystem, Alignments, Align NowMode AllRemote Command:CALibration:RF:CALibrati

Página 307 - GP (Gate Polarity)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathSystem, Alignments, Align NowMode AllRemote Command:CALibration:EMIXer:CALib

Página 308 - GRAT (Graticule)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemA successful Align Now, RF will set the Last Align RF temperature to the current tem

Página 309 - HD (Hold Data Entry)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathVisual annotation in the Show Alignment Statistics screenMode AllRemote Comm

Página 310

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemExample :CAL:TIME:LRF?Notes Value is the elapsed time, in seconds, since the last su

Página 311

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathVisual annotation in the Show Alignment Statistics screenMode AllRemote Comm

Página 312

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathVisual annotation in the Show Alignment Statistics screenMode AllRemote Comm

Página 313

3 Programming the AnalyzerSTATus SubsystemWhat Are Status RegistersThe status system contains multiple registers that are arranged in a hierarchical

Página 314 - ID (Identify)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathVisual annotation in the Show Alignment Statistics screenMode AllRemote Comm

Página 315 - IP (Instrument Preset)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemParameter SettingTimebase DAC CalibratedTimebase DAC setting Calibrated valueAuto Al

Página 316 - Resolution)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathSystem, AlignmentsMode AllRemote Command:CALibration:DATA:DEFaultExample :CA

Página 317 - KS( (Lock Registers)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemThe backup screen indicates the approximate amount of space required to contain the

Página 318 - KS, (Mixer Level)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemChanging the drive letter also modifies the path displayed in the dialog below. When

Página 319 - KS) (Unlock Registers)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemRemote Language Compatibility Measurement Application Reference 1285

Página 320 - KSA (Amplitude in dBm)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemThe restore operation checks the validity of the restore file using the database&apo

Página 321 - KSa (Normal Detection)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemChanging the drive letter also modifies the path displayed in the box below. When th

Página 322 - KSB (Amplitude in dBmV)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystem1288 Remote Language Compatibility Measurement Application Reference

Página 323 - KSb (Positive Peak Detection)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemPerform Backup (Remote Command Only)Invokes an alignment data backup operation to th

Página 324 - KSC (Amplitude in dBμV)

3 Programming the AnalyzerSTATus SubsystemWhat Are Status Register SCPI CommandsMost monitoring of the instrument conditions is done at the highest l

Página 325 - KSc (A Plus B to A)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemRemote Command:CALibration:DATA:RESTore <filename>Example :CAL:DATA:REST "

Página 326 - KSD (Amplitude in Volts)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemThe last Characterize Preselector Time and Temperature survives across the power cyc

Página 327 - KSd (Negative Peak Detection)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemIf the value of the Timebase DAC changes (by switching to Calibrated from User with

Página 328 - KSE (Title Mode)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemMode AllExample :CAL:FREQ:REF:MODE USERReadback Text xxx < where xxx is the

Página 329 - KSe (Sample Detection)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemI/O ConfigActivates a menu for identifying and changing the I/O configuration for re

Página 330 - KSG (Video Averaging On)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemenabled, the instrument can run software applications that use the instrument's

Página 331 - KSg (Display Off)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemExample :SYST:COMM:GPIB:CONT ON Will set GPIB port to ControllerInitial S/W

Página 332 - KSH (Video Averaging Off)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemRemote Command:SYSTem:COMMunicate:LAN:SCPI:SOCKet:ENABle OFF|ON|0|1:SYSTem:COMMunica

Página 333 - KSh (Display On)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemState Saved NoRange On | OffInitial S/W Revision Prior to A.02.00HiSLIP ServerTurns

Página 334

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemMode AllRemote Command:SYSTem:COMMunicate:LAN:SCPI:SOCKet:CONTrol?Example :SYST:COMM

Página 335 - KSj (View Trace C)

Table Of ContentsSyntax 283Legacy Products 283Description 283EDITDONE (Edit Done) 284Syntax 284Legacy Products 284Description 284EDITLIML (Edit Limit

Página 336 - KSK (Marker to Next Peak)

3 Programming the AnalyzerSTATus Subsystem• you are monitoring more than one device which supports SRQs• you need to have the controller do something

Página 337 - KSk (Blank Trace C)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemLXIOpens a menu that allows you to access the various LXI configuration properties.K

Página 338 - KSL (Marker Noise Off)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemFactoryThis key selects the factory setting, for example:“Agilent Technologies,N9020

Página 339

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemThis is reserved for future use. The T+M488 protocol is not supported on low speedco

Página 340 - KSM (Marker Noise On)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemRemote Command:SYSTem:COMMunicate:USB:PACKets?Example :SYST:COMM:USB:PACK?Notes Two

Página 341 - KSm (Graticule Off)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathSystem, Restore System DefaultsExample :SYST:DEF INPInitial S/W Revision Pri

Página 342 - KSN (Marker Minimum)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemRestore Align DefaultsThis selection causes the Alignment system settings to be a re

Página 343 - KSn (Graticule On)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemMiscellaneous Setting Default ValueGPIB Address 18Auto File Name Number 000Save Type

Página 344 - KSO (Marker Span)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemRestore Mode Defaults (All Modes)This selection resets all of the modes in the instr

Página 345 - KSo (Annotation Off)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathSystem, Restore System DefaultsExample :SYST:DEF ALLCouplings An All will ca

Página 346 - KSp (Annotation On)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemOr use Alt-Tab: press and hold the Alt key and press and release the Tab key until

Página 347 - KST (Fast Preset)

3 Programming the AnalyzerSTATus SubsystemUsing a Status RegisterEach bit in a register is represented by a numerical value based on its location. Se

Página 348 - KSV (Frequency Offset)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathSystemNotes No equivalent remote command for this key.Backwards Compatibilit

Página 349 - KSx (External Trigger)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemRemote Command:SYSTem:LKEY:LIST?NotesReturn Value:An <arbitrary block data> of

Página 350 - KSy (Video Trigger)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemNotes Return value is the host ID as a stringInitial S/W Revision Prior to A.02.00Se

Página 351 - KSZ (Reference Level Offset)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemRead-WriteSelection for allowing full read-write access to the USB ports.Key PathSys

Página 352 - L0 [zero] (Display Line Off)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemThe CXA models in which the AC/DC Switch field is called Fixed Atten and which omit

Página 353 - LF (Low Frequency Preset)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSysteml "Query the Mechanical Relay Cycle Count" on page 1315l "Query the O

Página 354 - LG (Logarithmic Scale)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemMode AllRemote Command:SYSTem:TEMPerature:HEXTreme?Example :SYST:TEMP:HEXT?Notes Val

Página 355

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemKey PathSystemModeAllNotes No equivalent remote command for this key.Initial S/W Rev

Página 356 - LIMIFAIL (Limits Failed)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemRemote Command:SYSTem:OPTions?Example :SYST:OPT?Notes The return string is a comma s

Página 357 - Query Data Type Codes

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemRemote Command:SYSTem:HELP:HEADers?Example :SYST:HELP:HEAD?Notes The output is an IE

Página 358

3 Programming the AnalyzerSTATus Subsystem4. So now the only output from the Status Questionable Integrity register will come from a bit 10positive t

Página 359

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSystemTime (Remote Command Only)Sets or queries the time in the instrument.Mode AllRemote

Página 360 - LIML (Lower-Limit Amplitude)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorTrace/DetectorThe Trace/Detector menu lets you control the acquisition, disp

Página 361 - LIMTSL (Slope Limit Line)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorState Saved The type of each trace is saved in instrument stateInitial S/W R

Página 362 - LIMU (Upper-Limit Amplitude)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorState Saved The state of Average is saved in Instrument State for ghosting p

Página 363 - LN (Linear Scale)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detector2. The trace is updated in a series of discrete steps, with measurement data

Página 364 - M1 [one] (Marker Off)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detectordetector for each trace, but the analyzer hardware provides the maximum flex

Página 365 - M2 [two] (Marker Normal)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorLDIF =Log DifferenceSelect TraceDetermines which trace the type control k

Página 366

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorSelect TraceDetermines which trace the type control keys will affect. When y

Página 367 - M3 [three] (Delta Marker)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorSelect TraceDetermines which trace the type control keys will affect. When y

Página 368

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorBackwards CompatibilityNotesPrevious to the X-Series, pressing Clear Write w

Página 369 - M4 [four] (Marker Zoom)

3 Programming the AnalyzerSTATus Subsystem• The instrument does not have a service request pending. (A service request is considered to bepending bet

Página 370

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detector• The average/hold count k is set to 1, so that the next time the average tr

Página 371 - MA (Marker Amplitude Output)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorWhenever you press Max Hold or send the equivalent SCPI command, Update is s

Página 372

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorRemember that restarting Min Hold also restarts Max Hold and averaging, as t

Página 373

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorView/BlankThis key lets you set the state of the two trace variables, Update

Página 374 - MDS (Measurement Data Size)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detectorthe trace in Trace On (Update On and Display On), even if that detector was

Página 375 - MEAN (Trace Mean)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorBackwardsCompatibility SCPI:TRACe:MODE BLANksets :TRACe:UPDate OFF, :TRACe:D

Página 376

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detector• View: Update Off and Display On• Blank: Update Off and Display Off• Backgr

Página 377 - MEAS (Meas)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorTrace Update State On/OffKey PathTrace/DetectorRemote Command:TRACe[1]|2|...

Página 378 - MF (Marker Frequency Output)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorMore InformationWhen a trace becomes inactive, the following things happen:•

Página 379 - MINH (Minimum Hold)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detectortype letter in the trace annotation panel in the Measurement bar. So in the

Página 380 - MINPOS (Minimum X Position)

3 Programming the AnalyzerSTATus SubsystemThe Status Byte RegisterThe RQS bit is read and reset by a serial poll. The same bit position (MSS) is read

Página 381 - MKA (Marker Amplitude)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detector:TRACe[1]|2|...6:UPDate[:STATe]?Example TRAC2:UPD 0 Makes trace 2 inactive (

Página 382 - MKACT (Activate Marker)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorInactive traces display across the entire X Axis of the instrument. Their ho

Página 383 - MKBW (Marker Bandwidth)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorSee "Trace Update State On/Off" on page 1342.See "Trace Displ

Página 384

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detectorand Blank were trace modes, set by TRACe:MODE command. In the X-Series, Vie

Página 385 - MKD (Marker Delta)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detector• if the trace is the target of a Copy or participant in an Exchange• if the

Página 386

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detectorpossible combinations:• Trace On: Update and Display both On• View: Update O

Página 387 - MKF (Marker Frequency)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorTrace Display State On/OffKey PathTrace/DetectorRemote Command:TRACe[1]|2|..

Página 388

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorTraces which are blanked (Display=off) do not display nor appear on printout

Página 389 - MKFC (Marker Counter)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorPOS =Positive peakSAMP =SampleNEG

Página 390 - 8566A/B, 8568A/B:

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detector[:SENSe]:DETector[:FUNCtion]?Example DET AVER Sets detector to average for a

Página 391

3 Programming the AnalyzerSTATus SubsystemBit Description0, 1 These bits are always set to 0.2 A 1 in this bit position indicates that the SCPI error

Página 392 - MKMIN (Marker Minimum)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detectorfrom the selected CISPR detector to an auto coupled detector according to th

Página 393 - MKN (Marker Normal)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorWhen the Detector choice is Auto, the detector selected depends on marker fu

Página 394

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorRemote Command[:SENSe]:DETector:AUTO ON | OFF | 1 | 0[:SENSe]:DETector:AUTO?

Página 395 - MKNOISE (Marker Noise)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detector• Log power (also known as video)• Power (also known as RMS)• Voltage envelo

Página 396 - MKOFF (Marker Off)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorPeak detection is used for CW measurements and some pulsed-RF measurements.

Página 397 - MKP (Marker Position)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorNegative PeakFor each bucket (interval) in the trace, Negative Peak detectio

Página 398 - MKPK (Marker Peak)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detectortraces for which this detector is selected (other than to slow down the swee

Página 399

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorNote that CISPR standard operation is to perform the envelope averaging on t

Página 400 - MKPT (Marker Threshold)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorExample DET:TRAC3 RAV Selects the RMS Average detector for trace 3.Notes Thi

Página 401 - MKPX (Marker Peak Excursion)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorAll Traces AutoThis is designed to quickly return the selected set of detect

Página 402 - MKREAD (Marker Readout)

3 Programming the AnalyzerSTATus SubsystemIn addition to the status byte register, the status byte group also contains the service requestenable regi

Página 403

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorKey PathTrace/DetectorRemote Command:TRACe:CLEar TRACE1|TRACE2|TRACE3|TRACE4

Página 404

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detectorfunction, the indicated function is performed during the sweep with the math

Página 405 - MKSP (Marker Span)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorCouplings Whenever a math function is turned on for a trace, that trace is s

Página 406 - MKSS (Marker to Step Size)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorRemote Language Compatibility Measurement Application Reference 1363

Página 407 - MKT (Marker Time)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorFor each active trace, the current trace point is processed for Trace 1, the

Página 408 - MKTRACE (Marker Trace)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorSelect TraceDetermines which trace the type control keys will affect. When y

Página 409 - MKTRACK (Marker Track)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorSelect TraceDetermines which trace the type control keys will affect. When y

Página 410 - MKTYPE (Marker Type)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorKey PathTrace/Detector, MathExample :CALC:MATH TRACE1,PDIF,TRACE4,TRACE5,,Se

Página 411 - ML (Mixer Level)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorDuring the sweep, the following formula is executed for each point in the tr

Página 412

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorThe values of the operand trace points are assumed to be in decibel units (a

Página 413 - MT0 [zero] (Marker Track Off)

3 Programming the AnalyzerSTATus SubsystemStandard Event Status RegisterThe standard event status register contains the following bits:Remote Languag

Página 414 - MT1 [one] (Marker Track On)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorNotes See Trace “Math”.State Saved The current trace math function is saved

Página 415 - MXMH (Maximum Hold)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorKey PathTrace/Detector, Math, Trace OperandsDependencies The First Trace can

Página 416

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorOperand 1Selects the first trace operand to be used for the trace math funct

Página 417

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorOperand 2Selects the second trace operand to be used for the trace math func

Página 418

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorReadback Trace <trace number>Initial S/W Revision Prior to A.02.00Oper

Página 419

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorPreset Trace number – 1 (wraps at 1). For example, for Trace 1, the Second T

Página 420

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorDependencies When Signal ID is on, this key is grayed out.PresetTRACE1, TRA

Página 421

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorFrom TraceSelects the trace to be copied to or exchanged with the To TraceKe

Página 422

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorFrom TraceSelects the trace to be copied to or exchanged with the To TraceKe

Página 423

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorTo TraceSelects the trace to be copied from or exchanged with the From Trace

Página 424

3 Programming the AnalyzerSTATus SubsystemBit Description0 A 1 in this bit position indicates that all pending operations were completedfollowing exe

Página 425

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorTo TraceSelects the trace to be copied from or exchanged with the From Trace

Página 426 - OL (Output Learn String)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorNormalize On/OffNormalize (On) activates the normalize function. On each swe

Página 427

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detector(as before). When Normalize is turned on, the result is the calibrated retu

Página 428

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/Detectortrace query results. In other words, both trace query result and marker Y be

Página 429 - OT (Output Trace Annotations)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorExample TRAC3:DISP 1 Shows the reference trace.Notes Use the TRAC3:DISP comm

Página 430

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorOpen/Short Guided CalOn pressing the Open/Short Cal softkey in the Normalize

Página 431 - PEAKS (Peaks)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorKey PathTrace/Detector, NormalizeRemote Command:DISPlay:WINDow[1]:TRACe:Y[:S

Página 432 - PKPOS (Peak Position)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorSend/Query Trace Data (Remote Command Only)This command allows trace data to

Página 433 - PLOT (Plot)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorMore InformationThe format and byte-ordering of the sent or received data de

Página 434 - PP (Preselector Peak)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorSmoothes the trace according to the number of points specified in :TRACe:MAT

Página 435 - PRINT (Print)

3 Programming the AnalyzerSTATus Subsystemnumber 7 and bit number 3 are enabled, the decimal sum of the 2 bits is 128 plus 8. So thedecimal value 136

Página 436 - PWRBW (Power Bandwidth)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorRefer to the figure above for a discussion of this end-point smoothing pheno

Página 437

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrace/DetectorMean Trace Data (Remote Command Only)Included for ESA compatibility. Not rec

Página 438

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerTriggerAccesses a menu of keys to control the selection of the trigger source and t

Página 439 - R2 [two] (End-of-Sweep SRQ)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerOther trigger-related commands are found in the INITiate and ABORt SCPI commandsubs

Página 440

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerMeas Mode Preset for RF Preset for IQ NotesSwept SA SA IMM IQ notsupportedCHP SA, W

Página 441

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerCMMB, ISDB-T,Digital Cable TV,MSRTx Power SA, GSM, TD-SCDMASA, GSM:RFBurstTD-SCDMA:

Página 442 - RB (Resolution Bandwidth)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerRFBurstAll others:IMMediateGSM/EDGE:IQMagAll others:IMMMediatePVT WIMAXOFDMA RFB IM

Página 443

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerPowerLE In-bandEmissionsBLUETOOTH IMM IQ notsupportedEDR In-bandSpuriousEmissionsBL

Página 444

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerfront panel. In some models, there is no second External input. In these models,

Página 445 - RC (Recall State)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrigger−IINPut - triggers on the I channel's ADC voltage−QINPut - triggers on the Q c

Página 446

Table Of ContentsLegacy Products 296Description 296FPKA (Fast Preselector Peak) 298Syntax 298Legacy Products 298Description 298FREF (Frequency Referen

Página 447 - RCLS (Recall State)

3 Programming the AnalyzerSTATus SubsystemThe bit is valid for most X-Series Modes.5 Waiting for trigger The instrument is waiting for the trigger co

Página 448 - REV (Revision)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerThe following examples show trigger setup parameters using an external trigger sour

Página 449 - RL (Reference Level)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerevent occurs and all the applicable trigger criteria have been met). A correspondin

Página 450

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerTrigger LevelSets a level for the video signal trigger. When the video signal cross

Página 451 - RMS (Root Mean Square Value)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerTrig SlopeControls the trigger polarity. It is set positive to trigger on a rising

Página 452

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerRemote Command:TRIGger[:SEQuence]:VIDeo:DELay <time>:TRIGger[:SEQuence]:VIDeo

Página 453

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerRemote Command:TRIGger[:SEQuence]:OFFSet <time>:TRIGger[:SEQuence]:OFFSet?:TR

Página 454

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerTrig SlopeControls the trigger polarity. It is set positive to trigger on a rising

Página 455

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerDefault Unit SBackwards CompatibilityNotesThe legacy :TRIGger[:SEQuence]:DELay comm

Página 456 - S1[one] (Continuous Sweep)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerExample TRIG:EXT1:LEV 0.4 VCouplings This same level is used for the Ext1 trigger s

Página 457 - S2 [two] (Single Sweep)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerTrig DelayControls a time delay during which the analyzer will wait to begin a swee

Página 458 - SAVES (Save State)

3 Programming the AnalyzerSTATus Subsystem"Operation Event Query" on page 142"Operation Negative Transition" on page 142"Ope

Página 459 - SER (Serial Number)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerKey PathTriggerExample TRIG:SOUR EXT2 Swept SA measurementTRIG:<

Página 460 - SETDATE (Set Date)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerCompatibility SCPIInitial S/W Revision Prior to A.02.00Trig SlopeControls the trigg

Página 461 - SETTIME (Set Time)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerNotes Video trigger delay may be set to negative values, in time domain, FFT and ev

Página 462 - SMOOTH (Smooth Trace)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerBackwards CompatibilityNotesThe legacy command::TRIGger[:SEQuence]:RFBurst:FSELecti

Página 463 - SNGLS (Single Sweep)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerDefault Unit depends on the current selected Y-Axis unitBackwardsCompatibility SCPI

Página 464 - SP (Frequency Span)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrigger3. If the new absolute RF Burst level differs from the previous by more than 0.5 dB

Página 465

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerRemote Command:TRIGger[:SEQuence]:RFBurst:SLOPe POSitive|NEGative:TRIGger[:SEQuence

Página 466

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerBackwards CompatibilityNotesThe legacy :TRIGger[:SEQuence]:DELay command affects th

Página 467 - SRQ (Service Request)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerfrequencies. For example, we might want to measure spurious transmissions at an off

Página 468

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerto OFF, and the external sync source rate is changed for some reason, the periodic

Página 469

3 Programming the AnalyzerSTATus SubsystemdependenciesInitial S/W Revision Prior to A.02.00Operation Event QueryThis query returns the decimal value

Página 470 - ST (Sweep Time)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerRemote Command:TRIGger[:SEQuence]:FRAMe:OFFSet <time>:TRIGger[:SEQuence]:FRAM

Página 471

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerRemote Command:TRIGger[:SEQuence]:FRAMe:ADJust <time>Example TRIG:FRAM:ADJ 1.

Página 472 - STB (Status Byte Query)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerSync SourceSelects a signal source for you to synchronize your periodic timer trigg

Página 473

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerExternal 1Pressing this key, when it is not selected, selects the external input po

Página 474 - SUM (Sum)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerTrig SlopeControls the trigger polarity. It is set positive to trigger on a rising

Página 475 - SV (Save State)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerTrigger LevelSets the value where the external 2 trigger input will trigger a new s

Página 476 - SWPCPL (Sweep Couple)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerRF BurstPressing this key, when it is not selected, selects the RF Burst as the tri

Página 477 - T1 [one] (Free Run Trigger)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerScope Meas GlobalRemote Command:TRIGger[:SEQuence]:RFBurst:LEVel:ABSolute <ampl&

Página 478 - T2 [two] (Line Trigger)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerModified at S/WRevisionA.04.00Relative Trigger LevelSets the relative trigger level

Página 479 - T3 [three] (External Trigger)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerGSM: –25 dBState Saved Saved in instrument stateMin –45 dBMax 0 dBDefault Unit dB

Página 480 - T4 [four] (Video Trigger)

3 Programming the AnalyzerSTATus SubsystemOperation Positive TransitionThis command determines which bits in the Operation Condition register will se

Página 481 - TA (Trace A)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerKey PathTrigger, Periodic TimerRemote Command:TRIGger[:SEQuence]:FRAMe:DELay <ti

Página 482 - TB (Trace B)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerField Two, lines 314 to 625As the Field is changed, the appropriate value for Line

Página 483 - TDF (Trace Data Format)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerwithin the TV trigger setup functions.Preset 17State Saved Saved in instrument stat

Página 484 - TH (Threshold)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerField OneWhen you select Field One it causes the selected line number to be viewed

Página 485

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerAs the TV standard is changed, the current line value is clipped as necessary to ke

Página 486 - THE (Threshold Enable)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerNTSC–4.43Sets the TV standard to NTSC–4.43.Key PathTrigger, TV, StandardExample TRI

Página 487 - TIMEDATE (Time Date)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerPAL-B,D,G,H,ISets the TV standard to PAL-B,D,G,H,IKey PathTrigger, TV, StandardExam

Página 488 - TITLE (Title)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTrigger• If Holdoff On and Type = Normal, readback value• If Holdoff On and Type = Above,

Página 489 - TM (Trigger Mode)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceTriggerKey PathTrigger, Auto/HoldoffRemote Command:TRIGger[:SEQuence]:HOLDoff <time>

Página 490

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceUser PresetUser PresetAccesses a menu that gives you the following three choices:• User Pr

Página 491

3 Programming the AnalyzerSTATus Subsystem"Questionable Negative Transition " on page 145"Questionable Positive Transition" on pa

Página 492

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceUser PresetThe User Preset function does the following:• Aborts the currently running meas

Página 493

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceUser Preset• Clears the input and output buffers.• Sets the Status Byte to 0.Key PathUser

Página 494 - TRDSP (Trace Display)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayView/DisplayThe View/Display key opens the Display Menu (common to most measur

Página 495 - TRIGPOL (Trigger Polarity)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayExample :DISP:VIEW ZSP sets the zone span viewDependencies All views except

Página 496 - TRSTAT (Trace State)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplaySee the figure below. Each type of annotation can be turned on and off individ

Página 497 - TS (Take Sweep)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayKey PathView/Display, Display, AnnotationRemote Command:DISPlay:ANNotation:MBA

Página 498 - VAVG (Video Average)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayKey PathView/Display, Display, AnnotationRemote Command:DISPlay:ANNotation:TRA

Página 499

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayKey PathView/Display, Display, AnnotationRemote Command:DISPlay:ACTivefunc[:ST

Página 500

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayExample DISP:ANN:TITL:DATA "This Is My Title"This example is for the

Página 501 - VB (Video Bandwidth)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayKey PathView/Display, DisplayRemote Command:DISPlay:WINDow[1]:TRACe:GRATicule:

Página 502

3 Programming the AnalyzerSTATus SubsystemdependenciesInitial S/W Revision Prior to A.02.00Questionable Event QueryThis query returns the decimal val

Página 503

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayMin –∞ (minus infinity) in current unitsMax +∞ (plus infinity) in current unit

Página 504

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayAnnotation Local SettingsThis is a Mode Global override of the meas local anno

Página 505 - VIEW (View Trace)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayThemeThis key allows you to change the Display theme. This is similar to the T

Página 506

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayFlat ColorSelects a format that is best when the screen is to be printed on an

Página 507 - VTL (Video Trigger Level)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayKey PathView/Display, Display, System Display Settings, BacklightExample DISP:

Página 508

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/Display• If, while in External Mixing, the Mixer Preset changes (for example, if you

Página 509 - XCH (Exchange)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/Displayput into Display Trace 1, and all the other traces “roll up.” Once the trace d

Página 510

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayNote that since the spectrogram is intended to give a view of spectral behavio

Página 511 - Panel &SCPI Reference

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayRepresentation of TimeIn the Spectrogram view, zero time is the point where th

Página 512

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/Displayinaccuracies within each trace are therefore reconciled with the start of the

Página 513

3 Programming the AnalyzerSTATus SubsystemQuestionable Positive TransitionThis command determines which bits in the Questionable Condition register w

Página 514

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayThe zoom region is indicated by a blue shading over the whole region. In the

Página 515

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayTransition RulesWhen you enter the Trace Zoom view, the top window of Trace Zo

Página 516 - AMPTD Y Scale

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayMore InformationIn Zone Span, the window with the focus (the selected window)

Página 517 - Attenuation

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayMeasurement Swept SARemote Command:TRACe:DISPlay:VIEW:SPECtrogram:POSition <

Página 518

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayBackwards CompatibilityNotesIn ESA/PSA, Full Screen was turned on with a softk

Página 519 - (Mech) Atten

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/DisplayNotesInitial S/W Revision Prior to A.02.00Remote Language Compatibility Measur

Página 520

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceView/Display1466 Remote Language Compatibility Measurement Application Reference

Página 521 - Enable Elec Atten

3 Programming the AnalyzerSTATus SubsystemMode AllRemote Command:STATus:QUEStionable:CALibration:CONDition?Example STAT:QUES:CAL:COND?Preset 0Status

Página 522

3 Programming the AnalyzerSTATus SubsystemExample STAT:QUES:CAL?Preset 0Status Bits/OPC dependencies Sequential commandInitial S/W Revision Prior to

Página 523

3 Programming the AnalyzerSTATus SubsystemPreset 32767Min 0Max 32767Status Bits/OPC dependencies Sequential commandInitial S/W Revision Prior to A.02

Página 524 - Elec Atten

Table Of ContentsDescription 308HD (Hold Data Entry) 309Syntax 309Legacy Products 309Description 309I1 [one] (Set RF Coupling to DC) 310Syntax 310Lega

Página 525 - Pre-Adjust for Min Clip

3 Programming the AnalyzerSTATus SubsystemMode AllRemote Command:STATus:QUEStionable:CALibration:SKIPped:ENABle <integer>:STATus:QUEStionable:C

Página 526 - COMBined

3 Programming the AnalyzerSTATus Subsystem:STATus:QUEStionable:CALibration:SKIPped:NTRansition?Example STAT:QUES:CAL:SKIP:NTR 1Align RF skipped is no

Página 527 - (Mech) Atten Step

3 Programming the AnalyzerSTATus Subsystem"Questionable Calibration Extended Failure Positive Transition" on page 154Questionable Calibrati

Página 528 - Max Mixer Level

3 Programming the AnalyzerSTATus SubsystemQuestionable Calibration Extended Failure Event QueryThis query returns the decimal value of the sum of the

Página 529 - Scale Type

3 Programming the AnalyzerSTATus SubsystemQuestionable Calibration Extended Failure Positive TransitionThis command determines which bits in the Ques

Página 530 - Presel Center

3 Programming the AnalyzerSTATus SubsystemRemote Command:STATus:QUEStionable:CALibration:EXTended:NEEDed:CONDition?Example STAT:QUES:CAL:EXT:NEED:CON

Página 531 - Preselector Adjust

3 Programming the AnalyzerSTATus SubsystemPreset 0Status Bits/OPC dependencies Sequential commandInitial S/W Revision Prior to A.02.00Questionable Ca

Página 532

3 Programming the AnalyzerSTATus SubsystemPreset 32767Min 0Max 32767Status Bits/OPCdependenciesSequential commandInitial S/WRevisionPrior to A.02.00Q

Página 533 - Y Axis Unit

3 Programming the AnalyzerSTATus Subsystem5) in the Questionable Register. The variable <integer> is the sum of the decimal values of the bitsy

Página 534

3 Programming the AnalyzerSTATus Subsystemhas a negative transition (1 to 0). The variable <integer> is the sum of the decimal values of thebit

Página 535

Table Of ContentsKSA (Amplitude in dBm) 320Syntax 320Legacy Products 320Description 320KSa (Normal Detection) 321Syntax 321Legacy Products 321Descript

Página 536

3 Programming the AnalyzerSTATus SubsystemQuestionable Integrity RegisterThe following commands and queries are available for this register:"Que

Página 537 - Antenna Unit

3 Programming the AnalyzerSTATus SubsystemPreset 32767Min 0Max 32767StatusBits/OPCdependenciesSequential commandInitial S/WRevisionPrior to A.02.00Qu

Página 538

3 Programming the AnalyzerSTATus SubsystemQuestionable register.Preset 0Min 0Max 32767StatusBits/OPCdependenciesSequential commandInitial S/WRevision

Página 539

3 Programming the AnalyzerSTATus Subsystem"Questionable Integrity Signal Event Query" on page 164"Questionable Integrity Signal Negati

Página 540 - Reference Level Offset

3 Programming the AnalyzerSTATus SubsystemQuestionable Integrity Signal Event QueryThis query returns the decimal value of the sum of the bits in the

Página 541

3 Programming the AnalyzerSTATus Subsystemregister bit has a positive transition (0 to 1). The variable <integer> is the sum of the decimalvalu

Página 542 - [:SENSe]:POWer[:RF]:MW:PATH?

3 Programming the AnalyzerSTATus SubsystemPreset 0Status Bits/OPC dependencies Sequential commandInitial S/W Revision Prior to A.02.00Questionable In

Página 543 - Standard Path

3 Programming the AnalyzerSTATus SubsystemPreset 0Status Bits/OPC dependencies Sequential commandInitial S/W Revision Prior to A.02.00Questionable In

Página 544 - Internal Preamp

3 Programming the AnalyzerSTATus Subsystemregister.Preset 32767Min 0Max 32767StatusBits/OPCdependenciesSequential commandInitial S/WRevisionPrior to

Página 545 - Full Range

3 Programming the AnalyzerSTATus SubsystemQuestionable Power EnableThis command determines which bits in the Questionable Power Condition Register wi

Página 546

Table Of ContentsSyntax 329Legacy Products 329Description 329KSG (Video Averaging On) 330Syntax 330Legacy Products 330Description 330KSg (Display Off)

Página 547 - Auto Couple

3 Programming the AnalyzerSTATus SubsystemQuestionable Power Negative TransitionThis command determines which bits in the Questionable Power Conditio

Página 548 - Auto/Man 1-of-N keys

3 Programming the AnalyzerSTATus SubsystemMax 32767StatusBits/OPCdependenciesSequential commandInitial S/WRevisionPrior to A.02.00Questionable Temper

Página 549

3 Programming the AnalyzerSTATus SubsystemMode AllRemoteCommand:STATus:QUEStionable:TEMPerature:ENABle <integer>:STATus:QUEStionable:TEMPeratur

Página 550

3 Programming the AnalyzerSTATus SubsystemMode AllRemoteCommand:STATus:QUEStionable:TEMPerature:NTRansition <integer>:STATus:QUEStionable:TEMPe

Página 551 - Video BW

3 Programming the AnalyzerSTATus Subsystem174 Remote Language Compatibility Measurement Application Reference

Página 552

4 List of Legacy Analyzer CommandsThe following table ("Alphanumeric List of all LegacyCommands with N9061A Support" on page 176) lists all

Página 553 - VBW:3dB RBW

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportAlphanumeric List of all Legacy Commands with N9061A Su

Página 554 - Auto Rules

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionACPE

Página 555 - Span:3dB RBW

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionACTV

Página 556 - RBW Control

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series Descriptionfunc

Página 557 - Filter Type

Table Of ContentsLegacy Products 338Description 338KSl (Transfer Trace B to Trace C) 339Syntax 339Legacy Products 339Description 339KSM (Marker Noise

Página 558 - Filter BW

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionCA C

Página 559

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series Descriptionline

Página 560

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionDA D

Página 561

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series Descriptionfreq

Página 562

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionFFTC

Página 563 - :INITiate:CONTinuous?

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionGDRV

Página 564

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionHN H

Página 565 - File Explorer

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionKS)

Página 566 - Page Setup

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionKSC

Página 567 - :HCOPy:ABORt

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series Descriptionfunc

Página 568 - Minimize

Table Of ContentsDescription 347KSV (Frequency Offset) 348Syntax 348Legacy Products 348Description 348KSx (External Trigger) 349Syntax 349Legacy Produ

Página 569

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionHP85

Página 570 - FREQ Channel

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionLIMI

Página 571 - Zoom Center

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series Descriptioninte

Página 572 - Auto Tune

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionMERG

Página 573 - Center Freq

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionMKMI

Página 574 - Center Frequency Presets

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series Descriptionmemo

Página 575 - Input 2:

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionO3 O

Página 576 - RF Center Freq

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionOT O

Página 577 - Ext Mix Center Freq

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionPREA

Página 578 - I/Q Center Freq

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series Descriptionrefe

Página 579 - Start Freq

Notices© Agilent Technologies, Inc., 2008-2013No part of this manual may bereproduced in any form or by anymeans (including electronic storageand retr

Página 580

Table Of ContentsQuery Data Type Codes 357LIMIPURGE (Delete Current Limit Line) 358Syntax 358Legacy Products 358Description 358LIMIREL (Relative Limit

Página 581 - Stop Freq

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series Descriptionfrom

Página 582

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionSRCP

Página 583

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionSYNC

Página 584

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionTOIR

Página 585

4 List of Legacy Analyzer CommandsAlphanumeric List of all Legacy Commands with N9061A SupportCommand Description 8566 85688560Series DescriptionUSTA

Página 586

5 Legacy Command DescriptionsThis chapter lists all the supported 8560Series, 8566A/B and8568A/B commands in alphanumeric order, and gives briefdetai

Página 587 - Input/Output

5 Legacy Command DescriptionsCommand SyntaxCommand SyntaxCommand syntax is represented pictorially.l Ovals enclose command mnemonics. The command mne

Página 588 - :INPut:MIXer?

5 Legacy Command DescriptionsCommand SyntaxSyntaxComponent Definition/RangeAnalyzercommandAny command in this chapter, with required parameters and t

Página 589 - RF Input

5 Legacy Command DescriptionsCommand DescriptionsCommand DescriptionsAll supported commands are listed here, with descriptions and cross-references t

Página 590 - Input Z Correction

5 Legacy Command DescriptionsA1 [one] (Clear Write for Trace A)A1 [one] (Clear Write for Trace A)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDe

Página 591 - RF Coupling

Table Of ContentsM4 [four] (Marker Zoom) 369Syntax 369Legacy Products 369Description 369MA (Marker Amplitude Output) 371Syntax 371Legacy Products 371D

Página 592 - Baseband I/Q (Option BBA)

5 Legacy Command DescriptionsA2 [two] (Maximum Hold for Trace A)A2 [two] (Maximum Hold for Trace A)SyntaxLegacy Products8560 series, 8566A/B, 8568A/B

Página 593

5 Legacy Command DescriptionsA3 [three] (View Mode for Trace A)A3 [three] (View Mode for Trace A)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDe

Página 594

5 Legacy Command DescriptionsA4 [four] (Blank Trace A)A4 [four] (Blank Trace A)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionBlanks Tr

Página 595

5 Legacy Command DescriptionsACPALPHA (Adjacent Channel Power Alpha Weighting)ACPALPHA (Adjacent Channel Power Alpha Weighting)SyntaxLegacy Products8

Página 596 - I/Q Path

5 Legacy Command DescriptionsACPALTCH (Adjacent Channel Power Alternate Channels)ACPALTCH (Adjacent Channel Power Alternate Channels)SyntaxLegacy Pro

Página 597

5 Legacy Command DescriptionsACPBRPER (Adjacent Channel Power Burst Period)ACPBRPER (Adjacent Channel Power Burst Period)SyntaxLegacy Products8560 se

Página 598 - I Differential Input

5 Legacy Command DescriptionsACPBRWID (Adjacent Channel Power Burst Width)ACPBRWID (Adjacent Channel Power Burst Width)SyntaxLegacy Products8560 seri

Página 599 - :INPut[1]:IQ[:I]:IMPedance?

5 Legacy Command DescriptionsACPBW (Adjacent Channel Power Bandwidth)ACPBW (Adjacent Channel Power Bandwidth)SyntaxLegacy Products8560 seriesDescript

Página 600

5 Legacy Command DescriptionsACPCOMPUTE (Adjacent Channel Power Compute)ACPCOMPUTE (Adjacent Channel Power Compute)SyntaxLegacy Products8560 seriesDe

Página 601

5 Legacy Command DescriptionsACPFRQWT (Adjacent Channel Power Frequency Weighting)ACPFRQWT (Adjacent Channel Power Frequency Weighting)SyntaxLegacy P

Página 602 - :INPut:IMPedance:IQ?

Table Of ContentsSyntax 379Legacy Products 379Description 379MINPOS (Minimum X Position) 380Syntax 380Legacy Products 380Description 380MKA (Marker Am

Página 603

5 Legacy Command DescriptionsACPLOWER (Lower Adjacent Channel Power)ACPLOWER (Lower Adjacent Channel Power)SyntaxLegacy Products8560 seriesDescriptio

Página 604 - :INPut:IQ:Q:DIFFerential?

5 Legacy Command DescriptionsACPMAX (Maximum Adjacent Channel Power)ACPMAX (Maximum Adjacent Channel Power)SyntaxDescriptionReturns the maximum adjac

Página 605 - :INPut[1]:IQ:Q:IMPedance?

5 Legacy Command DescriptionsACPMEAS (Measure Adjacent Channel Power)ACPMEAS (Measure Adjacent Channel Power)SyntaxLegacy Products8560 seriesDescript

Página 606

5 Legacy Command DescriptionsACPMSTATE (Adjacent Channel Power Measurement State)ACPMSTATE (Adjacent Channel Power Measurement State)SyntaxLegacy Pro

Página 607

5 Legacy Command DescriptionsACPMSTATE (Adjacent Channel Power Measurement State)Query Data Type CURR|DFLTSCPI Equivalent Commands NonePreset Default

Página 608 - :CALibration:IQ:PROBe:Q

5 Legacy Command DescriptionsACPPWRTX (Adjacent Channel Power Total Power Transmitted)ACPPWRTX (Adjacent Channel Power Total Power Transmitted)Syntax

Página 609 - I/Q Cable Calibrate…

5 Legacy Command DescriptionsACPRSLTS (Adjacent Channel Power Measurement Results)ACPRSLTS (Adjacent Channel Power Measurement Results)SyntaxLegacy P

Página 610 - :CALibration:IQ:ISOLation

5 Legacy Command DescriptionsQuery Data Type DetailsQuery Data Type DetailsAlternateChannelsChannels used for CalculationNumber of Values Returned0l

Página 611 - I/Q Probe Setup

5 Legacy Command DescriptionsACPSP (Adjacent Channel Power Channel Spacing)ACPSP (Adjacent Channel Power Channel Spacing)SyntaxLegacy Products8560 se

Página 612

5 Legacy Command DescriptionsACPT (Adjacent Channel Power T Weighting)ACPT (Adjacent Channel Power T Weighting)SyntaxLegacy Products8560 seriesDescri

Página 613 - :INPut:OFFSet:I

Table Of ContentsLegacy Products 391Description 391MKMIN (Marker Minimum) 392Syntax 392Legacy Products 392Description 392MKN (Marker Normal) 393Syntax

Página 614 - :INPut:COUPling:I

5 Legacy Command DescriptionsACPUPPER (Upper Adjacent Channel Power)ACPUPPER (Upper Adjacent Channel Power)SyntaxLegacy Products8560 seriesDescriptio

Página 615

5 Legacy Command DescriptionsADJALL (LO and IF Adjustments)ADJALL (LO and IF Adjustments)SyntaxLegacy Products8560 seriesDescriptionActivates the RF

Página 616 - RF Calibrator

5 Legacy Command DescriptionsAMB (A minus B into A)AMB (A minus B into A)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSubtracts the p

Página 617 - :CALibration:SOURce:STATe?

5 Legacy Command DescriptionsAMBPL (A minus B plus Display Line into A)AMBPL (A minus B plus Display Line into A)SyntaxLegacy Products8560 series, 85

Página 618 - External Gain

5 Legacy Command DescriptionsANNOT (Annotation)ANNOT (Annotation)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionTurns on or off all ann

Página 619 - Ext Preamp

5 Legacy Command DescriptionsAPB (Trace A Plus Trace B to A)APB (Trace A Plus Trace B to A)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescript

Página 620

5 Legacy Command DescriptionsAT (Input Attenuation)AT (Input Attenuation)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSpecifies the R

Página 621

5 Legacy Command DescriptionsAT (Input Attenuation)"(Mech) Atten Step" on page 527)[:SENSE]:POWer[:RF]:ATTenuation:AUTO (OFF|ON|0|1) (see &

Página 622 - I Ext Gain

5 Legacy Command DescriptionsAUNITS (Absolute Amplitude Units)AUNITS (Absolute Amplitude Units)Syntax8560 series8566A/B, 8568A/BLegacy Products8560 s

Página 623 - Q Ext Gain

5 Legacy Command DescriptionsAUNITS (Absolute Amplitude Units)If your selected remote language is any of the 8560 Series analyzers, and if the AUNITS

Página 624 - Corrections

Table Of ContentsDescription 402MKRL (Marker to Reference Level) 404Syntax 404Legacy Products 404Description 404MKSP (Marker Span) 405Syntax 405Legacy

Página 625 - Select Correction

5 Legacy Command DescriptionsAUTOCPL (Auto Coupled)AUTOCPL (Auto Coupled)SyntaxLegacy Products8560 seriesDescriptionSets video bandwidth, resolution

Página 626

5 Legacy Command DescriptionsAXB (Exchange Trace A and Trace B)AXB (Exchange Trace A and Trace B)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDe

Página 627 - Correction On/Off

5 Legacy Command DescriptionsB1 [one] (Clear Write for Trace B)B1 [one] (Clear Write for Trace B)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDe

Página 628 - Properties

5 Legacy Command DescriptionsB2 [two] (Maximum Hold for Trace B)B2 [two] (Maximum Hold for Trace B)SyntaxLegacy Products8560 series, 8566A/B, 8568A/B

Página 629

5 Legacy Command DescriptionsB3 [three] (View Mode for Trace B)B3 [three] (View Mode for Trace B)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDe

Página 630

5 Legacy Command DescriptionsB4 [four] (Blank Trace B)B4 [four] (Blank Trace B)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionBlanks Tr

Página 631 - UA

5 Legacy Command DescriptionsBL (Trace B minus Display Line to Trace B)BL (Trace B minus Display Line to Trace B)SyntaxLegacy Products8566A/B, 8568A/

Página 632

5 Legacy Command DescriptionsBLANK (Blank Trace)BLANK (Blank Trace)Syntax8560 Series:8566A/B, 8568A/B:Legacy Products8560 series, 8566A/B, 8568A/BDes

Página 633

5 Legacy Command DescriptionsBLANK (Blank Trace)(See "View/Blank " on page 1344)Preset TRB, TRC Blank.Notes The functions of BLANK are iden

Página 634

5 Legacy Command DescriptionsBML (Trace B Minus Display Line)BML (Trace B Minus Display Line)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescri

Página 635

Table Of ContentsMT1 [one] (Marker Track On) 414Syntax 414Legacy Products 414Description 414MXMH (Maximum Hold) 415Syntax 415Legacy Products 415Descri

Página 636

5 Legacy Command DescriptionsBTC (Transfer Trace B to Trace C)BTC (Transfer Trace B to Trace C)SyntaxLegacy Products8566A/B, 8568A/BDescriptionTransf

Página 637 - Apply Corrections

5 Legacy Command DescriptionsBXC (Exchange Trace B and Trace C)BXC (Exchange Trace B and Trace C)SyntaxLegacy Products8566A/B, 8568A/BDescriptionExch

Página 638 - Delete All Corrections

5 Legacy Command DescriptionsC1 [one] (Set A Minus B Mode Off)C1 [one] (Set A Minus B Mode Off)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDesc

Página 639 - Freq Ref In

5 Legacy Command DescriptionsC2 [two] (A Minus B Into A)C2 [two] (A Minus B Into A)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSubtr

Página 640 - [:SENSe]:ROSCillator:SOURce?

5 Legacy Command DescriptionsCA (Couple Attenuation)CA (Couple Attenuation)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionDuring normal

Página 641 - Ext Ref Freq

5 Legacy Command DescriptionsCARROFF (Carrier Off Power)CARROFF (Carrier Off Power)SyntaxLegacy Products8560 seriesDescriptionMeasures the average an

Página 642 - External Reference Lock BW

5 Legacy Command DescriptionsCARRON (Carrier On Power)CARRON (Carrier On Power)SyntaxLegacy Products8560 seriesDescriptionMeasures the average power

Página 643 - External Ref Coupling

5 Legacy Command DescriptionsCF (Center Frequency)CF (Center Frequency)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSpecifies the val

Página 644

5 Legacy Command DescriptionsCF (Center Frequency)Default unit is HZ.CF UPCF DNStep size: see Description above.CF OACF?Query Data Type <freq>

Página 645

5 Legacy Command DescriptionsCHANNEL (Channel Selection)CHANNEL (Channel Selection)SyntaxLegacy Products8560 seriesDescriptionIncrements or decrement

Página 646

Table Of ContentsLegacy Products 424Description 424OCCUP (Percent Occupied Power Bandwidth) 425Syntax 425Legacy Products 425Description 425OL (Output

Página 647

5 Legacy Command DescriptionsCHANPWR (Channel Power)CHANPWR (Channel Power)SyntaxLegacy Products8560 seriesDescriptionMeasures the power within the s

Página 648

5 Legacy Command DescriptionsCHPWRBW (Channel Power Bandwidth)CHPWRBW (Channel Power Bandwidth)SyntaxLegacy Products8560 seriesDescriptionQueries or

Página 649 - Output Config

5 Legacy Command DescriptionsCLRAVG (Clear Average)CLRAVG (Clear Average)SyntaxLegacy Products8566A/B, 8568A/BDescriptionRestarts the VAVG command by

Página 650 - NEGative

5 Legacy Command DescriptionsCLRW (Clear Write)CLRW (Clear Write)Syntax8560 series8566A/B, 8568A/BLegacy Products8560 series, 8566A/B, 8568A/BDescrip

Página 651

5 Legacy Command DescriptionsCLRW (Clear Write)The following table describes the parameters set by N9061A in the X-series instrument when thelegacy c

Página 652 - Trig Out (1 and 2)

5 Legacy Command DescriptionsCONTS (Continuous Sweep)CONTS (Continuous Sweep)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSets the in

Página 653

5 Legacy Command DescriptionsCOUPLE (Input Coupling)COUPLE (Input Coupling)SyntaxLegacy Products8560 seriesDescriptionSelects AC or DC coupling.Forma

Página 654

5 Legacy Command DescriptionsCR (Couple Resolution Bandwidth)CR (Couple Resolution Bandwidth)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescri

Página 655 - Analog Out

5 Legacy Command DescriptionsCS (Couple Frequency Step Size)CS (Couple Frequency Step Size)SyntaxLegacy Products8566A/B, 8568A/BDescriptionCouples th

Página 656 - OUTPut:ANALog:AUTO OFF

5 Legacy Command DescriptionsCT (Couple Sweep Time)CT (Couple Sweep Time)SyntaxLegacy Products8566A/B, 8568A/BDescriptionCouples the sweep time to th

Página 657 - OUTPut:ANALog:AUTO?

Table Of ContentsDescription 436R1 [one] (Illegal Command SRQ) 438Syntax 438Legacy Products 438Description 438R2 [two] (End-of-Sweep SRQ) 439Syntax 43

Página 658

5 Legacy Command DescriptionsCV (Couple Video Bandwidth)CV (Couple Video Bandwidth)SyntaxLegacy Products8566A/B, 8568A/BDescriptionCouples the video

Página 659

5 Legacy Command DescriptionsDA (Display Address)DA (Display Address)SyntaxLegacy Products8566A/B, 8568A/BDescriptionReturns the contents of the give

Página 660

5 Legacy Command DescriptionsDELMKBW (Occupied Power Bandwidth Within Delta Marker)DELMKBW (Occupied Power Bandwidth Within Delta Marker)SyntaxLegacy

Página 661 - I/Q Cal Out

5 Legacy Command DescriptionsDET (Detection Mode)DET (Detection Mode)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSelects the type of

Página 662 - :OUTPut:IQ:OUTPut?

5 Legacy Command DescriptionsDL (Display Line)DL (Display Line)Syntax8560 series:8566A/B, 8568A/B:Legacy Products8560 series, 8566A/B, 8568A/B274 Rem

Página 663 - Aux IF Out

5 Legacy Command DescriptionsDL (Display Line)DescriptionDefines the level of the display line and displays it on the instrument screen.Format DL <

Página 664

5 Legacy Command DescriptionsDLE (Display Line Enable)DLE (Display Line Enable)SyntaxLegacy Products8566A/B, 8568A/BDescriptionEnables or disables th

Página 665 - :OUTPut:AUX:AIF?

5 Legacy Command DescriptionsDLYSWP (Delay Sweep)DLYSWP (Delay Sweep)SyntaxLegacy Products8560 seriesDescriptionDelays the start of the sweep until t

Página 666 - I/Q Guided Calibration

5 Legacy Command DescriptionsDONE (Done)DONE (Done)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionAllows you to determine when the inst

Página 667

5 Legacy Command DescriptionsDR (Display Read)DR (Display Read)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSends the contents of the current disp

Página 668 - Marker Control Mode

Table Of ContentsRL (Reference Level) 449Syntax 449Legacy Products 449Description 450RMS (Root Mean Square Value) 451Syntax 451Legacy Products 451Desc

Página 669

5 Legacy Command DescriptionsE1[one] (Peak Marker)E1[one] (Peak Marker)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionPositions the mar

Página 670

5 Legacy Command DescriptionsE2 [two] (Marker to Center Frequency)E2 [two] (Marker to Center Frequency)SyntaxLegacy Products8560 series, 8566A/B, 856

Página 671

5 Legacy Command DescriptionsE3 [three] (Delta Marker Step Size)E3 [three] (Delta Marker Step Size)SyntaxLegacy Products8560 series, 8566A/B, 8568A/B

Página 672

5 Legacy Command DescriptionsE4 [four] (Marker to Reference Level)E4 [four] (Marker to Reference Level)SyntaxLegacy Products8560 series, 8566A/B, 856

Página 673 - Select Marker

5 Legacy Command DescriptionsEDITDONE (Edit Done)EDITDONE (Edit Done)SyntaxLegacy Products8560 seriesDescriptionThis command is used at the completio

Página 674

5 Legacy Command DescriptionsEDITLIML (Edit Limit Line)EDITLIML (Edit Limit Line)SyntaxLegacy Products8560 seriesDescriptionThis command is used to i

Página 675

5 Legacy Command DescriptionsERR (Error)ERR (Error)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionThe ERR? query returns a list of thre

Página 676

5 Legacy Command DescriptionsERR (Error)X-Series Error Code Description8560 Series Error CodeDescription42 RF Alignment Failure 758 SYSTEM: Unknownsy

Página 677

5 Legacy Command DescriptionsERR (Error)X-Series Error Code Description8560 Series Error CodeDescription521 External Ref missing or out of range 905,

Página 678

5 Legacy Command DescriptionsET (Elapsed Time)ET (Elapsed Time)SyntaxLegacy Products8560 seriesDescriptionReturns to the controller the elapsed time

Página 679 - Fixed Marker Y Axis Value

Table Of ContentsSyntax 461Legacy Products 461Description 461SMOOTH (Smooth Trace) 462Syntax 462Legacy Products 462Description 462SNGLS (Single Sweep)

Página 680 - Fixed Marker Z Axis Value

5 Legacy Command DescriptionsEX (Exchange Trace A and Trace B)EX (Exchange Trace A and Trace B)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDesc

Página 681

5 Legacy Command DescriptionsFA (Start Frequency)FA (Start Frequency)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSpecifies the start

Página 682

5 Legacy Command DescriptionsFA (Start Frequency)function to the start frequency.FA?Query Data Type <real> in HZ.SCPI Equivalent Commands [:SEN

Página 683

5 Legacy Command DescriptionsFB (Stop Frequency)FB (Stop Frequency)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSpecifies the stop fr

Página 684

5 Legacy Command DescriptionsFB (Stop Frequency)FB?Query Data Type <real> in HZ.SCPI Equivalent Commands [:SENSE]:FREQuency:STOP <real> (

Página 685

5 Legacy Command DescriptionsFDSP (Frequency Display Off)FDSP (Frequency Display Off)SyntaxLegacy Products8560 seriesDescriptionTurns the frequency a

Página 686

5 Legacy Command DescriptionsFOFFSET (Frequency Offset)FOFFSET (Frequency Offset)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSelects

Página 687

5 Legacy Command DescriptionsFOFFSET (Frequency Offset)FOFFSET?Query Data Type <real>SCPI Equivalent Commands [:SENSE]:FREQuency:OFFSet <num

Página 688

5 Legacy Command DescriptionsFPKA (Fast Preselector Peak)FPKA (Fast Preselector Peak)SyntaxLegacy Products8566A/BDescriptionAutomatically adjusts the

Página 689 - Relative To

5 Legacy Command DescriptionsFREF (Frequency Reference)FREF (Frequency Reference)SyntaxLegacy Products8560 seriesDescriptionSpecifies whether an exte

Página 690

Table Of ContentsTable Of ContentsN9061A Remote Language Compatibility Measurement Application User's &Programmer'sReference 1Table Of

Página 691

Table Of ContentsLegacy Products 474Description 474SV (Save State) 475Syntax 475Legacy Products 475Description 475SWPCPL (Sweep Couple) 476Syntax 476L

Página 692

5 Legacy Command DescriptionsFS (Full Span)FS (Full Span)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionl 8560 series: Sets the frequen

Página 693

5 Legacy Command DescriptionsFS (Full Span)N9030A-503 N9030A-508 N9030A-513 N9030A-526Remote Language Frequency Range Frequency Range Frequency Range

Página 694 - X Axis Scale

5 Legacy Command DescriptionsFS (Full Span)N9010A-503 N9010A-507 N9010A-513 N9010A-526Remote Language Frequency Range Frequency Range Frequency Range

Página 695

5 Legacy Command DescriptionsGATE (Gate)GATE (Gate)SyntaxPreset State: GATE OFFLegacy Products8560 seriesDescriptionTurns the time-gating function on

Página 696

5 Legacy Command DescriptionsGATECTL (Gate Control)GATECTL (Gate Control)SyntaxLegacy Products8560 seriesDescriptionSelects between the edge and leve

Página 697 - Marker Trace

5 Legacy Command DescriptionsGD (Gate Delay)GD (Gate Delay)SyntaxLegacy Products8560 seriesDescriptionSets the delay time from when the gate trigger

Página 698

5 Legacy Command DescriptionsGL (Gate Length)GL (Gate Length)SyntaxLegacy Products8560 seriesDescriptionSets the length of time the time gate is turn

Página 699

5 Legacy Command DescriptionsGP (Gate Polarity)GP (Gate Polarity)SyntaxLegacy Products8560 seriesDescriptionSets the polarity (positive or negative)

Página 700

5 Legacy Command DescriptionsGRAT (Graticule)GRAT (Graticule)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionTurns the graticule on or o

Página 701

5 Legacy Command DescriptionsHD (Hold Data Entry)HD (Hold Data Entry)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionDisables data entry

Página 702

Table Of ContentsDescription 483TH (Threshold) 484Syntax 484Legacy Products 484Description 484THE (Threshold Enable) 486Syntax 486Legacy Products 486D

Página 703

5 Legacy Command DescriptionsI1 [one] (Set RF Coupling to DC)I1 [one] (Set RF Coupling to DC)SyntaxLegacy Products8568A/BDescriptionSets the RF coupl

Página 704

5 Legacy Command DescriptionsI1 [one] (Set RF Coupling to DC)DC Coupled Range AC Coupled RangeInstrument Model (N9020A) Min. Freq. Max. Freq. Min. Fr

Página 705

5 Legacy Command DescriptionsI2 [two] (Set RF Coupling to AC)I2 [two] (Set RF Coupling to AC)SyntaxLegacy Products8568A/BDescriptionSets the RF coupl

Página 706

5 Legacy Command DescriptionsI2 [two] (Set RF Coupling to AC)DC Coupled Range AC Coupled RangeInstrument Model (N9010A) Min. Freq. Max. Freq. Min. Fr

Página 707

5 Legacy Command DescriptionsID (Identify)ID (Identify)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionThe ID? query returns the current

Página 708

5 Legacy Command DescriptionsIP (Instrument Preset)IP (Instrument Preset)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionPerforms an ins

Página 709

5 Legacy Command DescriptionsKS= (8566A/B: Automatic Preselector Tracking, 8568A/B: Marker Counter Resolution)KS= (8566A/B: Automatic Preselector Tra

Página 710

5 Legacy Command DescriptionsKS( (Lock Registers)KS( (Lock Registers)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSecures the contents of state re

Página 711

5 Legacy Command DescriptionsKS, (Mixer Level)KS, (Mixer Level)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSpecifies the maximum signal level tha

Página 712

5 Legacy Command DescriptionsKS) (Unlock Registers)KS) (Unlock Registers)SyntaxLegacy Products8566A/B, 8568A/BDescriptionUnlocks the state registers,

Página 713

Table Of ContentsTRIGPOL (Trigger Polarity) 495Syntax 495Legacy Products 495Description 495TRSTAT (Trace State) 496Syntax 496Legacy Products 496Descri

Página 714

5 Legacy Command DescriptionsKSA (Amplitude in dBm)KSA (Amplitude in dBm)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSets the amplitude readout (

Página 715

5 Legacy Command DescriptionsKSa (Normal Detection)KSa (Normal Detection)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSelects normal input detecti

Página 716

5 Legacy Command DescriptionsKSB (Amplitude in dBmV)KSB (Amplitude in dBmV)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSets the amplitude readout

Página 717

5 Legacy Command DescriptionsKSb (Positive Peak Detection)KSb (Positive Peak Detection)SyntaxLegacy Products8566A/B, 8568A/BDescriptionEnables positi

Página 718

5 Legacy Command DescriptionsKSC (Amplitude in dBμV)KSC (Amplitude in dBμV)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSets the amplitude readout

Página 719

5 Legacy Command DescriptionsKSc (A Plus B to A)KSc (A Plus B to A)SyntaxLegacy Products8566A/B, 8568A/BDescriptionPerforms a point-by-point addition

Página 720

5 Legacy Command DescriptionsKSD (Amplitude in Volts)KSD (Amplitude in Volts)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSets the amplitude reado

Página 721 - Marker Table

5 Legacy Command DescriptionsKSd (Negative Peak Detection)KSd (Negative Peak Detection)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSelects negati

Página 722 - Marker Count

5 Legacy Command DescriptionsKSE (Title Mode)KSE (Title Mode)SyntaxLegacy Products8566A/B, 8568A/BDescriptionActivates the title mode, writing a mess

Página 723 - Query Count Value

5 Legacy Command DescriptionsKSe (Sample Detection)KSe (Sample Detection)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSelects sample input detecti

Página 724

Table Of ContentsSyntax 509Legacy Products 509Description 5096 RLC Swept SA Measurement Front-Panel &SCPI Reference 511AMPTD Y Scale 516Reference

Página 725

5 Legacy Command DescriptionsKSG (Video Averaging On)KSG (Video Averaging On)SyntaxLegacy Products8566A/B, 8568A/BDescriptionEnables video averaging.

Página 726 - Gate Time

5 Legacy Command DescriptionsKSg (Display Off)KSg (Display Off)SyntaxLegacy Products8566A/B, 8568A/BDescriptionTurns the instrument’s display Off.For

Página 727 - All Markers Off

5 Legacy Command DescriptionsKSH (Video Averaging Off)KSH (Video Averaging Off)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSwitches video averagi

Página 728

5 Legacy Command DescriptionsKSh (Display On)KSh (Display On)SyntaxLegacy Products8566A/B, 8568A/BDescriptionTurns the instrument’s display On.Format

Página 729 - Marker Function

5 Legacy Command DescriptionsKSi (Exchange Trace B and Trace C)KSi (Exchange Trace B and Trace C)SyntaxLegacy Products8566A/B, 8568A/BDescriptionExch

Página 730 - Interval Markers

5 Legacy Command DescriptionsKSj (View Trace C)KSj (View Trace C)SyntaxLegacy Products8566A/B, 8568A/BDescriptionDisplays Trace C.Format KSjQuery Dat

Página 731

5 Legacy Command DescriptionsKSK (Marker to Next Peak)KSK (Marker to Next Peak)SyntaxLegacy Products8566A/B, 8568A/BDescriptionIf there is a marker o

Página 732

5 Legacy Command DescriptionsKSk (Blank Trace C)KSk (Blank Trace C)SyntaxLegacy Products8566A/B, 8568A/BDescriptionBlanks Trace C.Format KSkQuery Dat

Página 733 - 2. Span Pair Compatibility

5 Legacy Command DescriptionsKSL (Marker Noise Off)KSL (Marker Noise Off)SyntaxLegacy Products8566A/B, 8568A/BDescriptionDisables the noise density f

Página 734 - 6. Offscreen Markers

5 Legacy Command DescriptionsKSl (Transfer Trace B to Trace C)KSl (Transfer Trace B to Trace C)SyntaxLegacy Products8566A/B, 8568A/BDescriptionTransf

Página 735 - 7. Direct Marker Positioning

Table Of ContentsAuto Tune 572Center Freq 573Start Freq 579Stop Freq 581CF Step 583Freq Offset 584Input/Output 587Input/Output variables - Preset beha

Página 736

5 Legacy Command DescriptionsKSM (Marker Noise On)KSM (Marker Noise On)SyntaxLegacy Products8566A/B, 8568A/BDescriptionDisplays the noise density at

Página 737

5 Legacy Command DescriptionsKSm (Graticule Off)KSm (Graticule Off)SyntaxLegacy Products8566A/B, 8568A/BDescriptionBlanks the graticule on the instru

Página 738

5 Legacy Command DescriptionsKSN (Marker Minimum)KSN (Marker Minimum)SyntaxLegacy Products8566A/B, 8568A/BDescriptionMoves the marker to the minimum

Página 739

5 Legacy Command DescriptionsKSn (Graticule On)KSn (Graticule On)SyntaxLegacy Products8566A/B, 8568A/BDescriptionTurns on the graticule on the instru

Página 740 - Marker Noise

5 Legacy Command DescriptionsKSO (Marker Span)KSO (Marker Span)SyntaxLegacy Products8566A/B, 8568A/BDescriptionThis command operates only when the de

Página 741 - Band/Interval Power

5 Legacy Command DescriptionsKSo (Annotation Off)KSo (Annotation Off)SyntaxLegacy Products8566A/B, 8568A/BDescriptionBlanks the annotation on the ins

Página 742 - Band/Interval Density

5 Legacy Command DescriptionsKSp (Annotation On)KSp (Annotation On)SyntaxLegacy Products8566A/B, 8568A/BDescriptionActivates the annotation on the in

Página 743

5 Legacy Command DescriptionsKST (Fast Preset)KST (Fast Preset)SyntaxLegacy Products8566A/B, 8568A/BDescriptionPerforms an instrument preset, setting

Página 744 - Band Adjust

5 Legacy Command DescriptionsKSV (Frequency Offset)KSV (Frequency Offset)SyntaxLegacy Products8566A/B, 8568A/BDescriptionAllows you to specify a valu

Página 745

5 Legacy Command DescriptionsKSx (External Trigger)KSx (External Trigger)SyntaxLegacy Products8566A/B, 8568A/BDescriptionActivates the normal externa

Página 746 - Band/Interval Left

Table Of ContentsFixed marker functions 730Interval Markers 730Band Function Backwards Compatibility 7311. Marker Mode compatibility 7332. Span Pair C

Página 747

5 Legacy Command DescriptionsKSy (Video Trigger)KSy (Video Trigger)SyntaxLegacy Products8566A/B, 8568A/BDescriptionActivates the normal video trigger

Página 748 - Band/Interval Right

5 Legacy Command DescriptionsKSZ (Reference Level Offset)KSZ (Reference Level Offset)SyntaxLegacy Products8566A/B, 8568A/BDescriptionOffsets all ampl

Página 749 - Band Span Auto/Man

5 Legacy Command DescriptionsL0 [zero] (Display Line Off)L0 [zero] (Display Line Off)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionDis

Página 750

5 Legacy Command DescriptionsLF (Low Frequency Preset)LF (Low Frequency Preset)SyntaxLegacy Products8566A/BDescriptionPerforms a low frequency preset

Página 751 - Measure at Marker

5 Legacy Command DescriptionsLG (Logarithmic Scale)LG (Logarithmic Scale)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSpecifies the a

Página 752

5 Legacy Command DescriptionsLIMF (Limit Line Frequency Value)LIMF (Limit Line Frequency Value)SyntaxLegacy Products8560 seriesDescriptionThis comman

Página 753

5 Legacy Command DescriptionsLIMIFAIL (Limits Failed)LIMIFAIL (Limits Failed)SyntaxLegacy Products8560 seriesDescriptionReturns a number between 0 an

Página 754 - Meas at Marker Window

5 Legacy Command DescriptionsQuery Data Type CodesQuery Data Type CodesResult Meaning0 The active trace passed both the upper and the lower limit tes

Página 755 - Detectors

5 Legacy Command DescriptionsLIMIPURGE (Delete Current Limit Line)LIMIPURGE (Delete Current Limit Line)SyntaxLegacy Products8560 seriesDescriptionDel

Página 756

5 Legacy Command DescriptionsLIMIREL (Relative Limit Lines)LIMIREL (Relative Limit Lines)SyntaxLegacy Products8560 seriesDescriptionSpecifies whether

Página 757

Table Of ContentsFFT IF Gain 842Analog Demod Tune & Listen 845Noise Source 853Meas Preset 859Mode 860More Information 861Spectrum Analyzer 862EMI

Página 758

5 Legacy Command DescriptionsLIML (Lower-Limit Amplitude)LIML (Lower-Limit Amplitude)SyntaxLegacy Products8560 seriesDescriptionAssigns the lower-lim

Página 759

5 Legacy Command DescriptionsLIMTSL (Slope Limit Line)LIMTSL (Slope Limit Line)SyntaxLegacy Products8560 seriesDescriptionMakes the selected limit-li

Página 760

5 Legacy Command DescriptionsLIMU (Upper-Limit Amplitude)LIMU (Upper-Limit Amplitude)SyntaxLegacy Products8560 seriesDescriptionAssigns the upper-lim

Página 761 - Detector 2

5 Legacy Command DescriptionsLN (Linear Scale)LN (Linear Scale)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionScales the amplitude (ver

Página 762

5 Legacy Command DescriptionsM1 [one] (Marker Off)M1 [one] (Marker Off)SyntaxLegacy Products8566A/B, 8568A/BDescriptionBlanks any markers showing on

Página 763

5 Legacy Command DescriptionsM2 [two] (Marker Normal)M2 [two] (Marker Normal)SyntaxLegacy Products8566A/B, 8568A/BDescriptionMoves the active marker

Página 764

5 Legacy Command DescriptionsM2 [two] (Marker Normal)Query Data Type <real>. See "MKF (Marker Frequency)" on page 387.SCPI Equivalent

Página 765

5 Legacy Command DescriptionsM3 [three] (Delta Marker)M3 [three] (Delta Marker)SyntaxLegacy Products8566A/B, 8568A/BDescriptionComputes the frequency

Página 766 - Detector 3

5 Legacy Command DescriptionsM3 [three] (Delta Marker)Format M3 <real>HZ|KHZ|MHZ|GHZ|KZ|MZ|GZ|S|MS|US|SCM3 UP|DNUP or DN increments 10% of span

Página 767

5 Legacy Command DescriptionsM4 [four] (Marker Zoom)M4 [four] (Marker Zoom)SyntaxLegacy Products8566A/B, 8568A/BDescriptionThis command increases or

Página 768

Table Of ContentsMode Preset 883How-To Preset 884Mode Setup 886HP8560 series, HP8566/68 886Cmd Error 887Logging 888Preferences 889Restore Mode Default

Página 769

5 Legacy Command DescriptionsM4 [four] (Marker Zoom)M4 OAThe OA option only returns the current value to the controller; it does not set the activefu

Página 770

5 Legacy Command DescriptionsMA (Marker Amplitude Output)MA (Marker Amplitude Output)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionRet

Página 771

5 Legacy Command DescriptionsMC0 [zero] (MarkerFrequencyCounterOff)MC0 [zero] (MarkerFrequencyCounterOff)SyntaxLegacy Products8568A/BDescriptio

Página 772

5 Legacy Command DescriptionsMC1 [one] (MarkerFrequencyCounterOn)MC1 [one] (MarkerFrequencyCounterOn)SyntaxLegacy Products8568A/BDescriptionTur

Página 773 - BW & Avg Type

5 Legacy Command DescriptionsMDS (Measurement Data Size)MDS (Measurement Data Size)SyntaxLegacy Products8566A/B, 8568A/BDescriptionFormats binary dat

Página 774 - :CALCulate:MAMarker:COUPling?

5 Legacy Command DescriptionsMEAN (Trace Mean)MEAN (Trace Mean)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionReturns the mean value of

Página 775

5 Legacy Command DescriptionsMEANPWR (Mean Power measurement)MEANPWR (Mean Power measurement)SyntaxLegacy Products8560 seriesDescriptionMeasures the

Página 776

5 Legacy Command DescriptionsMEAS (Meas)MEAS (Meas)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionReturns the current sweep status.l If

Página 777 - Center Presel On/Off

5 Legacy Command DescriptionsMF (Marker Frequency Output)MF (Marker Frequency Output)Syntax8560 series:8566A/B, 8568A/B:Legacy Products8560 series, 8

Página 778 - Marker To

5 Legacy Command DescriptionsMINH (Minimum Hold)MINH (Minimum Hold)SyntaxLegacy Products8560 seriesDescriptionUpdates the chosen trace with the minim

Página 779 - Mkr->Start

Table Of ContentsScreen Image 1113Mass Storage Catalog (Remote Command Only) 1119Mass Storage Change Directory (Remote Command Only) 1119Mass Storage

Página 780 - Mkr->Ref Lvl

5 Legacy Command DescriptionsMINPOS (Minimum X Position)MINPOS (Minimum X Position)SyntaxLegacy Products8566A/B, 8568A/BDescriptionReturns the X co-o

Página 781 - Mkr -> Zone Center

5 Legacy Command DescriptionsMKA (Marker Amplitude)MKA (Marker Amplitude)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionReturns the amp

Página 782 - Mkr ->Span

5 Legacy Command DescriptionsMKACT (Activate Marker)MKACT (Activate Marker)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSpecifies the active marke

Página 783

5 Legacy Command DescriptionsMKBW (Marker Bandwidth)MKBW (Marker Bandwidth)SyntaxLegacy Products8560 seriesDescriptionReturns the bandwidth at the sp

Página 784

5 Legacy Command DescriptionsMKCF (Marker to Center Frequency)MKCF (Marker to Center Frequency)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDesc

Página 785

5 Legacy Command DescriptionsMKD (Marker Delta)MKD (Marker Delta)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionComputes the frequency

Página 786

5 Legacy Command DescriptionsMKD (Marker Delta)Format MKD <real>HZ|KHZ|MHZ|GHZ|KZ|MZ|GZ|S|MS|US|SCMKD UP|DNUP or DN specifies 10% of the curren

Página 787

5 Legacy Command DescriptionsMKF (Marker Frequency)MKF (Marker Frequency)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSpecifies the f

Página 788 - Meas Setup

5 Legacy Command DescriptionsMKF (Marker Frequency)MKF?Query Data Type 8560 Series: The data is returned in ASCII format.For all other languages, the

Página 789 - Average Type

5 Legacy Command DescriptionsMKFC (Marker Counter)MKFC (Marker Counter)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionTurns on or off t

Página 790

Table Of ContentsPower On 1251Alignments 1260I/O Config 1294Restore Defaults 1303Control Panel… 1308Licensing… 1309Security 1312Diagnostics 1313Servic

Página 791 - Log-Pwr Avg (Video)

5 Legacy Command DescriptionsMKFCR (Marker Counter Resolution)MKFCR (Marker Counter Resolution)Syntax8560 series:8566A/B, 8568A/B:390 Remote Language

Página 792 - Pwr Avg (RMS)

5 Legacy Command DescriptionsMKFCR (Marker Counter Resolution)Legacy Products8560 series, 8566A/B, 8568A/BDescriptionSets the resolution of the marke

Página 793 - Select Limit

5 Legacy Command DescriptionsMKMIN (Marker Minimum)MKMIN (Marker Minimum)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionMoves the activ

Página 794

5 Legacy Command DescriptionsMKN (Marker Normal)MKN (Marker Normal)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionMoves the active mark

Página 795

5 Legacy Command DescriptionsMKN (Marker Normal)UP or DN specifies 10% of the current span.MKN OAMKN?Query Data Type See "MKF (Marker Frequency)

Página 796

5 Legacy Command DescriptionsMKNOISE (Marker Noise)MKNOISE (Marker Noise)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionDisplays the av

Página 797

5 Legacy Command DescriptionsMKOFF (Marker Off)MKOFF (Marker Off)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionTurns off either the ac

Página 798

5 Legacy Command DescriptionsMKP (Marker Position)MKP (Marker Position)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSpecifies the marker position

Página 799

5 Legacy Command DescriptionsMKPK (Marker Peak)MKPK (Marker Peak)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionExecuting MKPK HI, or s

Página 800

5 Legacy Command DescriptionsMKPK (Marker Peak)Peak)" on page 336, except that KSK does not take in to account the marker peakexcursion or marke

Página 801

Table Of ContentsUnits 62User-defined Functions 62Supported Commands 62EP Parameter 63OA Parameter 63Handling of Unsupported Commands and Queries 63Ha

Página 802

Table Of ContentsRF Trigger Source 1397I/Q Trigger Source 1398More Information 1399Free Run 1400Video (IF Envelope) 1401Line 1405External 1 1407Extern

Página 803

5 Legacy Command DescriptionsMKPT (Marker Threshold)MKPT (Marker Threshold)SyntaxLegacy Products8560 seriesDescriptionSets the minimum amplitude leve

Página 804

5 Legacy Command DescriptionsMKPX (Marker Peak Excursion)MKPX (Marker Peak Excursion)SyntaxPreset State: 6 dBLegacy Products8560 series, 8566A/B, 856

Página 805

5 Legacy Command DescriptionsMKREAD (Marker Readout)MKREAD (Marker Readout)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSelects the type of active

Página 806

5 Legacy Command DescriptionsMKREAD (Marker Readout)MKREADType Non-Zero SpanNon-Zero SpanDelta Zero Span Zero Span DeltaPER Period offrequency(Pulsem

Página 807

5 Legacy Command DescriptionsMKRL (Marker to Reference Level)MKRL (Marker to Reference Level)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescri

Página 808

5 Legacy Command DescriptionsMKSP (Marker Span)MKSP (Marker Span)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionThis command operates o

Página 809

5 Legacy Command DescriptionsMKSS (Marker to Step Size)MKSS (Marker to Step Size)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSets th

Página 810

5 Legacy Command DescriptionsMKT (Marker Time)MKT (Marker Time)SyntaxLegacy Products8560 seriesDescriptionPlaces a marker at a position that correspo

Página 811

5 Legacy Command DescriptionsMKTRACE (Marker Trace)MKTRACE (Marker Trace)SyntaxLegacy Products8566A/B, 8568A/BDescriptionMoves the active marker to t

Página 812 - LLINE5

5 Legacy Command DescriptionsMKTRACK (Marker Track)MKTRACK (Marker Track)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionMoves the signa

Página 813

1 About the AnalyzerThe X-Series signal analyzer measures and monitors complexRF and microwave signals. Analog baseband analysis isavailable on MXA. T

Página 814

5 Legacy Command DescriptionsMKTYPE (Marker Type)MKTYPE (Marker Type)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSpecifies the type of marker.Typ

Página 815 - TRACE5

5 Legacy Command DescriptionsML (Mixer Level)ML (Mixer Level)Syntax8560 series:8566A/B, 8568A/B:Legacy Products8560 series, 8566A/B, 8568A/BDescripti

Página 816

5 Legacy Command DescriptionsML (Mixer Level)l Mixer Level = Ref. Level - Attenuation + Ext. Amplifier GainThe external amplifier gain is not preset

Página 817

5 Legacy Command DescriptionsMT0 [zero] (Marker Track Off)MT0 [zero] (Marker Track Off)SyntaxLegacy Products8566A/B, 8568A/BDescriptionDisables the m

Página 818

5 Legacy Command DescriptionsMT1 [one] (Marker Track On)MT1 [one] (Marker Track On)SyntaxLegacy Products8566A/B, 8568A/BDescriptionMoves the signal o

Página 819

5 Legacy Command DescriptionsMXMH (Maximum Hold)MXMH (Maximum Hold)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionUpdates each trace el

Página 820

5 Legacy Command DescriptionsNORMLIZE (Normalize Trace Data)NORMLIZE (Normalize Trace Data)SyntaxLegacy Products8560 seriesDescriptionActivates or de

Página 821 - Test Limits

5 Legacy Command DescriptionsNORMLIZE (Normalize Trace Data)(See "Normalize On/Off" on page 1381)Preset OFFCouplings NORMLIZE sets Trace B

Página 822 - X-Axis Unit

5 Legacy Command DescriptionsNRL (Normalized Reference Level)NRL (Normalized Reference Level)SyntaxLegacy Products8560 seriesDescriptionSets the norm

Página 823 - Delete All Limits

5 Legacy Command DescriptionsNRPOS (Normalized Reference Position)NRPOS (Normalized Reference Position)SyntaxLegacy Products8560 seriesDescriptionAdj

Página 824 - <connect>}

1 About the AnalyzerInstalling Application SoftwareInstalling Application SoftwareIf you want to install a measurement application after your initial

Página 825

5 Legacy Command DescriptionsO1 [one] (Format - Display Units)O1 [one] (Format - Display Units)SyntaxLegacy Products8566A/B, 8568A/BDescriptionTransm

Página 826

5 Legacy Command DescriptionsO2 [two] (Format - Two 8-Bit Bytes)O2 [two] (Format - Two 8-Bit Bytes)SyntaxLegacy Products8566A/B, 8568A/BDescriptionTr

Página 827

5 Legacy Command DescriptionsO3 [three] (Format - Real Amplitude Units)O3 [three] (Format - Real Amplitude Units)SyntaxLegacy Products8566A/B, 8568A/

Página 828

5 Legacy Command DescriptionsO4 [four] (Format - One 8-Bit Byte)O4 [four] (Format - One 8-Bit Byte)SyntaxLegacy Products8566A/B, 8568A/BDescriptionTr

Página 829 - N dB Points

5 Legacy Command DescriptionsOA or ? (Query Active Function)OA or ? (Query Active Function)Legacy Products8566A/B, 8568A/BDescriptionQuery active fun

Página 830 - N dB Points Results Query

5 Legacy Command DescriptionsOCCUP (Percent Occupied Power Bandwidth)OCCUP (Percent Occupied Power Bandwidth)SyntaxLegacy Products8560 seriesDescript

Página 831

5 Legacy Command DescriptionsOL (Output Learn String)OL (Output Learn String)SyntaxLegacy Products8566A/B, 8568A/BDescriptionTransmits information to

Página 832

5 Legacy Command DescriptionsOL (Output Learn String)Byte Support Information20 Supported: Trigger Mode, Sweep Mode, TRB Clearwrite statusUnsupported

Página 833 - PhNoise Opt

5 Legacy Command DescriptionsOL (Output Learn String)Format OL <80-byte string>OL?Query Data Type See table above.SCPI Equivalent Commands None

Página 834

5 Legacy Command DescriptionsOT (Output Trace Annotations)OT (Output Trace Annotations)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSends 32 chara

Página 835

1 About the AnalyzerInstalling Application SoftwareInstalling a license key for the selected application can be done automatically using a USBmemory

Página 836 - Best Wide-offset Φ Noise

5 Legacy Command DescriptionsOT (Output Trace Annotations)Index Content20 "PL1 UNLOCK"21 "PL2 UNLOCK"22 "Y-I-0 UNLOCK"2

Página 837 - Fast Tuning

5 Legacy Command DescriptionsPEAKS (Peaks)PEAKS (Peaks)Syntax8560 series:8566A/B, 8568A/B:Legacy Products8560 series, 8566A/B, 8568A/BDescriptionSort

Página 838 - ADC Dither

5 Legacy Command DescriptionsPKPOS (Peak Position)PKPOS (Peak Position)SyntaxLegacy Products8568DescriptionReturns the X co-ordinate value of the max

Página 839

5 Legacy Command DescriptionsPLOT (Plot)PLOT (Plot)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionAllows you transfer trace data, grati

Página 840 - Swept IF Gain

5 Legacy Command DescriptionsPP (Preselector Peak)PP (Preselector Peak)SyntaxLegacy Products8560 series, 8566A/BDescriptionOptimizes preselector trac

Página 841

5 Legacy Command DescriptionsPRINT (Print)PRINT (Print)SyntaxLegacy Products8560 seriesDescriptionTransfers trace data, graticule and annotation of t

Página 842 - FFT IF Gain

5 Legacy Command DescriptionsPWRBW (Power Bandwidth)PWRBW (Power Bandwidth)Syntax8560 Series:8566A/B, 8568A/B:Legacy Products8560 series, 8566A/B, 85

Página 843

5 Legacy Command DescriptionsPWRBW (Power Bandwidth)SCPI Equivalent Commands NoneNotes If the percent total power is 100%, the power bandwidth equals

Página 844 - High Gain (Best Noise Level)

5 Legacy Command DescriptionsR1 [one] (Illegal Command SRQ)R1 [one] (Illegal Command SRQ)SyntaxLegacy Products8566A/B, 8568A/BDescriptionDeactivates

Página 845 - [:SENSe]:DEMod?

5 Legacy Command DescriptionsR2 [two] (End-of-Sweep SRQ)R2 [two] (End-of-Sweep SRQ)SyntaxLegacy Products8566A/B, 8568A/BDescriptionActivates the end-

Página 846 - [:SENSe]:DEMod:STATe ON

1 About the AnalyzerX-Series Options and AccessoriesX-Series Options and AccessoriesYou can view an online list of available Options and Accessories

Página 847

5 Legacy Command DescriptionsR3 [three] (Hardware Broken SRQ)R3 [three] (Hardware Broken SRQ)SyntaxLegacy Products8566A/B, 8568A/BDescriptionActivate

Página 848 - [:SENSe]:DEMod:FM:DEEMphasis?

5 Legacy Command DescriptionsR4 [four] (Units-Key-Pressed SRQ)R4 [four] (Units-Key-Pressed SRQ)SyntaxLegacy Products8566A/B, 8568A/BDescriptionActiva

Página 849

5 Legacy Command DescriptionsRB (Resolution Bandwidth)RB (Resolution Bandwidth)Syntax8560 series:8566A/B, 8568A/B:Legacy Products8560 series, 8566A/B

Página 850

5 Legacy Command DescriptionsRB (Resolution Bandwidth)bandwidths, and sweep time are normally coupled, but executing RB decouples them. Execute "

Página 851

5 Legacy Command DescriptionsRBR (Resolution Bandwidth to Span Ratio)RBR (Resolution Bandwidth to Span Ratio)SyntaxLegacy Products8560 seriesDescript

Página 852 - Demod Time

5 Legacy Command DescriptionsRC (Recall State)RC (Recall State)Syntax8560 series:8566A/B, 8568A/B:Legacy Products8560 series, 8566A/B, 8568A/BDescrip

Página 853 - Noise Source

5 Legacy Command DescriptionsRC (Recall State)See Description above.Query Data Type N/ASCPI Equivalent Commands *RCL <integer> (see "Recal

Página 854

5 Legacy Command DescriptionsRCLS (Recall State)RCLS (Recall State)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionRecalls instrument st

Página 855

5 Legacy Command DescriptionsREV (Revision)REV (Revision)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionReturns the firmware revision n

Página 856 - :SOURce:NOISe:TYPE?

5 Legacy Command DescriptionsRL (Reference Level)RL (Reference Level)Syntax8560 series:8566A/B, 8568A/B:Legacy Products8560 series, 8566A/B, 8568A/BR

Página 857

1 About the AnalyzerFront & Rear Panel FeaturesFront & Rear Panel FeaturesThe instrument's front and rear panel features are fully detai

Página 858 - :SOURce:NOISe[:STATe]?

5 Legacy Command DescriptionsRL (Reference Level)DescriptionSpecifies the amplitude level of the top graticule line on the display. This represents t

Página 859 - Meas Preset

5 Legacy Command DescriptionsRMS (Root Mean Square Value)RMS (Root Mean Square Value)SyntaxLegacy Products8566A/B, 8568A/BDescriptionReturns the RMS

Página 860

5 Legacy Command DescriptionsROFFSET (Reference Level Offset)ROFFSET (Reference Level Offset)Syntax8560 series:8566A/B, 8568A/B:Legacy Products8560 s

Página 861

5 Legacy Command DescriptionsROFFSET (Reference Level Offset)UP or DN increments one vertical divisionROFFSET OAROFFSET?Query Data Type <real>

Página 862 - Spectrum Analyzer

5 Legacy Command DescriptionsRQS (Request Service Conditions)RQS (Request Service Conditions)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescri

Página 863 - IQ Analyzer (Basic)

5 Legacy Command DescriptionsRQS (Request Service Conditions)Format RQS <bit number>RQS OARQS?Query Data Type The current bit mask.SCPI Equival

Página 864 - 802.16 OFDMA (WiMAX/WiBro)

5 Legacy Command DescriptionsS1[one] (Continuous Sweep)S1[one] (Continuous Sweep)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSets the instrument

Página 865 - Vector Signal Analyzer (VXA)

5 Legacy Command DescriptionsS2 [two] (Single Sweep)S2 [two] (Single Sweep)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSets the instrument to sin

Página 866 - Analog Demod

5 Legacy Command DescriptionsSAVES (Save State)SAVES (Save State)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSaves the current state

Página 867 - TD-SCDMA with HSPA/8PSK

5 Legacy Command DescriptionsSER (Serial Number)SER (Serial Number)SyntaxLegacy Products8560 seriesDescriptionReturns the X-series instrument serial

Página 868

1 About the AnalyzerDisplay AnnotationsDisplay AnnotationsDisplay Annotations are fully detailed under the chapter "Front and Rear Panel Feature

Página 869 - DTMB (CTTB)

5 Legacy Command DescriptionsSETDATE (Set Date)SETDATE (Set Date)SyntaxLegacy Products8560 seriesDescriptionSets the date of the real-time clock of t

Página 870 - Combined WLAN

5 Legacy Command DescriptionsSETTIME (Set Time)SETTIME (Set Time)SyntaxLegacy Products8560 seriesDescriptionSets the date of the real-time clock of t

Página 871 - 802.16 OFDM (Fixed WiMAX)

5 Legacy Command DescriptionsSMOOTH (Smooth Trace)SMOOTH (Smooth Trace)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSmooths the trace according to

Página 872 - 89601 VSA

5 Legacy Command DescriptionsSNGLS (Single Sweep)SNGLS (Single Sweep)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSets the instrument

Página 873 - SCPI Language Compatibility

5 Legacy Command DescriptionsSP (Frequency Span)SP (Frequency Span)Syntax8560 series:8566A/B, 8568A/B:464 Remote Language Compatibility Measurement A

Página 874 - Digital Cable TV

5 Legacy Command DescriptionsSP (Frequency Span)Legacy Products8560 series, 8566A/B, 8568A/BDescriptionChanges the total displayed frequency range sy

Página 875

5 Legacy Command DescriptionsSP (Frequency Span)Couplings If resolution and video bandwidths are coupled to the span width, the bandwidths changewith

Página 876 - :INSTrument:NSELect?

5 Legacy Command DescriptionsSRQ (Service Request)SRQ (Service Request)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSends a service r

Página 877 - :INSTrument:CATalog?

5 Legacy Command DescriptionsSS (Center Frequency Step Size)SS (Center Frequency Step Size)Syntax8560 series:8566A/B, 8568A/B:Legacy Products8560 ser

Página 878

5 Legacy Command DescriptionsSS (Center Frequency Step Size)DescriptionSpecifies the center frequency step size.Format SS <real>HZ|KHZ|MHZ|GHZ|

Página 879

1 About the AnalyzerWindow Control KeysWindow Control KeysThe instrument provides three front-panel keys for controlling windows. These are:l "M

Página 880 - Application Catalog Options

5 Legacy Command DescriptionsST (Sweep Time)ST (Sweep Time)Syntax8560 series:8566A/B, 8568A/B:Legacy Products8560 series, 8566A/B, 8568A/BDescription

Página 881 - Detailed List of Modes

5 Legacy Command DescriptionsST (Sweep Time)Format ST <real>S|MS|US|SCST UP|DNUP or DN: Increments in a 1,2,5 sequenceST AUTO|MAN (8560 series

Página 882

5 Legacy Command DescriptionsSTB (Status Byte Query)STB (Status Byte Query)SyntaxLegacy Products8560 seriesDescriptionReturns to the controller the d

Página 883 - Mode Preset

5 Legacy Command DescriptionsSTDEV (Standard Deviation of Trace Amplitudes)STDEV (Standard Deviation of Trace Amplitudes)SyntaxLegacy Products8566A/B

Página 884 - How-To Preset

5 Legacy Command DescriptionsSUM (Sum)SUM (Sum)Syntax8560 Series:8566A/B, 8568A/B:Legacy Products8560 seriesDescriptionReturns the sum of all the tra

Página 885

5 Legacy Command DescriptionsSV (Save State)SV (Save State)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSaves the current state of the instrument

Página 886 - Mode Setup

5 Legacy Command DescriptionsSWPCPL (Sweep Couple)SWPCPL (Sweep Couple)SyntaxLegacy Products8560 seriesDescriptionSelects either a stimulus-response

Página 887 - Cmd Error

5 Legacy Command DescriptionsT1 [one] (Free Run Trigger)T1 [one] (Free Run Trigger)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSets the instrumen

Página 888 - Cmd Error Log

5 Legacy Command DescriptionsT2 [two] (Line Trigger)T2 [two] (Line Trigger)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSets the instrument sweep

Página 889 - Preferences

5 Legacy Command DescriptionsT3 [three] (External Trigger)T3 [three] (External Trigger)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSets the instr

Página 890 - Atten Offset

1 About the AnalyzerWindow Control KeysZoom is a toggle function. Pressing this key once increases the size of the selected window;pressing the key a

Página 891 - Limit Swp Time

5 Legacy Command DescriptionsT4 [four] (Video Trigger)T4 [four] (Video Trigger)SyntaxLegacy Products8566A/B, 8568A/BDescriptionSets the instrument sw

Página 892 - Global Settings

5 Legacy Command DescriptionsTA (Trace A)TA (Trace A)SyntaxLegacy Products8566A/B, 8568A/BDescriptionReturns trace A amplitude values from the instru

Página 893

5 Legacy Command DescriptionsTB (Trace B)TB (Trace B)SyntaxLegacy Products8566A/B, 8568A/BDescriptionReturns trace B amplitude values from the instru

Página 894 - :GLOBal:DEFault

5 Legacy Command DescriptionsTDF (Trace Data Format)TDF (Trace Data Format)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionFormats trace

Página 895 - Peak Search

5 Legacy Command DescriptionsTH (Threshold)TH (Threshold)Syntax8560 series:8566A/B, 8568A/B:Legacy Products8560 series, 8566A/B, 8568A/BDescriptionBl

Página 896 - Next Pk Right

5 Legacy Command DescriptionsTH (Threshold)Format TH <real>DM|MV|UV|DBTH UP|DNUP or DN increments by one step sizeTH ON|OFF (8560 series only)T

Página 897 - Mkr->CF

5 Legacy Command DescriptionsTHE (Threshold Enable)THE (Threshold Enable)SyntaxLegacy Products8566A/B, 8568A/BDescriptionTurns the threshold on or of

Página 898 - Peak Criteria

5 Legacy Command DescriptionsTIMEDATE (Time Date)TIMEDATE (Time Date)SyntaxLegacy Products8560 seriesDescriptionSets and returns the date and time of

Página 899 - “Peak Search” Criteria

5 Legacy Command DescriptionsTITLE (Title)TITLE (Title)SyntaxLegacy Products8560 seriesDescriptionActivates the screen title mode, enabling you to en

Página 900 - “Next Peak” Criteria

5 Legacy Command DescriptionsTM (Trigger Mode)TM (Trigger Mode)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionSelects a trigger mode: f

Página 901

1 About the AnalyzerWindow Control KeysRemote Command:DISPlay:WINDow[:SELect] <number>:DISPlay:WINDow[:SELect]?Example :DISP:WIND 1Preset 1Min

Página 902

5 Legacy Command DescriptionsTM (Trigger Mode)Format TM FREE|VID|LINE|EXT|TVTM?Query Data Type FREE|VID|LINE|EXT|TVSCPI EquivalentCommands:TRIGger[:S

Página 903

5 Legacy Command DescriptionsTRA (Trace Data Input and Output)TRA (Trace Data Input and Output)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDesc

Página 904 - Peak Sort

5 Legacy Command DescriptionsTRB (Trace Data Input and Output)TRB (Trace Data Input and Output)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDesc

Página 905 - Peak Readout

5 Legacy Command DescriptionsTRC (Trace Data Input and Output)TRC (Trace Data Input and Output)SyntaxLegacy Products8566A/B, 8568A/BDescriptionTransf

Página 906

5 Legacy Command DescriptionsTRDSP (Trace Display)TRDSP (Trace Display)SyntaxLegacy Products8566A/B, 8568A/BDescriptionDisplays a trace or turns it o

Página 907 - Continuous Peak Search

5 Legacy Command DescriptionsTRIGPOL (Trigger Polarity)TRIGPOL (Trigger Polarity)SyntaxLegacy Products8560 seriesDescriptionSelects the edge (positiv

Página 908

5 Legacy Command DescriptionsTRSTAT (Trace State)TRSTAT (Trace State)SyntaxLegacy Products8566A/B, 8568A/BDescriptionReturns trace states to the cont

Página 909 - Pk-Pk Search

5 Legacy Command DescriptionsTS (Take Sweep)TS (Take Sweep)SyntaxLegacy Products8560 series, 8566A/B, 8568A/BDescriptionStarts and completes one full

Página 910 - Min Search

5 Legacy Command DescriptionsVAVG (Video Average)VAVG (Video Average)Syntax8560 series:8566A/B, 8568A/B:Legacy Products8560 series, 8566A/B, 8568A/BD

Página 911 - Peak Search All Traces

5 Legacy Command DescriptionsVAVG (Video Average)Condition Legacy Spectrum Analyzers N9061A applicationAll conditions. 8566 and 8568 only - Original

Página 912 - Quick Save

Table Of ContentsT 97U 100IEEE 488.2 Common Commands 101All 101Clear Status 103Standard Event Status Enable 104Standard Event Status Register Query 10

Página 913

1 About the AnalyzerMouse and Keyboard ControlMouse and Keyboard ControlIf you do not have access to the instrument front-panel, there are several wa

Página 914

5 Legacy Command DescriptionsVAVG (Video Average)[:SENSe]:AVERage:COUNT <integer>(See "Average/Hold Number" on page 788)Preset 100, O

Página 915

5 Legacy Command DescriptionsVB (Video Bandwidth)VB (Video Bandwidth)Syntax8560 series:8566A/B, 8568A/B:Legacy Products8560 series, 8566A/B, 8568A/BD

Página 916

5 Legacy Command DescriptionsVB (Video Bandwidth)When auto coupled, the video bandwidth is calculated as ResolutionBandwidth x VideoResolution Bandw

Página 917 - From File…

5 Legacy Command DescriptionsVBO (Video Bandwidth Coupling Offset)VBO (Video Bandwidth Coupling Offset)SyntaxLegacy Products8566A/B, 8568A/BDescripti

Página 918

5 Legacy Command DescriptionsVBR (Video Bandwidth to Resolution Bandwidth Ratio)VBR (Video Bandwidth to Resolution Bandwidth Ratio)SyntaxLegacy Produ

Página 919

5 Legacy Command DescriptionsVIEW (View Trace)VIEW (View Trace)Syntax8560 series:8566A/B, 8568A/B:Legacy Products8560 series, 8566A/B, 8568A/BDescrip

Página 920

5 Legacy Command DescriptionsVIEW (View Trace):TRACe[1|2|3|4|5|6:DISPlay[:STATe] ON(See "View/Blank " on page 1344)Notes The functions of V

Página 921

5 Legacy Command DescriptionsVTL (Video Trigger Level)VTL (Video Trigger Level)SyntaxLegacy Products8560 seriesDescriptionSets the signal level that

Página 922

5 Legacy Command DescriptionsVTL (Video Trigger Level)Preset 0 dBmNotes Setting a value for VTL sets the trigger mode to VIDEO, even if it was not al

Página 923

5 Legacy Command DescriptionsXCH (Exchange)XCH (Exchange)SyntaxLegacy Products8566A/B, 8568A/BDescriptionExchanges the contents of the source and des

Página 924

1 About the AnalyzerMouse and Keyboard ControlThis method can be used to access any of the front-panel keys by using a mouse; as for example ifyou ar

Página 925

5 Legacy Command DescriptionsXCH (Exchange)510 Remote Language Compatibility Measurement Application Reference

Página 926

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceThe Swept SA measurement uses both swept and FFT analysis,and the frequency and time domain

Página 927

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeasurement Commands and their Results for Swept SAThe INITiate and CONFigure syntax, as w

Página 928

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceCommand n Return Valuemeasurements that have already been combined, in theaveraging calcul

Página 929

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceCommand n Return Value2 Returns Trace 2 data as a series of x,y pairs3 Returns Trace 3 dat

Página 930

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceSwept SA Measurement Descriptionl Swept Spectrum Analysis (Freq Domain): The analyzer swee

Página 931

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleAMPTD Y ScaleThe Amplitude front-panel key activates the Amplitude menu and s

Página 932

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y Scale+23dBm + RL Offset – External Gain (for EXA and CXA)This maximum value is de

Página 933

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y Scaleconfiguration uses a single attenuator with combined mechanical and electroni

Página 934

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y Scale(note that Configuration 2 is not strictly speaking a dual-section attenuator

Página 935

1 About the AnalyzerMouse and Keyboard ControlPC KeyboardIf you have a PC keyboard plugged into the instrument (or via Remote Desktop), certain key c

Página 936

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y Scale[:SENSe]:POWer[:RF]:ATTenuation?[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0

Página 937

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y Scalebe increased with the knob and step keys, but not decreased.Max CXA N9000A–50

Página 938

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y Scaleand are negligible for front panel use. See "Using the Electronic Atten

Página 939

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleMore InformationMechanical Attenuator Transition RulesWhen the Electronic Att

Página 940

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleThe “finer steps” advantage of the electronic attenuator is beneficial in opt

Página 941

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleMax Dual attenuator configuration: 24dBSingle attenuator configuration: th

Página 942

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleCOMBined[:SENSe]:POWer[:RF]:RANGe:OPTimize:ATTenuation?Notes The SCPI paramet

Página 943

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleExample :POW:RANGe:OPT:ATT OFFInitial S/W Revision Prior to A.02.00Elec Atten

Página 944

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleExample POW:ATT:STEP 2Notes Note this feature works like a 1-N choice from th

Página 945

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleScale / DivSets the units per vertical graticule division on the display. Thi

Página 946

1 About the AnalyzerMouse and Keyboard ControlFront-panel key Key codeSingle CTRL+ALT+SCont CTRL+ALT+CZoom CTRL+SHIFT+ZNext Window CTRL+SHIFT+NSplit

Página 947

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleDependencies If Normalize is on, Scale Type forced to Log and is grayed out.C

Página 948

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleCouplings The active marker position determines where the centering will be a

Página 949

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleKey PathAMPTD Y ScaleScope Meas GlobalRemote Command[:SENSe]:POWer[:RF]:PADJu

Página 950

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleY Axis UnitDisplays the menu keys that enable you to change the vertical (Y)

Página 951

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y Scaletrace 1 is on the fifth graticule line from the top, it represents 5 dBm and

Página 952

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaledBmVSets the amplitude unit for the selected amplitude scale (log/lin) to dBm

Página 953

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleVSets the amplitude unit for the selected amplitude scale (log/lin) to volt.K

Página 954

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaledBµASets the amplitude unit for the selected amplitude scale (log/lin) to dBµ

Página 955

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleKey PathAMPTD Y Scale, Y Axis UnitDependencies Grayed out if no Amplitude Cor

Página 956

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleDependencies Grayed out if no Amplitude Correction with an Antenna Unit is on

Página 957

1 About the AnalyzerMouse and Keyboard ControlFront-panel key Key code1 12 23 34 45 56 67 78 89 90 0This is a pictorial view of the table:54 Remote L

Página 958

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleExample UNIT:POW DBUAMDependencies Grayed out if no Amplitude Correction with

Página 959

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleMore InformationOffsets are used when gain or loss occurs between a device un

Página 960

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleSelecting the Low Noise Path Enable option provides a lower noise floor, espe

Página 961

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleStandard PathThis path gives the best remote-control throughput, minimizes ac

Página 962

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleRemote Command[:SENSe]:POWer[:RF]:MW:PRESelector[:STATe] ON|OFF|0|1[:SENSe]:P

Página 963

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y Scalethe key is not shown.If a POW:GAIN:BAND FULL command is sent when a low band

Página 964

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAMPTD Y ScaleKey PathAMPTD Y Scale, Internal PreampExample :POW:GAIN ON:POW:GAIN:BAND FULL

Página 965

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAuto CoupleAuto CoupleThe Auto Couple feature provides a quick and convenient way to autom

Página 966

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceAuto CoupleAuto/Man 1-of-N keysAn Auto/Man 1-of-N key allows you to manually pick from a l

Página 967

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceBWBWThe BW key opens the bandwidth menu, which contains keys to control the ResolutionBand

Página 968

1 About the AnalyzerInstrument Security & Memory VolatilityInstrument Security & Memory VolatilityIf you are using the instrument in a secure

Página 969

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceBWExample BAND 1 KHZBAND?BWID:AUTO ONBWID:AUTO?Notes For numeric entries, all RBW Types ch

Página 970

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceBWMore InformationWhen the Res BW is manually selected, it may be returned to the coupled

Página 971

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceBWAn * is displayed next to the VBW annotation when certain detector types (Average, EMI A

Página 972

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceBWMax 50 MHzDefault Unit HzBackwards CompatibilityNotesFor backwards compatibility this co

Página 973

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceBWExample BAND:VID:RAT 2BAND:VID:RAT?BAND:VID:RAT:AUTO 0BAND:VID:RAT:AUTO?Notes The values

Página 974

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceBW1. If the detector is Peak and the EMC Standard is set to either CISPR or MIL, use 10.0

Página 975

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceBW[:SENSe]:FREQuency:SPAN:BANDwidth[:RESolution]:RATio:AUTO?Example FREQ:SPAN:BAND:RAT 200

Página 976

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceBWFilter TypeBesides the familiar Gaussian filter shape, there are certain special filter

Página 977

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceBWWhen the Flattop filter type is chosen, a new set of 134 RBW hardware settings are avail

Página 978

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceBWRemote Command[:SENSe]:BANDwidth|BWIDth:TYPE DB3|DB6|IMPulse|NOISe[:SENSe]:BANDwidth|BWI

Página 979

1 About the AnalyzerInstrument Security & Memory Volatility56 Remote Language Compatibility Measurement Application Reference

Página 980

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceBWThe Filter Type menu lets you choose the filter bandwidth (–3 dB, –6 dB, Noise or Impuls

Página 981

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceBWKey PathBW, RBW Control, Filter BWExample BAND:TYPE DB3Readback –3 dBInitial S/W Revisio

Página 982 - Data (Import)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceBW562 Remote Language Compatibility Measurement Application Reference

Página 983

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceCont (Continuous Measurement/Sweep)Cont (Continuous Measurement/Sweep)Sets the analyzer fo

Página 984

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceCont (Continuous Measurement/Sweep)number k stops incrementing. A measurement average usua

Página 985

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFileFileOpens a menu that enables you to access various standard and custom Windows functi

Página 986

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFileOr use Alt-Tab: press and hold the Alt key and press and release the Tab key until th

Página 987

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFileParameter Name Print ThemesParameter Type EnumMode AllRemote Command:SYSTem:PRINt:THEM

Página 988 - TRACE6,<filename>

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFileRestore DownThis key allows you to Restore Down the Instrument Application and reverse

Página 989

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFileKey PathFileMode AllNotesThe Instrument Application willclose. No further SCPI command

Página 990

2 About the N9061A MeasurementApplicationThis chapter provides general information about the N9061Aapplication. It includes the following topics:l &qu

Página 991

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ ChannelFREQ ChannelAccesses a menu of keys that allow you to control the Frequency an

Página 992

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ ChannelState Saved Saved in instrument stateMin Hardware dependent; Zone Span depende

Página 993

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ ChannelPreset 13.255 GHzState Saved Saved in instrument stateMin Start Frequency of t

Página 994

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ ChannelCenter FreqSets the frequency that corresponds to the horizontal center of the

Página 995

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ ChannelPreset and Max values are dependent on Hardware Options (5xx)If no terminator

Página 996

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ Channel503(N9000A)1.505 GHz 3.0 GHz 3.08 GHz507(all but N9000A)3.505 GHz 7.0GHz 7.1

Página 997

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ ChannelOption when turn TGon and can’ttune belowwhile TG on)clipped to thisFreq when

Página 998

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ Channelis independent in each mode and common across all the measurements in the mode

Página 999 - :INITiate:RESTart

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ Channelcome back to the band you were in with the Center Frequency that you had. How

Página 1000 - More Information

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ ChannelPreset 0 HzState Saved Saved in instrument state.Min –40.049995 MHzMax 40.0499

Página 1001

2 About the N9061A Measurement ApplicationN9061A Application DescriptionN9061A Application DescriptionN9061A is a Remote Language Compatibility appli

Página 1002 - To File . .

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ Channelsettings of Source Numerator, Source Denominator and Power Sweep.Couplings In

Página 1003

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ ChannelDefault Unit HzStatus Bits/OPCdependenciesNon-overlappedInitial S/W Revision P

Página 1004 - Save, State

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ Channelsettings of Source Numerator, Source Denominator and Power Sweep.Couplings In

Página 1005

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ Channelthe current mixer setup. It can be queried with the SCPI command :FREQ:STOP? M

Página 1006

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ ChannelDependencies Span, RBW, Center frequencyIf the electronic/soft attenuator is e

Página 1007

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ ChannelScope Meas GlobalRemote Command[:SENSe]:FREQuency:OFFSet <freq>[:SENSe]:

Página 1008

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceFREQ Channeloffset, even on a trace that is not updating will immediately change all of th

Página 1009

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputInput/OutputThe Input/Output features are common across multiple Modes and Mea

Página 1010

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Outputcurrently selected input.BackwardsCompatibility SCPI[:SENSe]:FEED IQ|IONLy|QO

Página 1011

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Output2. When INPut:MIXer INTernal is received, SENSe:FEED RF is executed.3. When IN

Página 1012

2 About the N9061A Measurement ApplicationDocumentation for the N9061A applicationDocumentation for the N9061A applicationScope of this DocumentThis

Página 1013

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputYY is AC or DCZZ is 50Ω or 75ΩInitial S/W Revision Prior to A.02.00Input Z Cor

Página 1014

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputRF CouplingSpecifies alternating current (AC) or direct current (DC) coupling

Página 1015

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputOn models that are always DC coupled, such as millimeter wave models (frequenc

Página 1016

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputThe Agilent 113x Series active probes can be used for both single ended and di

Página 1017

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Outputcalibrated together. The probe calibration is done for a specific channel (I o

Página 1018

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputThe [:SENSe]:FEED RF|IQ|IONLy|QONLy|AREFerence|IFALign command supports all pa

Página 1019

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputI/Q PathSelects which I/Q input channels are active. The LED next to each I/Q

Página 1020

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputNotes For R&S FSQ-B71 compatibilityPreset IQInitial S/W Revision Prior to

Página 1021

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputI SetupAccess the channel setup parameters for the I channel.Key PathInput/Out

Página 1022 - Register 1 thru Register 16

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputRemote Command:INPut[1]:IQ:BALanced[:STATe] OFF|ON|0|1:INPut[1]:IQ:BALanced[:S

Página 1023

Table Of ContentsLegacy Products 210Description 210A3 [three] (View Mode for Trace A) 211Syntax 211Legacy Products 211Description 211A4 [four] (Blank

Página 1024

2 About the N9061A Measurement ApplicationGeneral Rules and LimitationsGeneral Rules and LimitationsThe N9061A application has been designed to emula

Página 1025

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputI SkewSets the skew factor for the I channel. The skew will shift the channel

Página 1026

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputRemote Command[:SENSe]:CORRection:IQ:I|Q:ATTenuation:RATio <real>[:SENSe

Página 1027

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputKey PathInput/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, CouplingReadb

Página 1028

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Output:INP:IQ:DIFF OFF:INP:IQ:IMP 50Notes Provided for E4406A code compatibility.The

Página 1029

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Outputbe identical. This does not apply to Probe settings or to parameters that are

Página 1030

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Output159;Settings Alert;I/Q mismatch:Differential.Couplings Some active probes incl

Página 1031

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputThis is unaffected by Preset but is set to the default value on a "Restor

Página 1032

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputAttenuationThe attenuation is part of the calibration data stored with the pro

Página 1033

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputCalibrateInvokes the guided probe calibration. The guided probe calibration is

Página 1034

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Outputstate. When Differential Input is on, both the data for the probe attached to

Página 1035 - Edit Register Names

2 About the N9061A Measurement ApplicationGeneral Rules and LimitationsCouplingsFor optimal use of the X-Series instrument, N9061A uses the auto coup

Página 1036 - Trace (+State)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputThe guided calibration (front panel only) will show connection diagrams and gu

Página 1037 - TRACE6

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Outputcommand.State Saved No.Initial S/W Revision Prior to A.02.00ExitExits the cali

Página 1038 - Data (Export)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Outputstatus. The calibration data is stored permanently (survives a power cycle) an

Página 1039 - Amplitude Correction

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Output[:SENSe]:CORRection:IQ:I|Q:ATTenuation?Example Set the attenuation for the cur

Página 1040

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputCouplingSome probe types allow coupling to reject low frequencies. This will f

Página 1041

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputLFR2Turns on low frequency rejection, rejecting signal component lower than 0.

Página 1042

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputKey PathInput/Output, I/Q, I Setup | Q Setup, I Probe | Q ProbeRemote Command:

Página 1043

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputRemote Command[:SENSe]:FEED:AREFerence REF50|REF4800|OFF[:SENSe]:FEED:AREFeren

Página 1044

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputKey PathInput/Output, RF CalibratorExample :FEED:AREF REF50Readback 50MHzInit

Página 1045

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputEntering an External Gain value does not affect the Reference Level, therefore

Página 1046

2 About the N9061A Measurement ApplicationGeneral Rules and Limitationsconcept, except that the value to be passed as a parameter to the next command

Página 1047

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputKey PathInput/Output, External GainRemote Command[:SENSe]:CORRection:SA[:RF]:G

Página 1048

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputState Saved Saved in instrument state.Min –100dBMax 100dBReadback MS, <Ex

Página 1049

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputThis key is grayed out in modes that do not support BTS.Preset This is unaffec

Página 1050

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputExample Set the I Ext Gain to 10 dBCORR:IQ:I:GAIN 10Set the I Ext Gain to –10

Página 1051

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputRestore Input/Output DefaultsThis selection causes the group of settings and d

Página 1052

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputKey PathInput/Output, CorrectionsMode SA, DVB-T/H, DTMB, SEQAN, TDSCDMADepende

Página 1053

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputSelect CorrectionSpecifies the selected correction. The term "selected co

Página 1054

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputSelect CorrectionSpecifies the selected correction. The term "selected co

Página 1055

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputExample SENS:CORR:CSET1 ONDependencies Turning this on automatically turns on

Página 1056

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputSelect CorrectionSpecifies the selected correction. The term "selected co

Página 1057

2 About the N9061A Measurement ApplicationGeneral Rules and LimitationsDevice Clear is supported by N9061A, and causes a mode preset of the instrumen

Página 1058 - Measurement Results

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputSelect CorrectionSpecifies the selected correction. The term "selected co

Página 1059

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputAntenna UnitFor devices (like antennas) that make measurements of field streng

Página 1060

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputdBµV/mSets the antenna unit todBµV/m. If this correction is turned on, and Ap

Página 1061

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputExample :CORR:CSET:ANT GAUSReadback "dBG"Initial S/W Revision A.02.

Página 1062

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputTo effect this correction, the goal, for any particular start and stop frequen

Página 1063

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputIf we interpolate on a linear scale, we assume that the two points are connect

Página 1064

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputExample :CORR:CSET1:COMM "this is a comment"Notes 60 chars max; may

Página 1065

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputCorrections arrays are not affected by a Preset, because they are in the Input

Página 1066

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputDelete All CorrectionsErases all correction values for all 4 Amplitude Correct

Página 1067

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputThe values sent in the command will totally replace all existing correction po

Página 1068

2 About the N9061A Measurement ApplicationHardware and Firmware Requirements for N9061AHardware and Firmware Requirements for N9061AFor maximum compa

Página 1069

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputIf an external frequency reference is being used, you must enter the frequency

Página 1070

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputSenseThe external reference is used if a valid signal is sensed at the Ext Ref

Página 1071

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputFor the instrument to stay locked, the value entered must be within 5 ppm of t

Página 1072

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Output(60 Hz), to take advantage of that superior performance. When using an externa

Página 1073

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Output--:INIT:CONT OFF:INIT:IMM;*WAI?--:INIT:CONT OFF:READ?--:INIT:CONT OFF:MEASure?

Página 1074

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputKey PathInput/Output, Freq Ref InMode AllRemote Command[:SENSe]:ROSCillator:CO

Página 1075 - Peak Table

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Output:READ?--:INIT:CONT OFF:MEASure?When the acquisition ceases, in addition to the

Página 1076 - • Peak Sort (Freq

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputState Saved Not saved in instrument stateReadback Normal|Stop AcqInitial S/W R

Página 1077

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputWhen the acquisition ceases, in addition to the error condition(s) described a

Página 1078

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputOutput ConfigAccesses keys that configure various output settings, like the fr

Página 1079 - MeasResult

2 About the N9061A Measurement ApplicationHardware and Firmware Requirements for N9061AProduct Name Instrument Model Number Upper Frequency Limit Fir

Página 1080

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputPolaritySets the output to the Trig 1 Out or Trig 2 Out connector to trigger o

Página 1081

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputMeasuringSelects the Measuring trigger signal to be output to the Trig 1 Out o

Página 1082

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputExample TRIG1:OUTP GATEReadback GateInitial S/W Revision Prior to A.02.00Odd/E

Página 1083

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputState Saved Saved in instrument stateInitial S/W Revision Prior to A.02.00Pola

Página 1084

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputExample TRIG1:OUTP HSWPReadback SweepingInitial S/W Revision Prior to A.02.00M

Página 1085

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputGateSelects the gate signal to be output to the Trig 1 Out or Trig 2 Out conne

Página 1086

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputState Saved Saved in Input/Output StateReadback line 1-of-N selection [variabl

Página 1087

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputOUTPut:ANALog:AUTO?Example OUTP:ANAL:AUTO ONPreset ONState Saved Saved in Inpu

Página 1088

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Outputretrace, as the analyzer remains tuned to the Center Frequency and does not sw

Página 1089

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputDependencies Because the Log Video output uses one of the two IF processing ch

Página 1090

2 About the N9061A Measurement ApplicationInstalling the N9061A ApplicationInstalling the N9061A ApplicationN9061A is a licensed application for X-Se

Página 1091

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Outputretrace (after a sweep and before the next sweep starts).This function depends

Página 1092

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputDigital BusThis menu allows you to configure the LVDS connector located on the

Página 1093

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputKey PathInput/Output, Output ConfigRemote Command:OUTPut:IQ:OUTPut IQ1|IQ250|O

Página 1094

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputKey PathInput/Output, Output Config, I/Q Cal OutReadback OffInitial S/W Revisi

Página 1095

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Outputcauses the aux output type to be offReadback Text OffInitial S/W Revision A.04

Página 1096

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/Outputat the bottom of the bandwidth will “fold”. For example, with a 40 MHz bandw

Página 1097

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceInput/OutputKey PathInput/Output, Output Config, Aux IF OutExample OUTP:AUX LOGVideocauses

Página 1098

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerMarkerThe Marker key accesses the Marker menu. A marker can be placed on a trace to

Página 1099

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerPreset OFF (all markers)State Saved The marker control mode is saved in instrument s

Página 1100

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerNotes If no suffix is sent it will use the fundamental units for the current marker

Página 1101

2 About the N9061A Measurement ApplicationInstalling the N9061A ApplicationlOnline assistance: http://www.agilent.com/find/assistlIf you do not have

Página 1102

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerThe entered value in Trace Points is immediately translated into the current X Axis

Página 1103

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker:CALCulate:MARKer[1]|2|...12:Y?Example CALC:MARK2:MODE POS turns on marker 2 as a no

Página 1104

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerSetting or Querying the Marker Z PositionThe command below sets the Marker Z positio

Página 1105

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkeroffscreen they always returned a valid result. In the X-Series, markers which are o

Página 1106

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerSelect MarkerSpecifies the selected marker. The term “selected marker” is used throu

Página 1107 - Save As . .

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerSelect MarkerSpecifies the selected marker. The term “selected marker” is used throu

Página 1108

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerSelect MarkerSpecifies the selected marker. The term “selected marker” is used throu

Página 1109

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerSelect MarkerSpecifies the selected marker. The term “selected marker” is used throu

Página 1110

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerRemote Command:CALCulate:MARKer[1]|2|...12:STATe OFF|ON|0|1:CALCulate:MARKer[1]|2|..

Página 1111

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker• If a Marker Function is on, the X Axis value and the Y-axis function result the ma

Página 1112

2 About the N9061A Measurement ApplicationSetting up N9061ASetting up N9061ATo set up your X-Series instrument for emulation of one of the supported

Página 1113 - Screen Image

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerFixed Marker Z Axis ValueThe Marker Z position determines which of the 301 traces (0

Página 1114 - :MMEMory:STORe:SCReen:THEMe?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerFixed Marker X Axis ValueKey PathMarker, FixedExample :CALC:MARK:MODE FIX sets Marke

Página 1115

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerposition, will not move off screen in the Spectrogram Window if the storage size is

Página 1116 - Save As…

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerExample :CALC:MARK:MODE FIX sets Marker 1 to Fixed.Notes See the description under t

Página 1117

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerIf Spectrogram is on, the marker result block has a third line displaying the time v

Página 1118

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerPropertiesOpens a menu used to set certain properties of the selected marker.Key Pat

Página 1119 - :MMEMory:CDIRectory?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerKey PathMarkerNotes The selected marker is remembered even when not in the Marker me

Página 1120

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerKey PathMarkerNotes The selected marker is remembered even when not in the Marker me

Página 1121

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerKey PathMarkerNotes The selected marker is remembered even when not in the Marker me

Página 1122

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerKey PathMarkerNotes The selected marker is remembered even when not in the Marker me

Página 1123

2 About the N9061A Measurement ApplicationHints and TipsHints and TipsThis section provides hints and tips that will help you get the most from the X

Página 1124 - RF Output

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerRange (for SCPI command): 1 to 12. If the range is exceeded the value is clipped.Cou

Página 1125 - Amplitude

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerPreset Marker 1State Saved The number of the selected marker is saved in instrument

Página 1126 - Power Sweep

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerPreset Marker 1State Saved The number of the selected marker is saved in instrument

Página 1127

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerPreset Marker 1State Saved The number of the selected marker is saved in instrument

Página 1128 - Amptd Offset

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerPreset Marker 1State Saved The number of the selected marker is saved in instrument

Página 1129 - Frequency

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerState Saved Saved in instrument stateBackwards CompatibilityNotesThe X Axis Scale of

Página 1130 - Multiplier Numerator

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerFrequencySets the marker X Axis scale to Frequency, displaying the absolute frequenc

Página 1131 - Source Sweep Reverse

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerauto setting for time domain traces. In a delta-marker mode it is the (sweep) time i

Página 1132 - Freq Offset

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerSee "Auto Init OFF" on page 699.Key PathMarker, PropertiesRemote Command:C

Página 1133 - Source Mode

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerThis flowchart makes it clear that putting all lower-numbered traces in View is the

Página 1134 - :INSTrument:SOURce[:SELect]?

Table Of ContentsDescription 219ACPLOWER (Lower Adjacent Channel Power) 220Syntax 220Legacy Products 220Description 220ACPMAX (Maximum Adjacent Channe

Página 1135 - Select Source

2 About the N9061A Measurement ApplicationHints and TipsChanging ModesAfter changing into or out of N9061A mode, allow at least a 1 second delay befo

Página 1136 - Point Trigger

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerIf Auto Init is set to On for a marker and that marker is on, that marker’s Marker T

Página 1137 - Select Highlighted Source

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerMin 1Max 6Readback line [TraceN, Auto Init] or [TraceN, Manual] where N is the trace

Página 1138

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerThis flowchart makes it clear that putting all lower-numbered traces in View is the

Página 1139 - Source Preset

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerIf Auto Init is set to On for a marker and that marker is on, that marker’s Marker T

Página 1140 - :SOURce:PRESet

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerMin 1Max 6Readback line [TraceN, Auto Init] or [TraceN, Manual] where N is the trace

Página 1141 - :SOURce:SETtings?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerThis flowchart makes it clear that putting all lower-numbered traces in View is the

Página 1142 - SPAN X Scale

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerIf Auto Init is set to On for a marker and that marker is on, that marker’s Marker T

Página 1143

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerMin 1Max 6Readback line [TraceN, Auto Init] or [TraceN, Manual] where N is the trace

Página 1144 - Span Presets

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerThis flowchart makes it clear that putting all lower-numbered traces in View is the

Página 1145 - Zoom Span

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerIf Auto Init is set to On for a marker and that marker is on, that marker’s Marker T

Página 1146 - Full Span

2 About the N9061A Measurement ApplicationService and CalibrationService and CalibrationSince the Performance Verification and Adjustment Software us

Página 1147 - Zero Span

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerMin 1Max 6Readback line [TraceN, Auto Init] or [TraceN, Manual] where N is the trace

Página 1148 - Last Span

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerThis flowchart makes it clear that putting all lower-numbered traces in View is the

Página 1149 - Signal Track (Span Zoom)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerIf Auto Init is set to On for a marker and that marker is on, that marker’s Marker T

Página 1150

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerMin 1Max 6Readback line [TraceN, Auto Init] or [TraceN, Manual] where N is the trace

Página 1151

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerThis flowchart makes it clear that putting all lower-numbered traces in View is the

Página 1152 - Sweep/Control

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerIf Auto Init is set to On for a marker and that marker is on, that marker’s Marker T

Página 1153 - [:SENSe]:SWEep:TIME:AUTO?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerMin 1Max 6Readback line [TraceN, Auto Init] or [TraceN, Manual] where N is the trace

Página 1154 - Sweep Setup

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerThis flowchart makes it clear that putting all lower-numbered traces in View is the

Página 1155 - :SWEep:TIME:AUTO:MODE?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerIf Auto Init is set to On for a marker and that marker is on, that marker’s Marker T

Página 1156

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerMin 1Max 6Readback line [TraceN, Auto Init] or [TraceN, Manual] where N is the trace

Página 1157

2 About the N9061A Measurement ApplicationService and Calibration72 Remote Language Compatibility Measurement Application Reference

Página 1158 - Sweep Type

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerThis flowchart makes it clear that putting all lower-numbered traces in View is the

Página 1159 - [:SENSe]:SWEep:TYPE:AUTO?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerIf Auto Init is set to On for a marker and that marker is on, that marker’s Marker T

Página 1160 - Sweep Type Rules

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerthe Y-axis result. Additional information is shown for markers which have marker fun

Página 1161

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerNotes Fixed markers are not counted, but a Fixed marker will have a count stored in

Página 1162 - FFT Width

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkercount query. Note this result may simply mean that the first sweep after the counter

Página 1163 - [:SENSe]:SWEep:FFT:WIDTh?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker5. If the reference marker is not a fixed marker, the display shows the difference b

Página 1164

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerGate TimeControls the length of time during which the frequency counter measures the

Página 1165

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker:CALCulate:MARKer[1]|2|3|4:FCOunt:RESolution:AUTO OFF|ON|0|1 isaccepted and ignored:

Página 1166

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarkerCouplings Sets the selected marker to 1.Preset n/a.Initial S/W Revision Prior to A.0

Página 1167

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionMarker FunctionThe Marker Function key opens up a menu of softkeys that all

Página 1168

3 Programming the AnalyzerThis section provides introductory information about theprogramming documentation included with your product.l"What Pro

Página 1169

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionMore InformationThe units to be used for displaying Marker Function results

Página 1170

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker Functionmeasures the average power across some time interval in zero span.Interval

Página 1171

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker Function.In the X-Series the marker itself has a width attribute, which you set us

Página 1172

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionNote that all of the alias commands described below cause the specified mar

Página 1173

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker Function:CALCulate:MARKer[1]|2|3|4:X:STARt?:CALCulate:MARKer[1]|2|3|4:X:STOP

Página 1174

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker Functionnever offscreen. Users who depended on this clipping behavior by setting B

Página 1175

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionThe value in Trace Points is translated into the current X Axis Scale units

Página 1176

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionKey PathMarkerNotes The selected marker is remembered even when not in the

Página 1177

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionKey PathMarkerNotes The selected marker is remembered even when not in the

Página 1178

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionKey PathMarkerNotes The selected marker is remembered even when not in the

Página 1179

3 Programming the AnalyzerWhat Programming Information is Available?What Programming Information is Available?The X-Series Documentation can be acces

Página 1180

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionKey PathMarkerNotes The selected marker is remembered even when not in the

Página 1181

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionCALC:MARK:Y? returns the y-axis value of the Marker Noise function for mark

Página 1182

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionWhen Band/Interval Power is selected while in the Marker Function Off state

Página 1183

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionCALC:MARK:FUNC? returns the current setting of band function for the marker

Página 1184

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionOn time domain traces, interval density is the average power in the interva

Página 1185

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionIt is legal to change the width of the band even if there is no marker func

Página 1186

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionRemote Command:CALCulate:MARKer[1]|2|...4:X:POSition:SPAN <param>:CAL

Página 1187

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionThe unit of the parameter must match the current domain of the trace the se

Página 1188

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker Function(buckets) by translating back based on the current X Axis Scale settings at

Página 1189

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionPreset If 0, Band/Interval Span is set to 5% of span, when a marker functio

Página 1190

3 Programming the AnalyzerWhat Programming Information is Available?Resource DescriptionAgilent VISAUser's GuideDescribes the Agilent Virtual In

Página 1191

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker Function• If in Auto, then whenever the Span changes, the Band Span for Marker Nois

Página 1192

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionMeasure at MarkerThis key and all the keys in this menu only appear with th

Página 1193

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionDependencies If BW & Avg Type is in an Autocoupled state, the (up to th

Página 1194

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionName DetectorNormal NormalPeak PeakSample SampleNeg Peak Negative PeakRMS A

Página 1195 - Gate On/Off

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker Function• Each detector is then read in successive single-point zero span sweeps, u

Página 1196 - [:SENSe]:SWEep:EGATe[:STATe]?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionPreset OffState Saved Saved in instrument stateReadback Text On|OffInitial

Página 1197 - Gate View On/Off

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionIf the Measure at Marker window is being displayed, and one of the detector

Página 1198

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionSets the detector for measure at marker detector 1 to OffPreset PeakState S

Página 1199

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionReadback Text Detector nameInitial S/W Revision A.02.00Detector 1This menu

Página 1200 - Gate View Setup

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionDetector 1This menu lets you select the detector to be used for Detector 1,

Página 1201 - Gate View Start Time

3 Programming the AnalyzerList of Supported SCPI CommandsList of Supported SCPI CommandsWhen the N9061A application has been selected, the X-Series i

Página 1202 - Gate Length

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionKey PathMarker Function, Measure at Marker, DetectorsRemote Command See &qu

Página 1203 - [:SENSe]:SWEep:EGATe:METHod?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionSets the detector for measure at marker detector 1 to Quasi peak:CALC:MAM:D

Página 1204

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionExample :CALC:MAM:DET2 QPESets the detector for measure at marker detector

Página 1205 - Gate Source

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionDetector 2This menu lets you select the detector to be used for Detector 2,

Página 1206

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionDetector 2This menu lets you select the detector to be used for Detector 2,

Página 1207

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker Functionas a parameter, sets detector 2 to Quasi Peak. If sent with Off as a parame

Página 1208

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionPreset Quasi PeakState Saved Saved in instrument stateReadback Text Detecto

Página 1209

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionExample :CALC:MAM:DET3 QPESets the detector for measure at marker detector

Página 1210

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionDetector 3This menu lets you select the detector to be used for Detector 3,

Página 1211

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionDetector 3This menu lets you select the detector to be used for Detector 3,

Página 1212

3 Programming the AnalyzerList of Supported SCPI CommandsC:CALCulate:BWIDth|BANDwidth:NDB "N dB Points " on page 829:CALCulate:BWIDth|BANDw

Página 1213

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker Functionas a parameter, sets detector 3 to EMI Average. If sent with Off as a para

Página 1214

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionPreset EMI AverageState Saved Saved in instrument stateReadback Text Detect

Página 1215

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionExample :CALC:MAM:DET3 QPESets the detector for measure at marker detector

Página 1216

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionKey PathMarker Function, Measure at Marker, DetectorsRemote Command See &qu

Página 1217

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker Functionand Average Type are currently in Auto. If set to “As Set”, then the curre

Página 1218

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionIf set to “Autocoupled”, then the RBW and Average Type are selected by the

Página 1219

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionIf set to “Autocoupled”, then the RBW and Average Type are selected by the

Página 1220

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker FunctionCenter Presel On/OffThis key controls the automatic centering of the presel

Página 1221

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker ToMarker ToThe Marker -> key accesses menu keys that can copy the current marker

Página 1222

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker ToMkr->CF StepSets the center frequency (CF) step size of the analyzer to the ma

Página 1223

3 Programming the AnalyzerList of Supported SCPI Commands:CALCulate:LLINe{1:6}:COPY "Copy from Limit" on page 814:CALCulate:LLINe{1:6}:DATA

Página 1224

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker ToMkr->StopChanges the stop frequency to the frequency of the selected marker. T

Página 1225

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker Toassign that value to the reference level (in dBm). PSA would just assign the delt

Página 1226

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker ToRemote Command:CALCulate:MARKer[1]|2|...12[:SET]:ZSPan:CENTerExample :CALC:MARK2:

Página 1227

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMarker ToNotes Sending this command selects the subopcoded markerDependencies This functio

Página 1228 - :TRIGger[:SEQuence]:TV:LINE?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeasMeasPressing the Meas key displays a menu of measurements that are available in the cu

Página 1229 - :TRIGger[:SEQuence]:TV:FMODe?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeasThe INITiate command works in a slightly different fashion in the Spectrogram view. I

Página 1230 - NPAL

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeasCommand n Return Value13. Marker 3 value (x,y)14. Marker 4 value (x,y)15. Marker 5 val

Página 1231

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeasSwept SA Measurement Descriptionl Swept Spectrum Analysis (Freq Domain): The analyzer

Página 1232

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupMeas SetupThe Meas Setup key opens a menu of softkeys that allow you to control

Página 1233 - Control Edge/Level

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupMore InformationWhen in Single, the sweep stops when N is reached. You can add m

Página 1234 - Gate Holdoff

3 Programming the AnalyzerList of Supported SCPI Commands{1:12}:FUNCtion:BAND:RIGHt:CALCulate:MARKer{1:12}:FUNCtion:BAND:SPAN"Band/Interval Span

Página 1235 - [:SENSe]:SWEep:EGATe:MINFast?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupKey PathMeas SetupRemote Command[:SENSe]:AVERage:TYPE:AUTO OFF|ON|0|1[:SENSe]:AV

Página 1236

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas Setup1. Trace averaging (see “"Trace/Detector" on page 1321”) averages sig

Página 1237

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupThe equation for trace averaging on the log-pwr scale is shown below, where K is

Página 1238 - [:SENSe]:SWEep:POINts?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupVoltage AvgIn this Average type, all filtering and averaging processes work on t

Página 1239 - Pause/Resume

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupKey PathMeas Setup, LimitsNotes The selected limit is remembered even when not i

Página 1240 - Abort (Remote Command Only)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupNotes The selected limit is remembered even when not in the Limit Menu.Preset Li

Página 1241

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupPreset Limit 1, not affected by Mode Preset, preset by Restore Mode Defaults.Sta

Página 1242 - :SYSTem:SHOW?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas Setup:CALCulate:LLINe[1]|2|...6:DISPlay?Example :CALC:LLIN2:DISP ON turns on the disp

Página 1243 - :SYSTem:ERRor[:NEXT]?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupSelect LimitSpecifies the selected limit. The term “selected limit” is used thro

Página 1244

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupKey PathMeas Setup, Limits, PropertiesNotes The selected limit is remembered eve

Página 1245 - :SYSTem:ERRor:VERBose?

Table Of ContentsADJALL (LO and IF Adjustments) 231Syntax 231Legacy Products 231Description 231AMB (A minus B into A) 232Syntax 232Legacy Products 232

Página 1246

3 Programming the AnalyzerList of Supported SCPI Commands:CALCulate:MARKer{1:12}[:SET]:STEP "Mkr->CF Step" on page 779:CALCulate:MARKer{

Página 1247

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupA trace can have multiple limit lines simultaneously; in that case, only one upp

Página 1248

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupPreset Limits 1 and 2 preset to 1, Limits 3 and 4 preset to 2, Limits 5 and 6 pr

Página 1249

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupTest TraceSelects the trace you want the limit to test. A limit is applied to on

Página 1250 - Hardware

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupExample :CALC:LLIN3:TRAC 2 applies limit 3 to trace 2.Notes When the trace displ

Página 1251 - Power On

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupTest TraceSelects the trace you want the limit to test. A limit is applied to on

Página 1252 - Last State

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupPreset Upper for Line 1, 3, and 5; Lower for Line 2, 4, 6.Not affected by Mode P

Página 1253 - Power On Application

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupFor log amplitude interpolation and log frequency interpolation, the interpolati

Página 1254 - Configure Applications

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupFixed / RelativeOpens a menu which will allow you to specify that the selected l

Página 1255

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupState Saved Saved in instrument state.Initial S/W Revision A.02.00Relative to RL

Página 1256

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupExample :CALC:LLIN:DESC “European Emissions”Dependencies 60 characters maxPreset

Página 1257 - :SYSTem:PUP:PROCess

3 Programming the AnalyzerList of Supported SCPI Commands:CALibration:DATA:BACKup "Perform Backup (Remote Command Only)" on page 1289:CALib

Página 1258 - Restore Power On Defaults

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupWhen the Margin is selected, it may be turned off by pressing the Margin key unt

Página 1259

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupThe table editor will only operate properly if the analyzer is sweeping, because

Página 1260 - Alignments

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupAmplitudeLets you edit the Amplitude of the current row.Key PathMeas Setup, Limi

Página 1261 - Auto Align

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupCopy from LimitCopies an existing limit into the current limit, including all se

Página 1262

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupCopy from LimitCopies an existing limit into the current limit, including all se

Página 1263

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupBuild from TraceBuilds a limit using an existing trace. This command will overwr

Página 1264 - :CALibration:AUTO:MODE?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupExample :CALC:LLIN2:BUIL TRACE1 builds limit line 2 based on the data in trace 1

Página 1265 - :CALibration:AUTO:ALERt?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupNote that an upper limit will be built above the trace, while a lower limit will

Página 1266

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas Setuptraces (Description, Associated Trace, Type, Margin, Interpolation, Relative to

Página 1267

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupOffsetEnters a menu which allows you to offset the limit trace by a specified fr

Página 1268 - Align Now

3 Programming the AnalyzerList of Supported SCPI Commands:CALibration:TEMPerature:LALL? "Show Alignment Statistics" on page 1274:CALibratio

Página 1269

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupExample :CALC:LLIN:OFFS:Y –3 dB sets the Y axis offset to –3 dB.:CALC:LLIN:OFFSe

Página 1270

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupIf the domain of the selected limit does not match the domain of the X Axis, no

Página 1271 - :CALibration:NRF?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupIf the trace is at or within the bounds of all applicable limits, but outside th

Página 1272

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupChanging the X Axis Unit will erase all your limit lines. Are you sure you want

Página 1273 - :CALibration:RF?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas Setup<ampl>Amplitude values units default to dBm. Up to two amplitude values ca

Página 1274 - Show Alignment Statistics

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupLimit Line Fail? (Remote Command Only)Tests a limit line against its associated

Página 1275 - :SYSTem:PON:TIME?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupRemote Command:CALCulate:LIMit[1]|2|...6:CONTrol:POINts?Example :CALC:LIM:CONT:P

Página 1276 - :CALibration:TIME:LRF?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupRemote Command:CALCulate:LIMit[1]|2|...6:UPPer:POINts?Example :CALC:LIM:UPP:POIN

Página 1277 - :CALibration:TEMPerature:LRF?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupExample :CALC:LIM:FAIL?returns a zero if limit line 1’s associated trace has no

Página 1278 - :CALibration:AUTO:TIME:OFF?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas Setup:CALCulate:LLINe:CMODe?Example :CALC:LLIN:CMOD RELmakes all limit lines relative

Página 1279

3 Programming the AnalyzerList of Supported SCPI Commands:DISPlay:MENU[:STATe] "Full Screen" on page 1463:DISPlay:THEMe "Theme" o

Página 1280 - Restore Align Defaults

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupPreset Off, –3.01 dBOFFState Saved The on/off status and the offset value are bo

Página 1281 - Backup or Restore Align Data…

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas Setupskirts. There is one pixel between the arrow and the trace.N dB Points can be us

Página 1282 - Alignment Data Wizard

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas Setup832 Remote Language Compatibility Measurement Application Reference

Página 1283

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupPhNoise OptSelects the LO (local oscillator) phase noise behavior for various de

Página 1284

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas Setup• "All other Models" on page 835Key PathMeas Setup, PhNoise OptRemote

Página 1285

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupSpan > 22 MHz, or whenRBW > 400 kHz, or whenCF ≤ 12 MHz and Span < 495

Página 1286

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupOtherwise, Auto will choose Best Wide-offset Phase NoiseNote that in these model

Página 1287

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas Setupof measurements (especially remote measurements where the center frequency is ch

Página 1288

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupPXA with option EP1: [single loop]Models with option EP2: [medium loop bandwidth

Página 1289

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas Setupparameter is honored and the HIGH state set, and returned to a query, but the Me

Página 1290 - Advanced

3 Programming the AnalyzerList of Supported SCPI CommandsI:ID? Also a legacy command. See "ID (Identify)" on page 314.:INITiate:CONTinuous

Página 1291 - Timebase DAC

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupHigh (Best Log Accy)When ADC dither is set to High, the scale fidelity is especi

Página 1292

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas Setupanalyzer, the dynamic range is always better with it on than off. The Swept IF G

Página 1293

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupExample IF:GAIN:SWEP:AUTO ONPreset ONInitial S/W Revision Prior to A.02.00Low Ga

Página 1294 - I/O Config

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas Setupthe range to either the high or low gain setting. But you must know that your me

Página 1295

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupAutorange (Slower – Follows Signals)Turns the ADC ranging to automatic which pro

Página 1296 - SCPI LAN

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupAnalog Demod Tune & ListenThe Analog Demod Tune & Listen key opens the A

Página 1297 - SICL Server

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupState Saved Saved in instrument state.BackwardsCompatibility SCPI[:SENSe]:DEMod:

Página 1298

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupFMPressing this key, when it is not selected, selects and activates the FM demod

Página 1299 - Reset Web Password

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupDe-emphasis (FM Demod only)The De-emphasis setting controls a single-pole filter

Página 1300 - System IDN Response

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupDe-emphasis (FM Demod only)The De-emphasis setting controls a single-pole filter

Página 1301

3 Programming the AnalyzerList of Supported SCPI Commands:INSTrument[:SELect]:INSTrument[:SELect]?"Mode" on page 860Specify parameter RLC t

Página 1302

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupDe-emphasis (FM Demod only)The De-emphasis setting controls a single-pole filter

Página 1303 - Restore Defaults

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupΦMPressing this key, when it is not selected, selects and activates the ΦM demod

Página 1304

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupOffPressing this key, turns the demodulation function off.Key PathMeas Setup, An

Página 1305 - Restore Misc Defaults

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupDemod State (Remote Command Only)Sets or queries the state of the Analog Demod

Página 1306

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupMore InformationThere are 2 types of noise sources: a Smart Noise Source (SNS),

Página 1307

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupExample SOUR:NOIS:TYPE NORMCouplings If no SNS is connected, this parameter will

Página 1308 - Control Panel…

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas Setup• Set the Detector to AVERAGE• Set the sweep time to 16ms - sets the variance co

Página 1309 - Licensing…

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas Setupdata is issued in printed form when an SNS is purchased or can be read from the

Página 1310

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupMore InformationThere are 2 types of noise sources: a Smart Noise Source (SNS),

Página 1311 - :SYSTem:HID?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMeas SetupCouplings If an SNS is connected, and the Type is set to SNS, this parameter tur

Página 1312 - Security

3 Programming the AnalyzerList of Supported SCPI Commands:MMEMory:STORe:TRACe "Trace (+State)" on page 1036:MMEMory:STORe:TRACe:DATA "

Página 1313 - Diagnostics

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeModeThe Mode key allows you to select the available Measurement Applications or “Modes

Página 1314

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModefollowing::INST:SEL EMI:CONF FSCThis results in the instrument being placed in the EMI

Página 1315 - :SYSTem:TEMPerature:LEXTreme?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeremains allocated even when it is not running, and is not released until the analyzer

Página 1316 - Internet Explorer…

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeIf you are using the Help feature, this mode must be currently active to access its de

Página 1317 - SYSTem:PDOWn [NORMal

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeINST:NSEL 8Initial S/W Revision Prior to A.02.00W-CDMA with HSPA+Selects the W-CDMA wi

Página 1318 - :SYSTem:KLOCk?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeIf you are using the Help feature, this mode must be currently active to access its de

Página 1319 - Date (Remote Command Only)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModePhase NoiseThe Phase Noise mode provides pre-configured measurements for making genera

Página 1320 - Time (Remote Command Only)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeExample INST:SEL ADEMODINST:NSEL 234Initial S/W Revision Prior to A.02.00TD-SCDMA with

Página 1321 - Trace/Detector

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeIf you are using the Help feature, this mode must be currently active to access its de

Página 1322 - [:SENSe]:AVERage[:STATe]?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeDVB-T/H with T2Selects the DVB-T/H mode for measurements of digital video signals usin

Página 1323 - Trace Update Indicator

3 Programming the AnalyzerList of Supported SCPI Commands[:SENSe]:BWIDth|BANDwidth:SHAPe "Filter Type " on page 557[:SENSe]:BWIDth|BANDwidt

Página 1324 - Trace Annunciator Panel

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeKey PathModeExample INST:SEL ISDBTINST:NSEL 239Initial S/W Revision A.03.00CMMBSelects

Página 1325 - Trace Annotation

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeCombined Fixed WiMAXSelects the Combined Fixed WiMAX mode for general purpose measurem

Página 1326 - Select Trace

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeIf you are using the Help feature, this mode must be currently active to access its de

Página 1327

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMode• RFID• Digital satellite video and other satellite signals, radar, LMDS• Up to 400K b

Página 1328 - Clear Write

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeAfter changing into or out of this mode, allow a 1 second delay before sending any sub

Página 1329 - Trace Average

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeINST:NSEL 106Initial S/W Revision A.10.01WLANSelects the WLAN mode for general purpose

Página 1330 - Max Hold

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeGSM/EDGE/EDGE Evo 13 EDGEGSM802.16 OFDMA (WiMAX/WiBro) 75 WIMAXOFDMAVector Signal Anal

Página 1331 - Min Hold

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeThe command must be sequential: i.e. continued parsing of commands cannot proceed unti

Página 1332

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeCurrent Application ModelReturns a string that is the Model Number of the currently se

Página 1333 - View/Blank

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeThe format is the name as the *OPT? or SYSTem:OPTion command: a comma separated list o

Página 1334 - Trace Display State On/Off

3 Programming the AnalyzerList of Supported SCPI Commands[:SENSe]:DEMod "Analog Demod Tune & Listen" on page 845[:SENSe]:DEMod:AM:BANDw

Página 1335

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeString length is COUNt * 7 – 1. (7 = Model Number length + 1 for comma. –1 = no comma

Página 1336

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModeDetailed List of ModesThis topic contains an alphabetical list of measurement applicat

Página 1337

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceModel "Noise Figure" on page 866l "Phase Noise" on page 866l "Rem

Página 1338

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMode PresetMode PresetReturns the active mode to a known state.Mode Preset does the follow

Página 1339

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMode Presetdoes a User Preset. The old PRESet:TYPE command is ignored (without generating

Página 1340 - :TRACe:MODE BLANk

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMode PresetRestore Input/OutputDefaults:SYSTem:DEFault INPut System Menu; RestoreSystem De

Página 1341

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMode SetupMode SetupThis menu allows you to select 8560 series analyzer emulation, or 8566

Página 1342

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMode SetupSelecting any legacy analyzer from this menu performs an instrument preset and s

Página 1343

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMode SetupThis string is limited to the first 20 characters of the input string (message u

Página 1344

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMode SetupKey Path Mode Setup, LoggingRefreshTo update the log page with new entries, sele

Página 1345

3 Programming the AnalyzerList of Supported SCPI Commands[:SENSe]:FREQuency:SPAN:FULL "Full Span " on page 1146[:SENSe]:FREQuency:SPAN:PREV

Página 1346

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMode SetupPreset ONNotes If the selected RLC Language is HP8566A/B or HP8568A/B, setting t

Página 1347 - Detector

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMode SetupSwp Type RuleChanges the Auto rules for determining whether the instrument uses

Página 1348

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMode SetupRestore Mode DefaultsResets the state for the currently active mode by resetting

Página 1349 - [:SENSe]:DETector[:FUNCtion]?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMode SetupWhen the Global Center Freq key is switched to On in any mode, the current mode’

Página 1350

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceMode SetupRemote Command:INSTrument:COUPle:DEFaultExample INST:COUP:DEFBackwardsCompatibil

Página 1351 - Multiple Detectors

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak SearchPeak SearchPressing the Peak Search key displays the Peak Search menu and place

Página 1352 - Average (Log/RMS/V)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak Searchmany times as required to navigate back through the previously accessed menus u

Página 1353

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak SearchNext Pk LeftPressing Next Pk Left moves the selected marker to the nearest peak

Página 1354

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak SearchDependencies Same as specified under Marker ToInitial S/W Revision Prior to A.0

Página 1355 - Quasi Peak

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak Search“Peak Search” CriteriaThis menu lets you decide what kind of search you want to

Página 1356 - EMI Average

Table Of ContentsSyntax 242Legacy Products 242Description 242B2 [two] (Maximum Hold for Trace B) 243Syntax 243Legacy Products 243Description 243B3 [th

Página 1357 - RMS Average

3 Programming the AnalyzerList of Supported SCPI Commands[:SENSe]:RLC:ATTenuation:STATe[:SENSe]:RLC:ATTenuation:STATe?"Atten Offset" on pag

Página 1358 - Preset Detectors

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak SearchExample CALC:MARK:PEAK:SEAR:MODE MAXReadback Highest PeakInitial S/W Revision P

Página 1359 - Clear Trace

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak SearchIn the event that a sequence of trace points with precisely the same values rep

Página 1360 - Preset All Traces

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak SearchPk ThresholdTurns the peak threshold requirement on/off and sets the threshold

Página 1361

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak SearchMore InformationThe Peak Threshold line is green and has the value of the peak

Página 1362 - Math: More Information

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak SearchThe Peak Table is updated after each sweep. The list of peaks in the Peak Table

Página 1363

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak SearchPreset AMPLitudeState Saved Saved in instrument state.BackwardsCompatibility SC

Página 1364

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak SearchMore InformationIf the Display Line (see the Section “View/Display”) is turned

Página 1365

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak SearchAbove Display LineSets the peak table to display only the 20 highest peaks abov

Página 1366 - Power Diff (Op1-Op2)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak SearchKey PathPeak SearchRemote Command:CALCulate:MARKer[1]|2|...12:CPSearch[:STATe]

Página 1367 - Log Offset (Op1 + Offset)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak Searchmeasurement will be allowed to complete (i.e., all the averages taken up to the

Página 1368 - Log Diff (Op1-Op2+Ref)

3 Programming the AnalyzerList of Supported SCPI Commands[:SENSe]:SWEep:TIME:AUTO "Sweep Time" on page 1152[:SENSe]:SWEep:TIME:AUTO:RULes &

Página 1369

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak SearchCALC:MARK:Y? queries the delta amplitude value for marker 1.Notes Turns on the

Página 1370 - Trace Operands

6 RLC Swept SA Measurement Front-Panel &SCPI ReferencePeak SearchRemote Command:TRACe:MATH:PEAK[:DATA]?Example TRAC:MATH:PEAK? Will identify the

Página 1371

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceQuick SaveQuick SaveThe Quick Save front-panel key repeats the most recent save that was p

Página 1372

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceQuick SaveOne of the key features of Auto File Name is that we guarantee that the Auto Fil

Página 1373 - Operand 2

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallRecallThe Recall menu lets you choose what you want to recall, and where you want to

Página 1374

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallThe content of a state file includes all of the settings and data required to return

Página 1375 - Copy/Exchange

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallnumber, the recall functiontries to recall as much as possible and returns a warning

Página 1376 - From Trace

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallas they were when saved. data will be exactly as they werewhen saved. Any traces th

Página 1377

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallEnables you to navigate to the center of the dialog that contains the list of files

Página 1378 - To Trace

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallFrom File…When you press “From File”, the analyzer brings up a Windows dialog and a

Página 1379

3 Programming the AnalyzerList of Supported SCPI Commands:SOURce[:EXTernal]:POWer[:LEVel][:IMMediate][:AMPLitude]"Amplitude" on page 1125:S

Página 1380 - Normalize

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallThe Look In field shows the path from which the file will be recalled and allows you

Página 1381 - Normalize On/Off

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1382

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1383 - Show Ref Trace (Trace 3)

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1384 - Open/Short Cal

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1385 - Norm Ref Posn

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1386

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1387 - Query Trace Data

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1388

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1389 - Remote Command

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1390 - :TRACe:MATH:SMOoth:POINts?

3 Programming the AnalyzerList of Supported SCPI CommandsStion-able:CA-Libration:EXTended:FAILure:PTRansition:STA-Tus:QUE-Stionable:CALibration:EXTen

Página 1391

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1392

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1393 - Trigger Source Presets

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1394

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1395

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1396

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1397 - RF Trigger Source

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1398 - I/Q Trigger Source

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1399

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1400 - Free Run

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallAfter the recall completes, the message "Register <register number> recal

Página 1401 - Video (IF Envelope)

3 Programming the AnalyzerList of Supported SCPI Commands:STATus:QUEStionable:FREQuency:CONDition? "Questionable Frequency Condition" on pa

Página 1402 - Trigger Level

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallKey PathRecall, StateExample *RCL 1Range 1–16 from front panel, 1–128 from SCPIReadb

Página 1403 - Trig Delay

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallA custom name of up to 30 characters entered using the Save, State,Edit Register Nam

Página 1404

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallRegister 1 thru Register 16Selecting any one of these register keys causes the State

Página 1405 - :TRIGger[:SEQuence]:OFFSet?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallRegisters are shared by all modes, so recalling from any one of the registers will c

Página 1406

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallsaved in the Spectrum Analyzer Mode, then after the recall, you will still be in the

Página 1407 - External 1

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallExample *RCL 1Range 1–16 from front panel, 1–128 from SCPIReadback Date and time wit

Página 1408

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecall“(empty)" if no prior save operation has been performed to this register.Initia

Página 1409 - External 2

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallRegister 1 thru Register 16Selecting any one of these register keys causes the State

Página 1410

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallRegisters are shared by all modes, so recalling from any one of the registers will c

Página 1411

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallsaved in the Spectrum Analyzer Mode, then after the recall, you will still be in the

Página 1412 - RF Burst

3 Programming the AnalyzerList of Supported SCPI Commands:STATus:QUEStionable:NTRansition "Questionable Negative Transition " on page 145:S

Página 1413 - Absolute Trigger Level

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallExample *RCL 1Range 1–16 from front panel, 1–128 from SCPIReadback Date and time wit

Página 1414 - Relative Trigger Level

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecall“(empty)" if no prior save operation has been performed to this register.Initia

Página 1415 - Trigger Slope

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallTrace (+State)The Recall Trace (+State) menu lets you choose a register or file from

Página 1416

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallrecalled.Once the state is loaded, the trace data must be loaded. The internal flags

Página 1417

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallIf the .trace file is an "all trace" file, "ToTrace" is ignored

Página 1418

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallIf the .trace file is an "all trace" file, "ToTrace" is ignored

Página 1419

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallIf the .trace file is an "all trace" file, "ToTrace" is ignored

Página 1420

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallRegisters are shared by all modes, so recalling from any one of the registers will c

Página 1421 - Reset Offset Display

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallsaved in the Spectrum Analyzer Mode, then after the recall, you will still be in the

Página 1422 - Sync Source

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallExample *RCL 1Range 1–16 from front panel, 1–128 from SCPIReadback Date and time wit

Página 1423

3 Programming the AnalyzerList of Supported SCPI Commands:SY-STem:CO-MMunicate:LAN:SCPI:TELNet:ENABle"SCPI Telnet" on page 1296:SYSTem:COMM

Página 1424

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecall“(empty)" if no prior save operation has been performed to this register.Initia

Página 1425

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallFrom File…When you press “From File”, the analyzer brings up a Windows dialog and a

Página 1426

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallThe Look In field shows the path from which the file will be recalled and allows you

Página 1427

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1428

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1429

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1430

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1431

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1432 - Entire Frame

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1433 - Standard

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1434

3 Programming the AnalyzerList of Supported SCPI CommandsAPPLication:VMEMory:USED:NAME?:SYSTem:PON:ETIMe? "Query the Elapsed Time since First Po

Página 1435

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1436 - Auto/Holdoff

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1437 - Trig Holdoff

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1438 - :TRIGger[:SEQuence]:HOLDoff?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1439

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1440 - User Preset All Modes

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1441 - Save User Preset

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1442 - View/Display

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1443 - Annotation

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1444 - Meas Bar On/Off

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1445

3 Programming the AnalyzerList of Supported SCPI Commands:TRACe:MATH:MEAN? "Mean Trace Data (Remote Command Only) " on page 1391:TRACe:MATH

Página 1446 - Active Function Values On/Off

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1447 - :DISPlay:ACTivefunc[:STATe]?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallListed below are the functions of the various fields in the dialog, and the correspo

Página 1448 - Graticule

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallUser specified paths are remembered when you leave and return to a Mode and are rese

Página 1449 - Display Line

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecalldata. Pressing the Open key in this menu brings up the Open dialog and Open menu tha

Página 1450 - System Display Settings

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallExample :MMEM:LOAD:CORR 2 "myAmpcor.csv" recalls the Amplitude Correction

Página 1451

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallNotes auto returnDependencies Only Correction 1 may be used to load a Correction tha

Página 1452 - :DISPlay:THEMe?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallNotes auto returnDependencies Only Correction 1 may be used to load a Correction tha

Página 1453 - :DISPlay:BACKlight?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallNotes auto returnDependencies Only Correction 1 may be used to load a Correction tha

Página 1454 - :DISPlay:BACKlight:INTensity?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallTRACE6,<filename>Example :MMEM:LOAD:TRAC DATA TRACE2, "myTrace2.csv"

Página 1455 - Spectrogram

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallSelect TraceThese keys let you pick the Trace into which to import the data; either

Página 1456

3 Programming the AnalyzerList of Supported SCPI Commands:TRIGger[:SEQuence]:FRAMe:RFBurst:SLOPe"Trigger Slope " on page 1429:TRIGger[:SEQu

Página 1457

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallSelect TraceThese keys let you pick the Trace into which to import the data; either

Página 1458 - Representation of Time

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallSelect TraceThese keys let you pick the Trace into which to import the data; either

Página 1459 - Trace Zoom

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallCouplings When a limit line is loaded from mass storage, it is automatically turned

Página 1460

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallAn example of using this menu is: If you select 2 and continue to the File Open menu

Página 1461

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallPreset Not part of Preset, but is reset to LLINE1 by Restore Mode Defaults; survives

Página 1462 - Display Trace

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallOpen…When you press “Open”, the analyzer brings up a Windows dialog and a menu entit

Página 1463 - Full Screen

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallOpen…When you press “Open”, the analyzer brings up a Windows dialog and a menu entit

Página 1464 - :DISPlay:ENABle?

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallOpen…When you press “Open”, the analyzer brings up a Windows dialog and a menu entit

Página 1465

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRecallOpen…When you press “Open”, the analyzer brings up a Windows dialog and a menu entit

Página 1466

6 RLC Swept SA Measurement Front-Panel &SCPI ReferenceRestartRestartThe Restart function restarts the current sweep, or measurement, or set of a

Comentários a estes Manuais