Agilent-technologies AGILENT U2761A Manual do Utilizador Página 93

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 123
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 92
Features and Functions 2
U2761A User’s Guide 75
Generate PSK Modulation Using SCPI Commands
Example 1
-> *CLS; *RST //To reset the U2761A to
default power-on state, this
command can be ignored if
this operation is not
required.
-> PSK:STAT ON //Enables PSK modulation.
-> FUNC SIN //Sets the carrier waveform to
Sine wave.
-> FREQ 1000 //Sets the carrier frequency to
1 kHz.
-> VOLT 5 VPP //Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0 //Sets the output offset to 0.
-> PSK:INT:RATE 10 //Sets the PSK rate to
10 Hz.
-> PSK:DEV 180 //Sets the PSK deviation to
180°.
-> OUTP ON //Turns on output.
Vista de página 92
1 2 ... 88 89 90 91 92 93 94 95 96 97 98 ... 122 123

Comentários a estes Manuais

Sem comentários